An automatically balancing isolation amplifier for the decoupling of measuring systems with high impedances

  • Instruments and Techniques
  • Published:
Pflügers Archiv Aims and scope Submit manuscript

Summary

It is theoretically shown that different conditions of coupling and grounding cause error voltages with the simultaneous use of two high-impedance measuring systems (e.g. pH electrodes). By the use of special amplifiers using optical transmission systems these error voltages are prevented. The transmission characteristics of the couplers are sensitive to changes in temperature due to the incorporation of infrared emitters in the system. An amplifying equipment with high d. c. stability was developed. The temperature drift was suppressed by the use of a special chopper stabilisation.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

Abbreviations

E E :

potential of the galvanic element glass electrode

E B :

potential of the galvanic element reference electrode

E G :

boundary liquid-wall of the reservoirs

f o :

switching frequency of the MOS FET-switches 300 Hz

j :

Gauss number (j 2=−1)

K 1 :

d. c. voltage gain of the amplifier (V 1): 36

K 2 :

d. c. voltage gain of the optical transmission system: 0.1

R b :

resistance of the reference electrode

R e :

resistance of the glass electrode

R EE :

input resistance of the amplifier (input glass electrode): 1012Ω

R EB :

(input reference electrode): 1012Ω

R 1–3,R K 1–4:

electrical resistances of the liquid between the electrodes and the wall of the reservoir

R G ,R G 1–2,T 1 :

time constant of the operational amplifier (V 1): 1.6 · 10−4 s

T 2 :

time constant of the optical transmission system: 3 · 10−6 s

T :

time constant of the hold elements: 0.1 · 10−6 s

(l):

radian frequency: s−1

u(t) :

time function of the voltage

U :

complex function of the voltage

References

  • Loeschcke, H. H., Sugioka, K.: pH of cerebrospinal fluid in the cisterna magna and on the surface of the choroid plexus of the 4th ventricle and its effect on ventilation in experimental disturbances of acid base balance. Transient and steady states. Pflügers Arch.312, 161–188 (1969)

    Google Scholar 

  • Opelt, W.: Kleines Handbuch technischer Regelvorgänge. 3. Aufl. Weinheim: Verlag Chemie 1960

    Google Scholar 

  • Süß, R.: Ein Ersatzschaltbild für die Glaselektroden-Meßkette. Arch. Techn. Messen (ATM)V 332–21, 89–92 (1960)

    Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Rights and permissions

Reprints and permissions

About this article

Cite this article

Mückenhoff, K., Luttmann, A. An automatically balancing isolation amplifier for the decoupling of measuring systems with high impedances. Pflugers Arch. 370, 93–97 (1977). https://doi.org/10.1007/BF00707952

Download citation

  • Received:

  • Issue Date:

  • DOI: https://doi.org/10.1007/BF00707952

Key words

Navigation