Summary
-
1.
Unequal arm lever systems with a gear ratio of 500–2,500 provide in an appropriate assembly the possibility of weighing 5,000 kg weights with an error of the order of 0.001–0.005%.
-
2.
These lever systems can be used as reference grade 1 dynamometers for measuring a force with an error not exceeding ±0.05% of the measured value.
-
3.
The adoption of the above lever dynamometers for metrological measurements will substitute in part the missing link in the measuring system, namely, the grade 1 dynamometers.
Similar content being viewed by others
Literature cited
I. S. Gorpinchenko, Izmeritel'nayaTekhnika No. 1 (1957); No. 5 (1958).
I. S. Gorpinchenko, Application of Unequal Arm Multilever Systems for Measuring Force and Mass [in Russian] (Author's report of his dissertation, Novosibirsk, 1959).
Rights and permissions
About this article
Cite this article
Gorpinchenko, I.S. Reference 1st grade lever-type dynamometer. Meas Tech 3, 945–947 (1960). https://doi.org/10.1007/BF00977515