Summary
The overall sensitivity of apparatus and transducers for diagnostic ultrasonography is measured with the measuring device typ 107 described by Buschmann.
The readings are in millimeters of penetration depth in paraffin oil, the latter is standardized by comparison with 73% Laevulose — in water — solution. The sensitivity values obtained are therefore comparable.
The values measured for overall sensitivity of two apparatus and a collection of transducers during several years revealed considerable variations. It was proved, that these differences had been caused by aging and destruction of electronic parts of the apparatus. After successful repairs the measuring results always reached again the previous level.
Additionally it was proved that the measuring device typ 107 respectively the paraffin oil used therein, guaranted constant measuring conditions over a period of some years.
Zusammenfassung
Zur Messung der Gesamtempfindlichkeit ultraschalldiagnostischer Untersuchungseinheiten (Geräte und Schallköpfe) wird das Ölstreckenmeßgerät nach Buschmann verwendet. Es liefert vergleichbare Meßwerte infolge der Standardisierung an 73 %iger Lävuloselösung.
Die Gesamtempfindlichkeitswerte, die an zwei Ultraschalldiagnostikgeräten im Verlaufe mehrerer Jahre mit einer Reihe von Schallköpfen gemessen wurden, zeigten erhebliche Schwankungen. Es konnte nachgewiesen werden, daß diese Schwankungen auf Verschleiß elektronischer Bauteile der Geräte beruhten und jeweils nach erfolgreicher Reparatur die vorherigen Werte wieder erreicht wurden.
Ferner konnte gezeigt werden, daß die Ölstreckenmeßgeräte, bzw. das enthaltene Öl, über mehrjährige Zeiträume gleichbleibende Meßbedingungen bieten.
Similar content being viewed by others
Literatur
Bergmann, L.: Der Ultraschall und seine Anwendung in Wissenschaft und Technik, 6. Aufl. Stuttgart: Hirzel 1954.
Buschmann, W.: Technische Fortschritte in der ophthalmologischen Ultraschalldiagnostik. Wiss. Z. Ernst-Moritz-Arndt-Univ. Greifswald, math.-nat. R. 12, 59–63 (1963).
—: Probleme und Fortschritte der Ultraschalldiagnostik am Auge. Klin. Mbl. Augenheilk. 144, 321–347 (1964).
—: Zur Ultraschalldiagnostik intraocularer Tumoren. Berlin, Humboldt-Univ., Med. Hab.-Schr. 1964.
—: Einige Meßverfahren zur Beurteilung der ultraschalldiagnostischen Untersuchungseinrichtungen (Geräte und Schallköpfe). SIDUO-Kongreßbericht. Wiss. Z. Humboldt-Univ. Berlin, math.-nat. R. 14, 115–119 (1965).
—: Klinische Ergebnisse und methodische Fortschritte in der echographischen Diagnostik intraocularer Tumoren. Klin. Mbl. Augenheilk. 148, 625–641 (1966).
—: Einführung in die ophthalmologische Ultraschalldiagnostik, 1. Aufl. Leipzig: VEB Georg Thieme 1966.
Filipczyński, L., and J. T. Groniowski: Visualization of the inside of the abdomen by means of ultrasonics and two methods for measuring ultrasonic doses. Digest of the 7th Internat. Conference on Medical and Biological Engineering, Stockholm, Sweden, August 14–19, 1967, p. 320.
-, and G. Lypacewicz: Ultrasonic problems of the diagnostic equipment for the eye examination. Vortrag auf dem SIDUO II, 24.–26. Mai 1967, Brno.
Matauschek, J.: Einführung in die Ultraschalltechnik. 1. Aufl. Berlin: VEB Verlag Technik 1957.
Ossoinig, K., u. K. Steiner: Zum Problem der Normung der Ultraschalldiagnostik des Auges. Albrecht v. Graefes Arch. klin. exp. Ophthal. 169, 241–249 (1966).
—: Weitere Erfahrungen mit Phantomen als Hilfsmittel zur Ultraschalldiagnostik intraocularer Tumoren. Albrecht v. Graefes Arch. klin. exp. Ophthal. 174, 1–8 (1967).
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Herrmann, G., Buschmann, W. Kontrolle der Gesamtempfindlichkeit von Ultraschalldiagnostikgeräten und Schallköpfen über lange Zeiträume. Albrecht von Graefes Arch. Klin. Ophthalmol. 175, 223–234 (1968). https://doi.org/10.1007/BF00418449
Received:
Issue Date:
DOI: https://doi.org/10.1007/BF00418449