Summary
Two retrieved Moore type hip prostheses of cast Co-Cr-Mo-alloy were examined after metallic failure.
The first case concerns a fatigue fracture in the stem of the prosthesis due to an abrupt change in the cross-sectional area and the presence of inappropriate welding beads in the stem. The abrupt change occurring at a transition from an I-beam into a rectangular profile gives rise to stress concentrations thus increasing chance of fatigue crack initiation.
The second case is a fracture of the hollow head of the prosthesis through a weld. The two parts of the cast femoral head were incompletely welded. Comparison with similar welds in other femoral heads of prostheses of the same type reveals a poor welding uniformity.
Hence, both fractures are associated with the welding process. A critical investigation is needed to find optimal welding conditions for Co-Cr-Mo-alloy prostheses.
Zusammenfassung
Es wurden zwei gebrochene Femurkopfendoprothesen des Typs Moore aus Co-Cr-Mo-Gußlegierung untersucht. Beim ersten Fall handelt es sich um einen Ermüdungsbruch des Prothesenschaftes an der Stelle eines Wechsels im Schaftprofil and eines unzulässigen Schweißprozesses, der auf dem Schaft ausgeführt wurde. Die plötzliche Verdnderung, die beim Übergang von einem I-Profil in ein rechteckiges Profil entsteht, gibt Anlaß zu Spannungskonzentrationen, die die Entstehung von Ermüdungsrissen begünstigen.
Beim zweiten Fall handelt es sich um einen Bruch in der Circumferenz des hohlen Kugelkopfes der Prothese, an der Stelle einer minderwertigen Schweßnäht. Beide Teile des gegossenen Kopfes waren unvollständig miteinander verschweißt. Ein Vergleich der gleichartigen Schweißnähte anderer Prothesenköpfe vom gleichen Typ and Fabrikat zeigte, daß die Schweißprozesse sehr ungleichmäßig ausgeführt worden waren. Somit entstanden in beiden Fällen die Prothesenbrüche durch einen falschen Schweißprozeß. Die optimalen Schweißbedingungen von Prothesen aus Co-Cr-Mo-Legierungen müssen durch eine kritische Untersuchung noch erforscht werden.
Similar content being viewed by others
References
Charnley J (1975) Fracture of femoral prostheses in total hip replacement. A clinical study. Clin Orthop 111:105–120
Dobbs HS (1978) Welds in the stems of cast cobalt chrome femoral components for use in total hip replacements. Engin Med 7:31–33
Ducheyne P, De Meester P, Aernoudt E, Martens M, Mulier JC (1975) Fatigue fractures of the femoral component of Charnley and Charnley-Müller type total hip prostheses. J Biomed Mater Res Symp 6:199–219
Ducheyne P, De Meester P, Aernoudt E (1980) Performance analysis of total hip prostheses; some particular metallurgical observations. J Biomed Mater Res 14:31–40
Galante JO, Rostoker W, Doyle JM (1975) Failed femoral stems in total hip prostheses. A report of six cases. J Bone Jt Surg 57-A:230–236
Markolf KL, Amstutz HC (1976) A comparative experimental study of stresses in femoral total hip replacement components: the effect of prostheses orientation and acrylic fixation. J Biomech 9:73–79
Rostoker W, Chao EYS, Galante JO (1978) Defects in failed stems of hip prostheses. J Biomed Mater Res 12:635–651
Smethurst E, Waterhouse RB (1977) Causes of failure in total hip prostheses. J Mater Sci 12:1781–1792