Atrial septostomy: Stationary angioplasty balloon technique-experimental work and preliminary clinical applications

  • Original Articles
  • Published:
Pediatric Cardiology Aims and scope Submit manuscript

Summary

A new approach to creating an interatrial communication using a stationary fixeddiameter angioplasty balloon is presented.

Stationary balloon septostomy was used to create an atrial communication in 19 young swine using a transseptal technique through a closed atrial septum, and in 18 newborn lambs through a patent foramen ovale.

In animals with an intact septum, there was persistent patency of the atrial communication using balloons larger than 12 mm in diameter. The average diameter of the atrial communication was 96% of the expanded diameter of the angioplasty balloon. For newborn lambs with a patent foramen, a smooth oval atrial communication was noted by 4 weeks, and there was persistent patency of the opening using balloons larger than 15 mm in diameter. The average diameter of the opening was 81% of the diameter of the angioplasty balloon used.

The new procedure was used successfully in five patients. All five showed resolution of the clinical symptomology.

In conclusion, stationary angioplasty balloon atrial septostomy has been effective in producing interatrial communications, which remained patent for 4–14 weeks, in animals with intact septum, as well as in those with a patent foramen ovale.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

References

  1. Baker F, Baker L, Zoltun R, et al. (1971) Effectiveness of the Rashkind procedure in transposition of the great arteries in infants.Circulation 43:16

    Google Scholar 

  2. King TD, Thompson SL, Mills NL (1978) Measurement of atrial septal defect during cardiac catheterization.Am J Cardiol 41:537–542

    Google Scholar 

  3. Kratz C, Davidgnon A, Chartrand C, et al. (1977) Simple dtransposition of the great arteries: results of early balloon septostomy followed by two-stage surgical correction.J Thorac Cardiovasc Surg 73:707–711

    Google Scholar 

  4. Lew AS, Harper RW, Federman J, et al. (1983) Recent experience with transseptal catheterization.Cath Cardiovasc Diagn 9:601–609

    Google Scholar 

  5. Mitchell SE, White RJ Jr, Kan J, Tolkoff J (1984) Improved balloon catheters for large-vessel and valvular angioplasty.AJR 141:571

    Google Scholar 

  6. Neches WH, Mullins CE, McNamara DG, et al. (1973) Balloon atrial septostomy in congenital heart disease in infants.Am J Dis Child 125:371–375

    Google Scholar 

  7. Park S, Zuberbuhler JR, Neches WH, et al. (1975) A new atrial septostomy technique.Cath Cardiovasc Diagn 1:195–201

    Google Scholar 

  8. Park SC, Neches WH, Zuberbuhler R, et al. (1978) Clinical use of blade atrial septostomy.Circulation 58:600–606

    Google Scholar 

  9. Park SC, Neches WH, Mullins CE, et al. (1982) Blade atrial septostomy: collaborative study.Circulation 66:258–266

    Google Scholar 

  10. Rashkind WJ, Miller WW (1966) Creation of an atrial septal defect without thoracotomy: palliative approach to complete transposition of the great arteries.JAMA 196:991

    Google Scholar 

  11. Rashkind WJ, Miller WW (1968) Transposition of the great arteries: results of palliation by balloon atrioseptostomy in thirty-one infants.Circulation 38:453–462

    Google Scholar 

  12. Rashkind WJ (1971) Atrioseptostomy by balloon catheter in congenital heart disease.Radiol Clin North Am 9:193–202

    Google Scholar 

  13. Rashkind WJ (1983) Balloon atrioseptostomy revisited: the first fifteen years.Int J Cardiol 4:369–378

    Google Scholar 

  14. Rome JJ, Keane JF, Perry SB, Spevak PJ, Lock JE (1990) Double-umbrella closure of atrial defects-initial clinical applications.Circulation 82:751–758

    Google Scholar 

  15. Sondheimer HM, Kavey REW, Blackman MS (1982) Fatal over-distension of an atrioseptostomy catheter.Pediatr Cardiol 2:255–257

    Google Scholar 

  16. Venables AW (1970) Balloon atrial septostomy in complete transposition of great arteries in infancy.Br Heart J 32:61–65

    Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Rights and permissions

Reprints and permissions

About this article

Cite this article

Mitchell, S.E., Anderson, J.H., Swindle, M.M. et al. Atrial septostomy: Stationary angioplasty balloon technique-experimental work and preliminary clinical applications. Pediatr Cardiol 15, 1–7 (1994). https://doi.org/10.1007/BF00796998

Download citation

  • Issue Date:

  • DOI: https://doi.org/10.1007/BF00796998

Key Words

Navigation