Summary

Correlation has previously been observed between the incidence of metal hypersensitivity (Ni, Cr, Co) and the incidence of infection after internal fixation of bone fractures. As there is no doubt that exposure to high concentrations of metal can cause allergy, metal levels (Cr, Ni, Mo, Co) were analyzed in implant contact tissues by ASS. Both the absolute concentration of the ions and the Ni: Cr ratio in the tissues adjacent to metal implants is greater in infected cases than in non-infected. Thus infection as a cause of allergy has to be considered. In view of the resulting problems perhaps consideration could be given to the use of a different alloy for metal implants.

Zusammenfassung

In vorangegangenen Untersuchungen konnte ein Zusammenhang zwischen Allergie gegen implantatspezifische Metalle (Ni, Cr, Co) und Infektmorbidität bei Patienten mit Osteosynthesen nach Frakturen festgestellt werden. Da kein Zweifel daran besteht, daß eine Exposition gegen hohe Metallkonzentrationen eine Allergie verursachen kann, wurden im Implantatkontaktgewebe Metallkonzentrationen (Cr, Ni, Mo, Co) mit der Atomabsorptionsspektroskopie bestimmt. Dabei ergab sich, daß sowohl die Nickel- und Chromkonzentrationen wie auch die Ni: Cr-Relation im Implantatkontaktgewebe infizierter Osteosynthesen größer ist als bei nicht infizierten. Es kann daher angenommen werden, daß die Infektion über die Erhöhung der Metallgewebekonzentrationen eine erhöhte Allergierate hervorruft. Angesichts der daraus folgenden Probleme stellt sich daher die Frage nach der Verwendung von Stählen mit verbesserter Korrosionsbeständigkeit auch in der Chirurgie.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

References

  1. Jones DA, Lucas KH, O'Driscoll M, Price CHG, Wibberley B (1975) Cobalt toxicity after McKee hip arthroplasty. J Bone Joint Surg [Br] 57:289–296

    Google Scholar 

  2. Hierholzer S, Hierholzer G (in press) Metallallurgie als pathogenetischer Faktor für die Knocheninfektion nach Osteosynthesen. Unfallheilkunde

  3. Hofmann J, Michel R, Holm R, Zilkens J (1981) Corrosion behaviour of stainless steel implants in biological media. Surface and interface analysis, vol 3. Heyden and Son, London, pp 110–117

    Google Scholar 

  4. Zitter H, Schaschl-Outschar D (1981) Schadensfälle an chirurgischen Implantaten und deren Ursachen. Werkstoffe und Korrosion 32:324–331

    Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Rights and permissions

Reprints and permissions

About this article

Cite this article

Hierholzer, S., Hierholzer, G., Sauer, K.H. et al. Increased corrosion of stainless steel implants in infected plated fractures. Arch. Orth. Traum. Surg. 102, 198–200 (1984). https://doi.org/10.1007/BF00575233

Download citation

  • Received:

  • Issue Date:

  • DOI: https://doi.org/10.1007/BF00575233

Keywords

Navigation