Der Einfluß abdominalen Unterdrucks auf das Herzminutenvolumen

Hemodynamic effects induced by abdominal decompression

  • Published:
Archiv für Gynäkologie Aims and scope Submit manuscript

Summary

  1. 1.

    A lower body negative pressure according to the method of abdominal decompression (Heyns) was applied to 17 healthy volunteers. Systolic and diastolic blood pressure, heart rate and cardiac output were measured during rest, the first, 8th and 9th decompression sessions and the Intervalls following the first and 8th application of subatmospheric pressure.

  2. 2.

    Blood pressures and heart rate were recorded with the „Haemotonograph“ (Godart) using an arm cuff and a photoelectric element shaped as an ear clip. The determination of cardiac output was performed using the dichromatic Waters ear unit XE 302 and the Waters densitometer MD-40. Cardiogreen® was used as a dye.

  3. 3.

    During abdominal decompression cardiac output decreases significantly by an average of 20 percent. The stroke volume diminishes because of the simultaneously increasing heart rate by about 30 percent.

  4. 4.

    During abdominal decompression a state similar to a hypovolemic shock with decreased venous return as well as abdominal blood perfusion and blood pooling in the splanchnic area takes place temporarily. Immediately after the application of subatmospheric pressure the blood perfusion might be temporarily raised.

Zusammenfassung

  1. 1.

    17 gesunde Versuchspersonen wurden einem abdominalen Unterdruck von −50 bis −70 mm Hg nach der Methode der abdominalen Dekompression von Heyns ausgesetzt. Systolischer und diastolischer Blutdruck, Herzfrequenz und Herzminutenvolumen wurden in Ruhe, während der ersten Dekompression und der darauffolgenden Pause sowie während der 8. und 9. Dekompression und der dazwischen liegenden Pause bestimmt.

  2. 2.

    Blutdrucke und Herzfrequenz wurden über Armmanschette und Ohr-Fotodetektor mit dem Hämotonographen gemessen. Zur Bestimmung des Herzminutenvolumens verwendeten wir die dichromatische Waters-Ohremheit und das Waters-Densitometer. Als Indikator benutzten wir Indocyanin-Grün (Cardiogreen®).

  3. 3.

    Unter abdominaler Dekompression fällt das Herzminutenvolumen statistisch signifikant um durchschnittlich 20% ab; das Schlagvolumen nimmt durch die gleichzeitig einsetzende Tachykardie um etwa 30% ab.

  4. 4.

    Während abdominaler Dekompression kommt es kurzfristig zu Symptomen eines hypovolämischen Schocks mit vermindertem venösen Rückstrom, einer Abnahme der Durchblutung und einer Blut-Pool-Bildung im Splanchnicusgebiet. Unmittelbar nach Beendigung der Unterdruckanwendung ist die Durchblutung kurzfristig erhöht.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

Literatur

  • Ardill, B. L., Bannister, R. G., Fentem, P. H., Greenfield, D. M.: Circulatory responses of supine subjects to the exposure of parts of the body below the xiphisternum to subatmospheric pressure. J. Physiol. (Lond.)193, 57 (1967).

    Google Scholar 

  • Babenerd, J.: Zentraler Venendruck während der abdominalen Dekompression bei Nichtschwangeren, Schwangeren und Gebärenden. Arch. Gynäk.207, 45 (1969).

    Google Scholar 

  • Blecher, J. A.: Toxaemia of pregnancy treated by abdominal decompression. S. Afr. med. J.39, 960 (1965).

    Google Scholar 

  • —: The volume of the maternal placental circulation before and after abdominal decompression. S. Afr. J. med. Sci.33, 43 (1968).

    Google Scholar 

  • —, Heyns, O. S.: Abdominal decompression in the treatment of the toxaemias of pregnancy. Lancet1967 II, 621.

    Google Scholar 

  • Brown, F., Goei, J. S., Greenfield, A. D. M., Plassaras, G. C.: Circulatory responses to simulated gravitational shifts of blood in man induced by exposure of the body below the iliac crests to subatmospheric pressure. J. Physiol. (Lond.)183, 607 (1966).

    Google Scholar 

  • Coles, D. R., Kidd, B. S. L., Moffat, W.: Distensibility of blood vessels of the human calf determined by local application of subatmospheric pressures. J. appl. Physiol.10, 461 (1957).

    Google Scholar 

  • Edwards, A. W. T., Bassingthwaighte, J. B., Sutter, W. F., Wood, E. H.: Blood level of indocyanine green in the dog during multiple dye curves and its effect on instrumental calibration. Mayo Clin. Proc.35, 745 (1960).

    Google Scholar 

  • Fischer, W. M., Kallista, R. W.: Herz-Kreislaufveränderungen bei der Anwendung abdominalen Unterdrucks. 1971 (im Druck).

  • —, Schlunze, K., Haan, D.: Hämodynamische Untersuchungen während abdominaler Dekompression in der Schwangerschaft — ein vorläufiger Bericht. In: E. Saling und K. A. Hüter, Fortschritte der perinatalen Medizin. Stuttgart: Georg Thieme 1970.

    Google Scholar 

  • Fox, I. J., Brooker, L. G. S., Heseltine, D. W., Essex, H. E., Wood, E. H.: A tricarbocyanine dye for continuous recording of dilution curves in whole blood independent of variations in blood oxygen saturation. Mayo Clin. Proc.32, 478 (1957).

    Google Scholar 

  • —, Wood, E. H.: Indocyanine green: Physical and physiologic properties. Mayo Clin. Proc.35, 732 (1960).

    Google Scholar 

  • Heyns, O. S.: Abdominal decompression in pregnancy and labour. Med. Proc.5, 121 (1959a).

    Google Scholar 

  • —: Abdominal decompression in the first stage of labour. J. Obstet. Gynaec. Brit. Emp.66, 220 (1959b).

    Google Scholar 

  • —: A device to improve foetal oxygenation. Trans. Coll. Phycns S. Afr.5, 11 (1961).

    Google Scholar 

  • —: Use of abdominal decompression in pregnancy and labour to improve foetal oxygenation. Develop. Med. Child Neurol.4, 473 (1962a).

    Google Scholar 

  • —: The improvement of foetal oxygenation by abdominal decompression. Bull. Soc. Gynec. Belge32, 29 (1962b).

    Google Scholar 

  • —, Samson, J. M., Graham, J. A. C.: Influence of abdominal decompression on intra-amniotic pressure and foetal oxygenation. Lancet1962 I, 289.

    Google Scholar 

  • Hunton, D. B., Bollmann, J. L., Hoffmann, H. N.: Hepatic removal of indocyanine green. Mayo Clin. Proc.35, 752 (1960).

    Google Scholar 

  • Köster, E., Dorsch, J.: Untersuchungen über Farbmeßmethoden in der Kreislaufdiagnostik. Beitr. Klin. Tuberk.134, 207 (1967).

    Google Scholar 

  • Koubenec, H.-K.: Fetale Herzfrequenz und intrauteriner Druck bei Anwendung der abdominalen Dekompression unter der Geburt. Geburtsh. u. Frauenheilk.30, 781 (1970).

    Google Scholar 

  • Lawrence, S., Lilienfield, H. H., Kovach, R. D.: Proc. Soc. exp. Biol. (N.Y.)91, 595 (1956). Zit. nach K. W. Schneider und P. Hassenstein.

    Google Scholar 

  • Master, A. M., Lasser, R. P., Jaffe, H. L.: Blood pressure in white people over 65 years of age. Ann. intern. Med.48, 284 (1958).

    Google Scholar 

  • Mathews, D. D., Loeffler, F. E.: The effect of abdominal decompression on fetal oxygenation during pregnancy and early labour. J. Obstet. Gynaec. Brit. Cwlth75, 268 (1968).

    Google Scholar 

  • McGregor, M., Sekelj, P., Adam, W.: Measurement of cardiac output in man by dye dilution curves using simultaneous ear oxymeter and whole blood cuvette techniques. Circulat. Res.9, 1083 (1961).

    Google Scholar 

  • Milnor, W. R., Talbot, S. A., McKeever, W. P., Marye, R. B., Newman, E. V.: A photoelectric ear densitometer for continuously recording the arterial concentration of T-1824 in the dye dilution method. Circulat. Res.1, 117 (1953).

    Google Scholar 

  • Murray, R. H., Thompson, L. J., Bowers, J. A., Albright, C. D.: Hemodynamic effects of graded hypovolemia and vasodepressor syncope induced by lower body negative pressure. Amer. Heart. J.76, 799 (1968).

    Google Scholar 

  • Musgrave, S.: Thesis, University of Kentucky, 1967: zit. nach Murray, R. H. u. Mitarb.

  • Rueger, M. J.: Blood pressure variations in the two arms. Ann. intern. Med.35, 1023 (1951).

    Google Scholar 

  • Schneider, K. W.: Blutvolumen und Herzminutenvolumen. Ärztl. Forsch.20, 420 (1966).

    Google Scholar 

  • —, Cheikhzadeh, A., Müller, W.: Die Bestimmung des Herzminutenvolumens mit einem dichromatischen Ohrdensitometer und vereinfachter Berechnung durch Computer. Cardiologia (Basel)51, 349 (1967).

    Google Scholar 

  • —, Hassenstein, P.: Probleme der Herzminutenvolumen-Bestimmung bei Verwendung von Indocyanin-Indikatoren. Arch. Kreisl.-Forsch.42, 84 (1963).

    Google Scholar 

  • Schubiger, V., Hauser, G. A.: Abdominelle Dekompression. Gynaecologia (Basel)167, 293 (1969).

    Google Scholar 

  • Sekelj, P., Jegier, W., Johnson, A.: Automatic electronic computer for the estimation of arterial concentration of Evans blue dye. Amer. Heart. J.55, 485 (1958).

    Google Scholar 

  • Takaki, N. K., Mathews, C. W., Semmens, J. P.: Indocyanine green liver function studies on women taking Progestins. Obstet. Gynec.30, 220 (1967).

    Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Rights and permissions

Reprints and permissions

About this article

Cite this article

Fischer, W.M., Kallista, R.W. Der Einfluß abdominalen Unterdrucks auf das Herzminutenvolumen. Arch. Gynäk. 210, 111–123 (1971). https://doi.org/10.1007/BF00667802

Download citation

  • Received:

  • Issue Date:

  • DOI: https://doi.org/10.1007/BF00667802

Navigation