Methods for calibrating against the practical temperature scale in the region of 10–90°K

  • Published:
Measurement Techniques Aims and scope

Conclusions

  1. 1.

    A method for correcting platinum thermometer scales against a standard scale by means of formula

    $$W_T = W_T^k + \sigma (1 + \eta )(1 - W_T^k ) - \frac{{\eta ^2 \sigma ^2 }}{{\eta \sigma + \nu W_T^k }}$$

    is suggested, where W KT is a tabulated function of W stT , and the latter is a function of temperature; σ, η andγ are constants obtained from calibrating the thermometer at the boiling points of helium, hydrogen and oxygen. The method is only applicable if these constants are positive. The error obtained after correction by this method of 10 different grades of platinum, which were experimentally investigated earlier [3], amounts to 0.01° in the range of 12–90°K for sufficiently pure grades of platinum (α>0.003923). For less pure grades of platinum the correction error amounts to several hundredths of a degree.

  2. 2.

    Experimentally obtained data [3] of thermometers made of 10 different grades of platinum were compared with the calculated data:

  3. a)

    Scales calculated by means of the Nernst formula with one constant and the Z-function method (with two constants) were compared with experimental data, thus determining the accuracy and range of applicability of these simple methods;

  4. b)

    scales of thermometer type IKh-4 calculated by more complex correction methods [5, 6, 7] were compared with experimental data, showing the insufficient accuracy of these methods for this type of platinum;

  5. c)

    the comparison of scales corrected by Barber's method [8] indicated the unsatisfactory degree of accuracy of this correction method in the 20–90°K range for sufficiently pure platinum. The correction accuracy at temperatures below 20°K is considerably worse than that obtained by the method suggested by the present author.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

Literature cited

  1. P. G. Strelkov, Bulletin of the Acad. Sci. USSR, Physics series,14, 1, 115 (1950).

    Google Scholar 

  2. D. R. Stull, Fourth Conf. on low temper. calor.Cambr. Mass., Sept. (1949).

  3. D. I. Sharevskaya, P. G. Strelkov, A. S. Borovik-Romanov, D. N. Astrov, and G. Kh. Morozova, Izmeritel'naya Tekhnika, 7 (1960).

  4. H. J. Hoge, Rev. Sci. Instr.,21, 9, 815 (1950).

    Google Scholar 

  5. J. M. Loss and J. A. Morrison, Can. J. Phys.,29, 2, 142 (1951).

    Google Scholar 

  6. H. Van Dijk, Procès-verbaux CIPM,26A, pt. 42 (1959).

  7. H. Schultz, Zeit.Angew. Phys.,9, No. 9, 465 (1957).

    Google Scholar 

  8. C. R. Barber, Procès-verbaux CIPM,26A, pt. 52 (1959).

  9. M. Kohler and Neckar, Z. Phys.,126, 495 (1949).

    Google Scholar 

  10. M. P. Orlova, Izmeritel'naya Tekhnika, 2 (1961).

  11. D. I. Sharevskaya and P. G. Strelkov, Izmeritel'naya Tekhnika, 12 (1960).

  12. Standard Table of T=f (W) for temperatures of 10–90.19°K [in Russian] (Standartgiz, 1959).

  13. Stimson, Procès-verbaux CIPM,21, pt. 117 (1948).

Download references

Authors

Additional information

A. S. Borovik-Romanov participated in the discussion of the results obtained in the present work.

Rights and permissions

Reprints and permissions

About this article

Cite this article

Sharevskaya, D.I., Strelkov, P.G. Methods for calibrating against the practical temperature scale in the region of 10–90°K. Meas Tech 4, 110–115 (1961). https://doi.org/10.1007/BF00978229

Download citation

  • Issue Date:

  • DOI: https://doi.org/10.1007/BF00978229

Keywords

Navigation