Abstract
An interlaboratory testing program on the measurement of the linear thermal expansion coefficient of low-expansion glasses has been carried out. Three different types of interferometric dilatometers, each located at three different organizations, and two kinds of low-expansion glass materials were selected for the experiments. As a result of the comparison, a reasonable agreement among the different measuring instruments was confirmed, and it was determined that the thermal expansion coefficient for low-expansion glasses can be measured with an accuracy of ±4×10−8 K−1 by using commercially available interferometric dilatometers.
Similar content being viewed by others
References
S. J. Bennett, J. Phys. E 10:525 (1977).
R. B. Roberts, J. Phys. E 14:1386 (1981).
W. D. Drotning, Rev. Sci. Instrum. 52:1896 (1981).
A. P. Miiller and A. Cezairliyan, Int. J. Thermophys. 3:259 (1982).
M. Okaji and H. Imai, J. Phys. E 16:1208 (1983).
S. Nagai, M. Okaji, and H. Imai, Bull. Natl. Res. Lab. Metrol. 37:388 (1988) (in Japanese).
M. Okaji and H. Imai, J. Phys. E 17:669 (1984).
K. Takahashi et al., Thermophys. Prop. 8(8th Jap. Symp.): 179 (1987) (in Japanese).
H. Aikawa, M. Okaji, and H. Imai, In preparation.
J. W. Berthold III, S. F. Jacobs, and M. A. Norton, Metrologia 13:9 (1977).
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Imai, H., Okaji, M., Kishii, T. et al. Measurement of thermal expansivity of low-expansion glasses by interferometric methods: Results of an interlaboratory comparison. Int J Thermophys 11, 937–947 (1990). https://doi.org/10.1007/BF00503585
Received:
Issue Date:
DOI: https://doi.org/10.1007/BF00503585