Ein gesteuerter dynamischer Titrierautomat mit konduktometrischer Indizierung

A controlled dynamic titrator with conductometric indication

  • Originalabhandlungen
  • Published:
Fresenius' Zeitschrift für analytische Chemie Aims and scope Submit manuscript

Abstract

The reagent solution (concentration c) is fed with constant flow rate w to a mixing chamber to which at the same time the sample solution (sample concentration x) is fed with a linearly increasing flow rate v=kt. After passing through a heat exchanger, the thermostated reaction mixture reaches a conductivity flow cell. The recorded conductivity curves are identical with the analoguous diagrams obtained by conductometric batch titrations. Thirty titrations per hour are performed automatically. The standard deviation is ±0.3 %. The sample amount does not have to be measured.

Zusammenfassung

Der beschriebene Titrierautomat basiert auf folgendem Prinzip: Die Maßlösung (Titratorkonzentration c) wird mit der konstanten Geschwindigkeit w einer Mischkammer zugeführt, in die gleichzeitig die Probelösung (Titrandenkonzentration x) mit der linear ansteigenden Geschwindigkeit v=kt strömt. Nach Passieren eines Wärmeaustauschers gelangt das thermostatisierte Reaktionsgemisch in eine Durchflußleitfähigkeitsmeßzelle. Die von einem Schreiber registrierten Leitfähigkeitsdiagramme sind identisch mit den analogen Kurven bei der statischen, konduktometrischen Titration. In einer Stunde werden 30 Titrationen vollautomatisch durchgeführt. Die Standardabweichung liegt bei ±0,3%. Die Probelösung (maximal 1,2ml) erfordert keine Dosierung.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

Literatur

  1. Ashworth, M. R. F.: Titrimetric Organic Analysis, I. New York: Intersci. Publ. 1964

    Google Scholar 

  2. Autorenkollektiv: Analytikum. Leipzig: VEB Deutscher Verlag für Grundstoffindustrie 1971

    Google Scholar 

  3. Berry, E., Barter, D. (Beckman): US-Patent 3 574 554 (1971)

  4. Boardman, W.: Chem. Ind. (London) 14, 565 (1963)

    Google Scholar 

  5. Dubois, J. E., Walisch, W.: J. Chim. Phys. 52, 775 (1955)

    Google Scholar 

  6. Fisher Scientific Co.: Brit. Patent 95 86 92 (1964)

  7. Fleet, B., Ho, A. Y. W.: Anal. Chem. 46, 9 (1974)

    Google Scholar 

  8. Foissac, L.: Chim. Anal. (Paris) 48, 354 (1966)

    Google Scholar 

  9. Halfter, G. (Geigy AG): Schweiz. Patent 3 63 508 (1962)

  10. Humme, G., Jaenicke, O., Walisch, W.: Meßtechnik 77, 225 (1969)

    Google Scholar 

  11. Irving, H., Pettit, L. D.: Analyst 84, 641 (1959)

    Google Scholar 

  12. Mueller, T. R., Stelzner, R. W., Fisher, D. J., Jones, H. C.: Anal. Chem. 37, 13 (1965)

    Google Scholar 

  13. Oehme, F.: Fortschr. Chem. Forsch. 29, 73 (1972)

    Google Scholar 

  14. Pehrsson, L., Johansson, A.: Anal. Chim. Acta 66, 67 (1973)

    Google Scholar 

  15. Schwartz, M. K.: Anal. Chem. 45, 739A (1973)

    Google Scholar 

  16. Skeggs, L. T.: Am. J. Clin. Pathol. 28, 311 (1957)

    Google Scholar 

  17. Taylor, A. F.: Analyst 88, 145 (1963)

    Google Scholar 

  18. Walisch, W.: Ann. Univ. Sarav. (Sci.) 2, 270 (1953)

    Google Scholar 

  19. Walisch, W.: Chim. Anal. (Paris) 39, 63 (1957)

    Google Scholar 

  20. Walisch, W.: Dtsch. Bundespatent 1 051 386

  21. Walisch, W., Barthel, J.: Dtsch. Bundespatent 1 237 359 (1962)

  22. Walisch, W., Barthel, J.: Z. Phys. Chem. N. F. 39, 236 (1963)

    Google Scholar 

  23. Walisch, W., Becker, F.: Z. Phys. Chem. N. F. 34, 319 (1962)

    Google Scholar 

  24. Walisch, W., Humme, G.: Mikrochim. Acta 1968, 748

  25. Walisch, W., Humme, G., Jaenicke, O.: Meßtechnik 77, 149 (1969)

    Google Scholar 

  26. Walisch, W., Jaenicke, O.: Mikrochim. Acta 1967, 1147

Download references

Author information

Authors and Affiliations

Authors

Additional information

Diese Arbeit wurde als Projekt des Sonderforschungsbereichs Analytik durchgeführt. Wir danken der Deutschen Forschungsgemeinschaft für die Bereitstellung der Mittel.

Rights and permissions

Reprints and permissions

About this article

Cite this article

Ashworth, M.R.F., Walisch, W., Becker, W. et al. Ein gesteuerter dynamischer Titrierautomat mit konduktometrischer Indizierung. Z. Anal. Chem. 273, 275–281 (1975). https://doi.org/10.1007/BF00459835

Download citation

  • Received:

  • Issue Date:

  • DOI: https://doi.org/10.1007/BF00459835

Navigation