Choice of probe current parameters in channel time-division rheographic equipment

  • Research, Design, And Technology
  • Published:
Biomedical Engineering Aims and scope

Conclusions

1. Channel time division has been used in rheographic apparatus, where interaction between the pulsed probe currents and the electrode-object-electrode system may increase the error.

2. Such interactions can largely be eliminated if the probe-current spectrum coincides with the frequency range from 30–50 up to 300–400 kHz, in which the system is basically of resistive behavior.

3. A probe current composed of a radio-pulse train is optimum for rheographic purposes with time division, where the mark-space ratio is equal to the number of channels in the apparatus, the repetition frequency is a few hundred Hz, and the carrier frequency lies in the range 50–200 kHz.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

Literature cited

  1. L. M. Anan'ev, Ya. S. Pekker, and M. B. Psakhis, Med. Tekhnika, No. 5, 17–21 (1984).

  2. A. I. Naumenko and V. V. Skotnikov, The Principles of Electroplethysmography [in Russian], Leningrad (1975).

  3. V. I. Polishchuk and L. G. Terekhova, Techniques and Methods in Rheography and Rheoplethysmography [in Russian], Moscow (1983).

  4. G. B. Psakhis and M. B. Psakhis, Med. Ref. Zh.,22, No. 10, 32 (1984).

    Google Scholar 

  5. R. Kh. Tukshaitov and G. P. Novoshinov, Paraclinical Examination Methods in Neurological Clinical Practice [in Russian], Issue 2, Moscow (1969), pp. 51–59.

  6. Kh. Kh. Yarullin, Clinical Rheoencephalography [in Russian], Moscow (1983).

Download references

Authors

Additional information

Krasnoyarsk Polytechnical Insittute. Translated from Meditsinskaya Tekhnima, No. 2, pp. 31–36, March–April, 1987.

Rights and permissions

Reprints and permissions

About this article

Cite this article

Pekker, Y.S., Psakhis, M.B. Choice of probe current parameters in channel time-division rheographic equipment. Biomed Eng 21, 69–72 (1987). https://doi.org/10.1007/BF00563029

Download citation

  • Received:

  • Issue Date:

  • DOI: https://doi.org/10.1007/BF00563029

Keywords

Navigation