Abstract
Electronic dynamometry measurements, performed upon intraocular lens (IOL) haptics of prototype one-piece three-loop silicone lenses, accurately defined the relationships between elastic force and haptic displacement. Lens implantations in the capsular bag of dogs (loop span equal to capsular bag diameter, loops undeformed immediately after the operation) were evaluated macrophotographically 5–8 months postoperatively. The highly constant elastic property of silicon rubber permitted quantitative correlation of subsequent in vivo haptic displacement with the resultant force vectors responsible for tissue contraction. The lens optics were well centered in 17 (85%) and slightly offcenter in 3 (15%) of 20 implanted eyes. Of the 60 supporting loops, 28 could be visualized sufficiently well to permit reliable haptic measurement. Of these 28, 20 (71 %) were clearly displaced, ranging from 0.45 mm away from to 1.4 mm towards the lens' optic center. These extremes represented resultant vector forces of 0.20 and 1.23 mN respectively. Quantitative vector analysis permits better understanding of IOL-capsular interactions.
Similar content being viewed by others
References
Apple DJ, Mamalis N, Loftfield K, Googe JM (1984) Complications of intraocular lenses. A historical and histopathological review. Surv Ophthalmol 29:1–54
Baikoff G (1981) Insertion of the Simcoe posterior lens into the capsular bag. Am Intraocular Soc 7:267–269
Draeger J, Guthoff R, Wirt H, Neumann W, Lang GK (1987) Experimentelle Untersuchungen mit neuem Linsenmaterial, neuer Geometrie, insbesondere zur Gewebsverträglichkeit im Tierversuch. In: Jakobi KW, Schott K, Gloor B (eds) Proc erster Kongress der Deutschen Gesellschaft für Intraokularlinsen Implantation. Springer, Berlin Heidelberg New York, pp 15–19
Draeger J, Guthoff R, Abramo F (1989) Zur Biomechanik der Intraokularlinsen-Haptik: Vergleichender Messungen von Elastizitäts- und Rückstellkraft in Abhängigkeit von Materialeigenschaften und Konstruktion. In: Lang GL, Ruprecht KW, Jacobi KW, Schott K (eds) Proceedings des zweiten Kongresses der Deutschen Gesellschaft für Intraokularlinsen Implantation. Enke, Stuttgart, pp 42–44
Drews RC, Kreiner C (1987) Comparative study of the elasticity and memory of intraocular lens loops. J Cataract Refract Surg 13:525–530
Faulkner ED (1986) Early experiments with Staar silicon elastic lens implants. J Cataract Refract Surg 12:36–39
Green WR, McDonnell PJ (1985) Opacification of the posterior capsule. Trans Ophthalmol Soc UK 104:727–739
Guthoff R, Draeger J, Lang GK, Neumann W (1989) Kapselsackgestützte Silikonlinsen. Klinische und histopathologische Ergebnisse nach 8 Monaten Verweildauer im Hundeauge. In: Lang GL, Ruprecht KW, Jacobi KW, Schott K (eds) Proceedings des zweiten Kongresses der Deutschen Gesellschaft für Intraokularlinsen Implantation. Enke, Stuttgart, pp 174–175
Guthoff RF, Abramo F, Draeger J, Chumbley LC (1990) Measurement of elastic resisting forces of IOL haptic loops of varying geometrical design and material composition. J Cataract Refract Surg (in press)
Hartmann C, Krieglstein GH, Kolb M (1989) Kapselsackveränderungen nach endokapsulärer PMMA-Disk-Linsenimplantation. In: Lang GL, Ruprecht KW, Jacobi KW, Schott K (eds) Proceedings des zweiten Kongresses der Deutschen Gesellschaft für Intraokularlinsen Implantation. Enke, Stuttgart, pp 55–62
McDonnell PJ, Green WR, Maumenee AE, Iliff WJ (1983) Pathology of intraocular lenses in 33 eyes examined postmortem. Ophthalmology 90:386–403
McDonnell PJ, Stark WJ, Green WR (1984) Posterior capsule opacification. A specular microscopic study. Ophthalmology 91:853–856
Meyer W (1897) Über den physikalisch optischen Bau der Augen von Schaf und Hund. Inaugural dissertation, Rostock
Neuhann T (1987) Theorie und Operationstechnik der Kapsulrexis. Klin Monatsbl Augenheilkd 190:542–545
Rochels R, Nover A (1988) Untersuchung zur Häufigkeit und Entstehung der Dezentrierung kapselsackfixierter Hinterkammerlinsen. Klin Monatsbl Augenheilkd 193:585–588
Skorpik C, Menapace R, Gnad HD, Grasl M, Scheidel W (1987) Evaluation of 50 silicone posterior chamber lens implantations. J Cataract Refract Surg 13:640–643
Author information
Authors and Affiliations
Additional information
Supported in part by the Claere Jung Foundation, Hamburg, FRG. Presented in part at the Third Congress of the German Intraocular Lens Implantation Society, Vienna, Austria, March 3, 1989
Rights and permissions
About this article
Cite this article
Guthoff, R., Abramo, F., Draeger, J. et al. Forces on intraocular lens haptics induced by capsular fibrosis. Graefe's Arch Clin Exp Ophthalmol 228, 363–368 (1990). https://doi.org/10.1007/BF00920063
Received:
Accepted:
Issue Date:
DOI: https://doi.org/10.1007/BF00920063