Abstract
We report a new design for an acoustically resonant spectrophone well suited for intracavity operation with widely tunable lasers. Operation with Brewster angle windows or in a windowless mode is possible, and high quality factors (560 and 509, respectively) are achieved. Windowless operation permits continuous monitoring of ambient air. We discuss factors limiting the sensitivity, ∼10−7 cm−1, and present a double chopping scheme capable of significantly reducing the acoustical background encountered in windowless operation. The sensitivity of the spectrophone's performance to misalignment is also examined.
Similar content being viewed by others
References
C.F. Dewey, R.D. Kamm, C.E. Hackett: Appl. Phys. Lett.23, 633–635 (1973)
E. Max, L.-G. Rosengren: Opt. Commun11, 422–426 (1974)
L.B. Kreuzer: “The Physics of Signal Generation and Detection”, inOptoacoustic Spectroscopy and Detection, ed. by Y.-H. Pao (Academic Press, New York 1977) pp. 1–25
Another design for a windowless resonant spectrophone, differing drastically from ours, was presented by S. Shtrikman, M. Slatkine: Appl. Phys. Lett.31, 830–831 (1977)
A preliminary report on this design was presented at the OSA. Topical Meeting on Photoacoustic Spectroscopy, Ames, Iowa (1979)
R.R. Patty, G.M. Russwurm, W.A. McClenny, D.R. Morgan: Appl. Opt.13, 2850–2854 (1974)
W. Schnell, G. Fischer: Appl. Opt.14, 2058–2059 (1975)
A. Mayer, J. Comera, H. Charpentier, C. Jaussaud: Appl. Opt.17, 391–393 (1978)
M.S. Shumate, R.T. Menzies, J.S. Margolis, L.-G. Rosengren: Appl. Opt.15, 2480–2488 (1976)