Summary
Clinical experience with 130 endoprostheses, entirely made of ceramic, having an average 5 year follow-up is reported. Excellent to sufficient results were seen in 84% of all cases. A number of particular matters of interest (operative technique, radiological morphology of the bone/ ceramic bone interface etc.) are discussed in detail. Complications (loosening of components: 2%; component fracture: 6%, wear phenomena etc.) are described and documented including histological observations from capsular tissue of total ceramic hip joints. The whole system has basically stood the test, during the implantation period, modifications to improve the function of the system will be described in another publication.
Zusammenfassung
Es wird über 5jährige klinische Erfahrungen an Hand von 130 Keramik-Keramikhüft-endoprothesen berichtet. Die sehr guten bis befriedigenden Ergebnisse machen 84% aus. Über eine Reihe von Besonderheiten (Operationstechnik, Röntgenmorphologie der Grenzfläche um Keramikimplantate etc.) wird detailliert berichtet. Es werden verschiedene Komplikationen (Lockerungen: 2%, Keramikbrüche: 6%, Abriebphänomene) dargestellt, ramikbrüche: 6%, Abriebphänomene) dargestellt, wobei auch humanhistologische Befunde aus Kapselgewebe um Keramik-Keramikgelenke beschrieben werden. Das System hat sich in den Grundzügen bewährt, Änderungen, die die Prothese quo ad functionem sicherer machen, werden in einer 2. Publikation vorgestellt.
Similar content being viewed by others
References
Boutin P (1972) Arthroplastie totale de la hanche par prothese en Alumine fritte. Etude experimentale et premieres applications cliniques. Rev Chir Orthop 58:229–246
Boutin P, Blanquaert D, Sedel L (1980) Friction materials in total hip replacement: Characterisation of Aluminium oxide after implantation for 9 years. Abstract, 1st World Biomaterials Congress, Vienna/Austria, April 8–12
Buchhorn U, Griss P, Niederer PG, Willert H-G (1979) Klinische Relevanz von Lockerungszeichen bei Hüftendoprothesen. Z Orthop 117:685–690
Christel P, Leray JL (1980) Tissue tolerance of wear debris from arthroplasties. Abstract, 1st World Biomaterials Congress, Vienna/Austria, April 8–12
Doerre E, Dawihl W, Altmeyer G (1977) Dauerfestigkeit keramischer Hüftendoprothesen. Biomed Technik 22:3–7
Doerre E, Dawihl W (1978) Mechanische und tribologische Eigenschaften keramischer Endoprothesen. Biomed Technik 23:305–310
Griss P, Krempien B, Andrian-Werburg H Frhr v, Heimke G, Fleiner R (1973) Experimentelle Untersuchung zur Gewebsverträglichkeit oxidkeramischer (Al2O3) Abriebteilchen. Arch Orthop Unfall-Chir 76:270–279
Griss P, Heimke G, Andrian-Werburg H Frhr v (1975) Die Aluminiumoxidkeramik-Metall-Verbundendoprothese. Fine neue Hüftgelenktotalendoprothese zur teilweise zementfreien Implantation. Arch Orthop Unfall-Chir 81:259–266
Griss P, Werner E, Buchinger R, Heimke G (1977) Die Mannheimer Oxidkeramik-Metall-Verbundendoprothesen. Eine kritische Wertung unserer bisherigen klinischen Erfahrungen. Arch Orthop Unfall-Chir 87:73–84
Griss P, Heimke G, Werner E, Bleicher J, Jentschura G (1978a) Was bedeutet die Resorption des Calcar femoris nach der Totalendoprothesenoperation der Hüfte? Eine vergleichende Studie an Charnley-Müller-und Oxidkeramikendoprothesen (Typ Lindenhof). Arch Orthop Traumat Surg 92:225–232
Griss P, Jentschura G, Heimke G (1978b) Zur Technik der Pfannenimplantation bei dysplastischem Acetabulum. Arch Orthop Traumat Surg 93:57–63
Griss P, Hackenbroch MH, Jäger M, Preussner B, Schäfer Th, Seebauer R, Eimeren W van, Winkler W (1981) Ten year results of total hip replacement. A retrospective multicenter study based on a 10% random sample of 39,000 total hip replacements. In: Burri C, Herfarth D, Jager M (eds) Aktuelle Probleme in Chirurgie und Orthopädie. H Huber, Bern (in press)
Harms J, Mäusle E (1980) Biokompatibilität von Implantaten in der Orthopädie. Hefte Unfallheilkunde. Springer, Berlin Heidelberg New York
Heimke G, Beisler W, Andrian-Werburg H Frhr v, Griss P, Krempien B (1973) Untersuchungen an Implantaten aus Al2O3-Keramik. Ber Dtsch Keram Ges 50:4–8
Heimke G, Griss P (1981) Five years experience with ceramicmetal-composite hip endoprostheses. II. Arch Orthop Traumat Surg 98:165–171
Knahr K, Plenk H, Salzer M (1980) Erfahrungen mit biokeramischen Implantaten in der Orthopädie. Vortrag: Biomaterialien-Symposium Gießen, 14. Juni
Mittelmeier H, Harms J (1980) Implantate aus Aluminium-oxydkeramik: Vortrag, Biomaterialien-Symposium Gießen, 14. Juni
Plitz W (1980) Clinical and tribological aspects of the damage analysis of AlO 3 ceramic hip joints. Abstract, Ist World Biomaterials Congress, Vienna/Austria, April 8–12
Plitz W, Griss P (1980) Clinical, histomorphological and material related observations on removed alumina-ceramic hip joint components. Proceedings: Conference on Implant Retrieval: Material and biological analysis, Gaithersburg, Ma, May 1–2 (in press)
Salzer M, Locke H, Plenk H, Punzet G, Zweymüller K (1977) Bioceramic components in hip joint endoprostheses experiences in man. 9th Ann Int Biomaterials Symp. New Orleans, April 1977, Abstract
Semlitsch M, Willert H-G, Doerre E (1975) Neue Werkstoffpaarung Al2O3-Keramik, Polyaethylen zur Verminderung des Polyaethylenabriebes bei Gelenkpfannen von Hüftendoprothesen. Med Orthop Technik (MOT) 95:143–144
Semlitsch M, Lehmann M, Weber H, Doerre E, Willert H-G (1977) New prospects for a prolonged functional life span of artificial hip joints by using the material combination polyethylene/aluminium oxide ceramic/metal. J Biomed Mater Res 10:537–552
Willert H-G, Semlitsch M, Buchhorn G, Kriete U (1978) Materialverschleiß und Gewebereaktion bei künstlichen Gelenken (Histopathologie, Biokompatibilität, biologische und klinische Probleme). Orthopäde 7:62–83
Author information
Authors and Affiliations
Additional information
This work was partially sponsored by the German Federal Minister of Research and Technology (Grant No. 01 V6 256 A-ZK/NT/MT 267)
Rights and permissions
About this article
Cite this article
Griss, P., Heimke, G. Five years experience with ceramic-metal-composite hip endoprostheses. Arch. Orth. Traum. Surg. 98, 157–164 (1981). https://doi.org/10.1007/BF00632972
Received:
Issue Date:
DOI: https://doi.org/10.1007/BF00632972