Conclusions
Tests have shown that the errors of reproducing the temperature scale on the SPK equipment are related to the following phenomena:
-
1.
Instability of the photoelectric portion of the equipment (noise effects).
-
2.
Incomplete elimination of differences between the two channels of the external optical system (in comparing lamps).
-
3.
Instability of the lamp ribbon brightness (effects of the slope, room temperature, etc.).
-
4.
Nonuniformity of the lamp ribbon temperature over the working region.
-
5.
Inaccurate knowledge of the effective wavelength.
Above values of the error will be made more precise with further improvements in the equipment and technique of testing.
Similar content being viewed by others
Literature cited
A. N. Gordov and E. A. Lapina, Izmeritel'naya Tekhnika No. 5 (1958).
V. V. Kandyba and V. A. Kovalevskii, Dokl. Akad. Nauk SSSR 108, 4 (1956).
V. A. Kovalevskii, G. L. Iosel'son, and V. V. Kandyba, Izmeritel'naya Tekhnika No. 2 (1956).
A. N. Gordov, I. I. Kirenkov, and E. A. Lapina, Trud. VNIIM No. 35/95 (1958).
Rights and permissions
About this article
Cite this article
Kirenkov, I.I., Kovalevskii, V.A. & Krakhmal'nikova, G.A. Reproduction of the color-temperature scale by photometric methods. Meas Tech 3, 112–115 (1960). https://doi.org/10.1007/BF00979043
Issue Date:
DOI: https://doi.org/10.1007/BF00979043