Summary
Enzyme electrodes were assembled by coupling membrane-immobilized lactate monooxygenase and coimmobilized lactate monooxygenase and lactate dehydrogenase, respectively, to oxygen electrodes. The lactate monooxygenase sensor was used for lactate determination with the analyzer “Glukometer” as well as for sequential measurement of lactate and lactate dehydrogenase activity.
The bienzyme electrode is applicable for determining lactate and pyruvate in one measuring cycle. The operational parameters of the optimized sensors like measuring range, relative standard deviation, measuring frequency, and stability were studied. Determinations in biological fluids correlated well with spectrophotometric reference methods.
Similar content being viewed by others
Literatur
Scheller FW, Schubert F, Renneberg R, Müller H-G, Jänchen M, Weise H (1985) Biosensors 1:135–160
Schmidt H-L, Kittsteiner-Eberle R (1986) Naturwissenschaften 73:314–321
Blaedel WJ, Jenkins RA (1976) Anal Chem 48:1240–1247
Schelter-Graf A, Schmidt H-L, Huck H (1984) Anal Chim Acta 163:299–303
Turner APF (1986) Ciba Foundation Meeting “Biosensors”, London
Guarnaccia R, Racine Ph (1976) Biomed Techn 21:189–190
Clark LC, Noyes LK, Grooms TA, Gleason CA (1984) Clin Biochem 17:288–291
Mizutani F, Sasaki K, Shimura Y (1982) Anal Chem 55:35–38
Schindler JG, v Gülich M (1981) Fresenius Z Anal Chem 308:434–436
Empfehlungen der Deutschen Gesellschaft für Klinische Chemie (1972) Z Klin Chem Klin Biochem 10:182–192
Blaedel WJ, Engstrom RC (1980) Anal Chem 52:1691–1697
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Weigelt, D., Schubert, F. & Scheller, F. Enzymelektroden Zur Bestimmung von Lactat, Pyruvat und der Aktivität von Lactatdehydrogenase. Z. Anal. Chem. 328, 259–262 (1987). https://doi.org/10.1007/BF00481633