On the accurate calibration of an electrochemical oxygen meter in the 10−7 bar range

  • C. M. Mari
  • S. Pizzini
  • T. A. Giorgi
  • L. Rosai
  • M. Borghi
Article

Abstract

A method for the accurate standardization of an electrochemical oxygen meter, in the 10−6-10−7 bar range is described. The method is based on the use of an electrochemical oxygen leak and a chemical getter pump. In this manner problems connected with traces of hydrogen, originating from the pyrolysis of water, and traces of oxygen of unknown origin are overcome.

Keywords

Oxygen Hydrogen Physical Chemistry Pyrolysis Unknown Origin 
These keywords were added by machine and not by the authors. This process is experimental and the keywords may be updated as the learning algorithm improves.

Preview

Unable to display preview. Download preview PDF.

Unable to display preview. Download preview PDF.

References

  1. [1]
    T. H. Etsell and S. N. Flengas,Met. Trans,3 (1972) 27.Google Scholar
  2. [2]
    C. B. Alcock and S. Zador,J. Appl. Electrochem. 2 (1972) 289.Google Scholar
  3. [3]
    J. Fouletier, G. Vitter and M. Kleitz,ibid 5 (1975) 111.Google Scholar
  4. [4]
    L. Rosai and M. Borghi,J. Vac. Sci. Technol. 11 (1974) 347.Google Scholar
  5. [5]
    S. Pizzini and G. Bianchi,Chim. Ind. 55 (12) (1973) 966.Google Scholar
  6. [6]
    S. Pizzini, C. M. Mari and D. C. Hadjicostants, Proc. 8th I.S.R.S. Gothenburg (Sweden) (1976) p. 104.Google Scholar
  7. [7]
    J. Fouletier, H. Seinera and M. Kleitz,J. Appl. Electrochem. 4 (1974) 305.Google Scholar

Copyright information

© Chapman and Hall Ltd. 1977

Authors and Affiliations

  • C. M. Mari
    • 1
  • S. Pizzini
    • 1
  • T. A. Giorgi
    • 2
  • L. Rosai
    • 2
  • M. Borghi
    • 2
  1. 1.Electrochemistry InstituteUniversity of MilanMilanoItaly
  2. 2.S.A.E.S. Getters S.p.A.MilanoItaly

Personalised recommendations