Skip to main content

Determination of membrane potential in smooth muscle cells using microelectrodes with reduced tip potential

Summary

Conventionally 3 M KCl-filled, high resistance glass microelectrodes introduce large errors in the measurement of d.c. transmembrane potentials (MP) because of normally occurring tip potentials (TP). If, however, the 3 M KCl filling solution is adjusted to pH 2.0 with HCl, electrodes loose, almost entirely, the TP. Furthermore, the sensitivity of TP to the ionic composition of the test medium (0.125 M KCl and 0.125 M NaCl solution) is lost. Conventionally 3 M KCl-filled electrodes (pH 6.5) are found to have TP exceeding −30 mV in 0.125 M KCl solution. TP values measured in 0.125 M NaCl solution are about 29% higher than those measured in 0.125 M KCl solution. The MP of smooth muscle cells (isolated taenia coli of guinea-pig in modified Krebs solution at 32°C) determined using acidified electrodes shows an arithmetic mean of −51.2 mV with a S.D. of only 1.8 mV. However, MP determined by conventionally filled electrodes appears to be smaller by a variable amount. This reduction of MP is found to be linearly correlated over a wide range with the TP difference (Δ TP), as measured in the two test solutions. When comparing the MP determined by the two kinds of electrodes there is a good correlation provided that Δ TP is comparable. The results indicate that the reduction of MP determined by conventional electrodes is induced by their different TP in solutions of varying ionic composition. As a result, avoiding large TP by acidifying the filling solution can help to reduce TP-induced errors in MP measurements.

This is a preview of subscription content, access via your institution.

References

  • Adrian, R. H.: The effect of internal and external potassium concentration on the membrane potential of frog muscle. J. Physiol. (Lond.)133, 631–658 (1956)

    Google Scholar 

  • Golenhofen, K., v. Loh, D.: Elektrophysiologische Untersuchungen zur normalen Spontanaktivität der isolierten Taenia coli des Meerschweinchens. Pflügers Arch.314, 312–328 (1970)

    Google Scholar 

  • Küchler, G.: Zur Frage der Übertragungseigenschaften von Glasmikroelektroden bei der intracellulären Membranpotentialmessung. Pflügers Arch. ges. Physiol.280, 210–223 (1964)

    Google Scholar 

  • Wann, K. T., Goldsmith, M. W.: Reduction of tip potential artefacts in microelectrode measurements. Nat. physic. Sci.238, No. 81, 44–45 (1972)

    Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Additional information

Supported by the Deutsche Forschungsgemeinschaft (Ul 8/4).

Rights and permissions

Reprints and Permissions

About this article

Cite this article

Riemer, J., Mayer, CJ. & Ulbrecht, G. Determination of membrane potential in smooth muscle cells using microelectrodes with reduced tip potential. Pflugers Arch. 349, 267–275 (1974). https://doi.org/10.1007/BF00592454

Download citation

  • Received:

  • Issue Date:

  • DOI: https://doi.org/10.1007/BF00592454

Key words

  • Smooth Muscle
  • Microelectrodes
  • Membrane Potential