Summary
Fractures of metallic implants are frequently caused by unstable osteosynthesis which facilitates vibrations and inner tensions of the metal. In some unfavourable cases we will have to expect vibration-corrosion. Among about 950 pertrochanteric femoral fractures we could observe 4 nail-fractures, besides 3 plate fractures and 5 screw fractures. Two of about 400 Smith-Peterson nails implanted for arthrodesis of the hip, broke. Three of 400 Küntscher nails and two of 150 Rush-pins broke. In nearly all of these cases the fracture of the metal could have been avoided by a better operation technique, partially by application of AO plates in pertrochanteric fractures that have no weak point in the area of the fracture partially by better reposition of the bone fracture. Working with the Küntscher nail the bone marrow space should be drilled to such a width that the nail fills completely the inner space up to the corticalis Rush-pins are generally to thin for fixation of greater bones. If the metallic fixation of fractures is only unstable we must use a cast fixation externally in addition.
Résumé
Les fractures des metaux utilisés pour fixation interne d'une fracture osseuse, ont leur origine frequemment dans une ostéosynthèse non stabilisante. Dans ces cas les metaux devont subir des vibrations et des tensions internes, terminant parfois dans une corrosions par vibration. Nous trouvâmes 4 clous-Böhler fracturés en 950 operations pour une fracture transcondylaire du femur. Trois clous de Küntscher de 400 et deux Rush-pins en 150 étaient fracturés. Presque dans tous les cas une meilleure technique operatoire aurait empéché ces malheurs. Si l'on peut pas stabiliser une fracture osseuse par des metaux interne il était préférable d'appliquer une fixation externe quelquonque.
Zusammenfassung
Die Ursache der Brüche metallischer Implantate liegt sehr häufig in einer instabilen Osteosynthese und der dadurch ermöglichten Schwingung und Vorspannung des Metalls. Es kommt daher in einigen ungünstigen Fällen zur Schwingungsrißkorrosion. Unter etwa 950 pertrochanteren Oberschenkelfrakturen fanden wir 4 Nagelfrakturen, außerdem 3 Plattenbrüche und 5 Schraubenbrüche. Bei etwa 400 Nagelungen mit dem Smith-Peterson-Nagel wurden 2 Nagelbrüche nach Hüftgelenksarthrodese festgestellt. Unter etwa 400 Küntscher-Nägeln brachen 3, und von ca. 150 Rushpins brachen 2. In fast allen Fällen ließe sich eine Metallfrakturierung durch eine bessere Operationstechnik vermeiden, sei es, daß für die pertrochanteren Brüche AO-Platten verwendet werden, die im Bereich der Fraktur keine Kerbstelle besitzen, sei es, daß sorgfältiger reponiert wird. Bei der Küntscher-Nagelung muß auf eine solche Dicke aufgebohrt werden, daß ein entsprechender Nagel den Markraum innen im Bruchbereich fest ausfüllt. Rushpins sind zur Fixierung großer Röhrenknochen generell zu dünn. Ist die Fraktur nur instabil zu versorgen, so wird eine zusätzliche äußere Gipsfixierung notwendig.
Similar content being viewed by others
Literatur
Bauder, K.: Marknagelverbiegungen. Chir. Praxis 12, 591–595 (1968).
Böhler, L.: Technik der Knochenbruchbehandlung, Bd. III, S. 105. Wien: Maudrich 1956.
Bowden, F. P., I. B. P. Williamson, and P. G. Laing: Metallic transfer in screwing and its significance in bone surgery. Nature (Lond.) 173, 520–522 (1954).
Bürkle de la Camp, H.: Fehler und Gefahren in der Knochen- und Gelenkchirurgie. Langenbecks Arch. klin. Chir. 289, 463–475 (1958).
Clarke, E., and J. Hickman: An investigation into the correlation between the electrical potentials of metals and their behaviour in biological fluids. J. Bone Jt Surg. 35-B, 467–473 (1953).
Cohen, J.: Corrosion-testing for orthopedic implants. J. Bone Jt Surg. 44-A, 307–316 (1962).
Contzen, H.: Grundlagen der Alloplastik mit Metallen und Kunststoffen. Stuttgart: Thieme 1967.
Dürr, W., H. W. Bumm u. R. Helms: Über Allenthesen zur operativen Behandlung von Frakturen im Bereiche des Trochantermassivs. Chirurg 36, 165–175 (1966).
Emneus, H., and U. Stenram: Reaction to alloys used in osteosynthesis. Acta orthop. scand. 24, 315–330 (1960).
Ferguson, A. B., P. G. Laing, and E. S. Hodge: The ionization of metal implants in living tissues. J. Bone Jt Surg. 42-A, 77–90 (1960).
Fischer, S., u. H. Zitter: Verhalten von chemisch beständigen Stählen im menschlichen Körper. Werkstoffe und Korrosion 20, 753–758 (1963).
Fischer-Wasels, J., u. H. B. Schünemann: Zur Frage von Marknagelbrüchen. Arch. orthop. Unfall-Chir. 46, 207–228 (1953).
Gough, H. J. J.: Inst. Met. 49, 17 (1932); zit. n. H. Spähn.
Häbler, C.: Die Marknagelung nach Küntscher. Berlin-München: Urban & Schwarzenberg 1953.
Hermann, L.: Nagelbruch und Verbiegung als Spätkomplikation nach Marknagelung. Zbl. Chir. 85, 354–361 (1960).
Kemlein, W.: Extraktionsgerät für Schenkelhalsnagelbruchstücke mit V-förmigem Querschnitt. Chirurg 38, 385 (1968).
Kirsch, J.: Die Stabilität des Küntscher-Nagels. Mschr. Unfallheilk. 62, 143–147 (1959).
Koch, W.: Erfahrungen bei Anwendung der Spreizschraube zur Fixation von Olecranonbrüchen. Zbl. Chir. 93, 195–196 (1968).
Küntscher, G.: Praxis der Marknagelung. Wien: Maudrich 1959.
—— Die Osteosynthese. Zbl. Chir. 85, 828–840 (1960).
Letiz, G.: Typische Komplikationen nach Osteosynthesen und ihre mechanischen Ursachen. Arch. orthop. Unfall-Chir. 64, 285–297 (1968).
Longinow, A. W.: Stress corrosion testing of alloys. Materials protection 8, 33–39 (1966).
Seyfarth, H.: Beitrag zur Anwendung von Metallen in der plastischen Chirurgie. Arch. orthop. Unfall-Chir. 47, 656–663 (1955).
Spähn, H.: Zur Schwingungsrißkorrosion metallischer Werkstoffe. Metalloberfläche 16, 197–202, 233–239, 267–273, 299–307, 335–340, 369–373 (1967); 17, 1–9 (1968).
Täger, K. H.: Materialprüfung für Metalle in der Orthopädie und Chirurgie. Chirurg 33, 129–131 (1962).
Werhahn, O.: Beitrag zur Spannungsentwicklung in Knotenpunkten geschweißter Fachwerkbrücken. Dissertation Hannover 1955.
Wright, J. K., and J. H. Axon: Electrolysis and stainless steel in bone. J. Bone Jt Surg. 38, 745–753 (1956).
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
v. Elmendorff, H.F., Simmert, H. & Hupfauer, W. Über Brüche von metallischen Implantaten. Arch orthop Unfall-Chir 67, 141–154 (1969). https://doi.org/10.1007/BF00416834
Received:
Issue Date:
DOI: https://doi.org/10.1007/BF00416834