Osteotomies of the distal femur were fixed with two self-reinforced poly-L-lactic acid (SR-PLLA) plates and metallic screws placed through the plates on each side of the femur in 23 adult rabbits. They were followed-up after 3, 6, 12 and 24 weeks. After killing, radiological, histological, microradiographic and oxytetracycline fluorescence studies were performed. Except for one histologically confirmed fibrotic non-union at 24 weeks, the osteotomies healed, including one involving a rabbit which had suffered an ipsilateral femoral shaft fracture of unknown cause. No malformations were observed, and the macroscopically detected swelling was a normal postoperative reaction. This study showed that SR-PLLA plates implanted on both sides on the bone are suitable for the fixation of weight-bearing cancellous bone osteotomies in rabbits.
Similar content being viewed by others
References
Bergsma EJ, Rozema FR, Bos RRM, Bruijn WC de (1993) Foreign body reactions to resorbable poly(L-lactide) bone plates and screws used for the fixation of unstable zygomatic fractures. J Oral Maxillofac Surg 51:666–670
Bos RRM, Rozema FR, Boering G, Nijenhuis AJ, Pennings AJ, Verwey AB (1989) Bio-absorbable plates and screws for internal fixation of mandibular fractures. A study in six dogs. Int J Oral Maxillofac 18:365–369
Bos RRM, Rozema FR, Boering G, Nijenhuis AJ, Pennings AJ, Jansen HWB (1989) Bone-plates and screws of bioabsorbable poly (L-lactide) — an animal pilot study. Br J Oral Maxillofac Surg 27:467–476
Bos RRM, Rozema RF, Boering G, Nijenhuis AJ, Pennings AJ, Verwey AB, Nieuwenhuis P, Jansen HWB (1991) Degradation of and tissue reaction to biodegradable poly(L-lactide) for use as internal fixation of fractures: a study in rats. Biomaterials 12:32–36
Buchholz RW, Henry S, Henley MB (1994) Fixation with bioabsorbable screws for the treatment of fractures of the ankle. J Bone Joint Surg [Am] 76:319–324
Champy M, Blez P, Kohn JL, Muster D (1993) Utilisation de plaques résorbables dans Postéosynthese du tiers moyen de la face. In: Mainard D, Merle M, Delagoutte JP, Louis JP (eds) Actualité en biomatériaux, Vol 2. Romillat, Paris, pp 47–54
Eitenmüller J, Entenmann J, Muhr G (1988) Treatment of ankle fractures with complete biodegradable plates and screws of high molecular weight polylactide. In: Transactions of the 3rd World Biomaterials Congress, Kyoto, Japan, April 21–25, p 195
Goldner JA (1938) A modification of Masson trichrome technique for routine laboratory purposes. Am J Pathol 14:237–243
Hanafusa S, Matsusue Y, Yasunaga T, Yamamuro T, Oka M, Shikinami Y, Ikada Y (1995) Biodegradable plate fixation of rabbit femoral shaft osteotomies. Clin Orthop 315:262–271
Hara Y, Tagawa M, Eijma H, Orima H, Sugiyama M, Shikinami Y, Hyon S-H, Ikada Y (1994) Clinical evaluation of uniaxially oriented poly-L-lactide rod for fixation of experimental femoral diaphyseal fracture in immature cats. J Vet Med Sci 56:1041–1045
Hirvensalo E (1989) Fracture fixation with biodegradable rods. Forty-one cases of severe ankle fractures. Acta Orthop Scand 60:601–606
Hofmann GO (1995) Biodegradable implants in traumatology: a review on the state-of-art. Arch Orthop Trauma Surg 114:123–132
Koskikare K, Hirvensalo E, Pätiälä H, Rokkanen P, Pohjonen T, Törmälä P, Lob G (1995) Intra-osseous plating with absorbable self-reinforced poly-L-lactide plates in the fixation of distal femoral osteotomies on rabbits. J Biomed Mater Res
Majola A, Vainionpää S, Vihtonen K, Vasenius J, Törmälä P, Rokkanen P (1992) Intramedullary fixation of cortical bone osteotomies with self-reinforced polylactic rods in rabbits. Int Orthop 16:101–108
Manninen MJ, Päivärinta U, Pätiälä H, Rokkanen P, Taurio P, Tamminmäki M, Törmälä P (1992) Shear strength of cancellous bone after osteotomy fixed with absorbable self-reinforced polyglycolic acid and poly-L-lactic acid rods. J Mater Sci Mater Med 3:245–251
Päivärinta U, Böstman O, Majola A, Toivanen T, Törmälä P, Rokkanen P (1993) Intraosseous cellular response to biodegradable fracture fixation screws made of polyglycolide or polylactide. Arch Orthop Trauma Surg 112: 71–74
Partio EK (1992) Absorbable screws in the fixation of cancellous bone fractures and arthrodeses. A clinical study of 318 patients. Thesis, University of Helsinki
Rozema FR (1991) Resorbable poly (L-lactide) bone plates and screws. Thesis, University of Groningen
Ruf W, Schult W, Buhl K (1990) Die Stabilisierung von Malleolarfrakturen und Flakeverletzungen mit resorbierbaren Polyglykolid-Stiften (Biofix). Unfallchirurgie 16:202–209
Suuronen R, Pohjonen T, Vasnius J, Vainionpää S (1992) Comparison of absorbable self-reinforced multilayer poly-Llactide and metallic plates in the fixation of mandibular body osteotomies. An experimental study in sheep. J Oral Maxillofac Surg 50: 255–262
Suuronen R, Pohjonen T, Wessman L, Törmälä P, Vainionpää S (1992) New generation biodegradable plate for fracture fixation. Comparison of bending strengths of mandibular osteotomies fixed with absorbable self-reinforced multi-layer polyL-lactide plates and metallic plates. An experimental study in sheep. Clin Mater 9:77–84
Svensson P-J, Janarv P-M, Hirsch G (1994) Internal fixation with biodegradable rods in pediatric fractures: one-year followup of fifty patients. J Pediatr Orthop 14:220–224
Törmälä P, Rokkanen P, Laiho J, Tamminmäki M, Vainionpää S (1988) Material for osteosynthesis devices. US Patent No 4 743 257
Vert M, Chabot F, Leray J, Christel P (1981) Stereoregular bioresorbable polyesters for orthopaedic surgery. Macromol Chem Supp 15:30–41
Yamamuro T, Matsusue Y, Uchida A, Shimada K, Shimozaki E, Kitaoka K (1994) Bioabsorbable osteosynthetic implants of ultra high strength poly-l-lactide. Int Orthop 18:332–340
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Koskikare, K., Hirvensalo, E., Pätiälä, H. et al. Fixation of osteotomies of the distal femur with absorbable, self-reinforced, poly-L-lactide plates. Arch Orthop Trauma Surg 116, 352–356 (1997). https://doi.org/10.1007/BF00433989
Received:
Issue Date:
DOI: https://doi.org/10.1007/BF00433989