Abstract
We report two cases in which fracture of the tibial metal tray was thought to be due to improper design of the tray and heavy polyethylene wear. The bone beneath the fractured portion of the tray was deficient and had been replaced by granulomatous fibrous tissue including numerous polyethylene wear particles and foreign-body giant cells. Osteolysis occurred as a tissue reaction to these particles.
Similar content being viewed by others
References
Dannenmaier WC, Haynes DW, Nelson CL (1985) Granulomatous reaction and cystic bony destruction associated with high wear rate in a total knee prosthesis. Clin Orthop 198:224–230
Flivik B, Ljung P, Rydholm U (1990) Fracture of the tibial tray of the PCA knee. A case report of early failure caused by improper design. Acta Orthop Scand 61:26–28
Gradisar IA Jr, Hoffmann ML, Askew MJ (1989) Fracture of a fenestrated metal backing of a tibial knee component. A case report. J Arthroplasty 4:27–30
Harrington IJ (1983) Static and dynamic loading patterns in knee joints with deformities. J Bone Joint Surg [Am] 65:247–259
Mendes DG, Brandon D, Galor L, Roffman M (1984) Breakage of the metal tray in total knee replacement. Orthopaedics 7:860–862
Nolan JF, Bucknill TM (1992) Aggressive granulomatosis from polyethylene failure in uncemented knee replacement. J Bone Joint Surg [Br] 74:23–24
Pilliar RM (1983) Powder metal-made orthopedic implants with porous surface for fixation by tissue ingrowth. Clin Orthop 176:42–51
Ranawat CS, Johanson NA, Rimnac CM, Wright TM, Schwartz RE (1986) Retrieval analysis of porous-coated components for total knee arthroplasty. A case report of two cases. Clin Orthop 209:244–248
Scott RD, Ewald FC, Walker PS (1984) Fracture of metallic tibial tray following total knee replacement. Report of two cases. J Bone Joint Surg [Am] 66:780–782
Skinner HB, Mabey MF, Paganelli JV, Meagher JM (1987) Failure analysis of PCA revision total knee replacement tibial component. A preliminary study using the finite element method. Orthopaedics 10:581–584
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Maruyama, M., Terayama, K., Sunohara, H. et al. Fracture of the tibial tray following PCA knee replacement. Arch Orthop Trauma Surg 113, 330–333 (1994). https://doi.org/10.1007/BF00426182
Received:
Issue Date:
DOI: https://doi.org/10.1007/BF00426182