Some errors in calibrating polyhedral prisms

  • Linear and Angular Measurements
  • Published:
Measurement Techniques Aims and scope

Conclusions

  1. 1.

    The results of the calibrations can be influenced by the systematic error in the “scale factor,” which is due not only to the calibration error of the readout device of the autocollimators, but also by the extent to which identical conditions are observed in the measurements. In addition, one of the parameters of a polyhedral prism, viz. the deviations of the angles between working faces from the nominal value, also exert an indirect effect on the correctness of the calibration results.

  2. 2.

    In view of the possibility of a progressive change in the “constant” angle between the line-up axes of the autocollimators, it is appropriate to select a modified method of measurements, by rotating the polyhedral prism in one direction only, in the structuring of the physical standard arrangement. When the method outlined here is employed, the possibility of that happening is established on the basis of the di values obtained, and the systematic error can be duly determined and eliminated.

  3. 3.

    The systematic errors cited in the article are not the only ones encountered in the calibration of polyhedral prisms. In a joint resolution adopted by PKS SÉV in the field of angular measurements, the necessity of studying still other factors responsibility for errors in the measurements (effect of nonplanarity of the working surfaces accuracy in determining the plane of measurements, and so forth) is emphasized.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

Literature cited

  1. M. G. Boguslavskii, L. K. Kayak, L. I. Fedotova, and E. E. Sharova, Izmeritel'. Tekh., No. 5 (1972).

  2. Minutes of the conference of COMECON member-nation specialists on metrology, No. 59-8/267, Leningrad, September 1972.

  3. M. Peglow, Leitz Mitt. für Wiss. u. Tech., No. 7 (1964).

  4. J. G. Evans and C. O. Taylerson, Notes on Appl. Sci., No. 26 (1964).

  5. J. Hanneman and W. Würzner, Feingerätetechnik, No. 5 (1970).

  6. M. G. Boguslavskii, V. M. Éliashberg, E. E. Sharova, and L. I. Fedotova, Izmeritel'. Tekh., No. 7 (1972).

  7. I. Brezina and J. Mokros, Měrová Technika, No. 6 (1971).

  8. I. Brezina, Měrová Technika, No. 5 (1972).

  9. R. Noch, ASUG-Mitteilungen, No. 1 (1968).

Download references

Authors

Additional information

Translated from Izmeritel'naya Tekhnika, No. 5, pp. 17–18, May, 1974.

Rights and permissions

Reprints and permissions

About this article

Cite this article

Brezina, I. Some errors in calibrating polyhedral prisms. Meas Tech 17, 666–669 (1974). https://doi.org/10.1007/BF00812381

Download citation

  • Issue Date:

  • DOI: https://doi.org/10.1007/BF00812381

Keywords

Navigation