Das Mackay-Marg-Tonometer

II. Experimentelle Analyse des Tonogramms

The Mackay-Marg tonometer

II. Experimental analysis of the tonogram

  • Published:
Albrecht von Graefes Archiv für klinische und experimentelle Ophthalmologie Aims and scope Submit manuscript

Summary

The probe of the Mackay-Marg Tonometer (MMT) was mounted on the chin-rest and a transparent pressure-chamber on the light-arm of the slitlamp in such a manner that, observed through the pressure chamber, any size of applanation could be exactly adjusted. To each of the chosen applanation the response of the MMT was registered. The examinations were performed in open and closed stopcock of the chamber on a thin rubber-membrane and on the cornea of enucleated human eyes.

On the rubber membrane the MMT indicated properly the pressure of the chamber as soon as the contact area covered the plunger and the innerst zone of the footplate. On the cornea however first definitely higher values were indicated by the MMT in small applanations and the true level of the pressure not before the size of applanation increased to that of the whole footplate. Finally due to the studies performed the trough of the tonogram is correlated to a corneal applanation the diameter of which is equal to the diameter of the footplate.

Zusammenfassung

An der Spaltlampe wurden die Sonde des Mackay-Marg-Tonometers (MMT) an der Kinstütze und eine durchsichtige Druckkammer am Lichtarm so befestigt, daß durch die Druckkammer hindurch beobachtet, beliebig große Applanationen exakt eingestellt werden konnten. Durch kurzes Einschalten des Schreibers wurde die der jeweils eingestellten Applanation zugeordnete Amplitude des MMT registriert.

Die Untersuchungen wurden bei offenem und bei geschlossenem Hahn der Druckkammer an der Leichenhornhaut und an einer dünnen Gummimembran durchgeführt.

Durch die Gummimembran hindurch registrierte das MMT den Kammerdruck, sobald die Grenzen der Kontaktfläche die innere Fußplattenzone, durch die Leichenhornhaut dagegen erst bei einer Applanation, deren Grenzen den äußeren Rand der Fußplatte erreicht haben.

Das Tal der negativen Zacke im aufsteigenden Schenkel des Tonogramms entspricht, den gegenwärtigen Untersuchungen zufolge, einer Hornhautapplanation deren Durchmesser dem der Sondenfußplatte gleichzusetzen ist.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

Literatur

  • Hilton, G. F., Schaffer, R. N.: Electronic applanation tonometry. Amer. J. Ophthal. 62, 838–843 (1966).

    Google Scholar 

  • Mackay, R. S., Marg, E., Oechsli, R.: Automatic tonometer with exact theory. Various biological applications. Science 131, 1668–1669 (1960).

    Google Scholar 

  • Marg, E., Mackay, R. S., Oechsli, R.: Corneal bending and buckling in tonometry. Arch. Ophthal. 65, 67–74 (1961).

    Google Scholar 

  • Moses, R. A., Marg, E., Oechsli, R.: Investigation of the basic validity and clinical usefullness of the Mackay-Marg tonometer. Invest. Ophthal. 1, 78–85 (1962).

    Google Scholar 

  • Stepanik, J.: Das Verhalten der vorderen und hinteren Hornhautfläche während der Applanation. Albrecht v. Graefes Arch. klin. exp. Ophthal. 176, 277–282 (1968).

    Google Scholar 

  • —: Das Mackay-Marg-Tonometer. I. Klinische Studien. Albrecht v. Graefes Arch. klin. exp. Ophthal. 180, 104–112 (1970).

    Google Scholar 

  • - The Mackay-Marg Tonometer: Correlation of the tonogram to the corneal applanations induced by the tonometer. Acta Ophthal. (in press).

Download references

Author information

Authors and Affiliations

Authors

Additional information

Gefördert durch einen Beitrag aus einer Sonderdotation des Österreichischen Bundesministeriums für Unterricht.

Rights and permissions

Reprints and permissions

About this article

Cite this article

Stepanik, J. Das Mackay-Marg-Tonometer. Albrecht von Graefes Arch. Klin. Ophthalmol. 181, 57–64 (1970). https://doi.org/10.1007/BF00410811

Download citation

  • Received:

  • Issue Date:

  • DOI: https://doi.org/10.1007/BF00410811

Navigation