Summary
It is known that the loosening of artificial hip joints leads to their shift relative to the carrying bone structure. For a quantitative radiological detection of even minimal dislocations fix-points both in the carrying bone structure as well as in the prostheses have to be defined. While for the stem prostheses and cups such points can be defined directly from their geometrical structure, they have to be generated artificially for the carrying bone structure by implanted metal pins, which can easily be detected in the X-ray print.
Having defined in this way a complete and appropriate set of fix-points, the position of the prosthesis relative to the bone structure is uniquely defined by the distance vectors between the different fix-points. The main biometrical steps of such a concept as well as their computational formulation and realization are derived; the main advantages and the accuracy of the technique are elucidated and critically discussed.
Zusammenfassung
Kommt es zu einer Lockerung eines alloplastischen Hüftersatzes, so verlagert sich dieser relativ zu den ihn verankerten Knochenstrukturen. Um minimale Positionsänderungen des künstlichen Gelenkkörpers bereits radiologisch erfaßbar zu machen, müsen sowohl im Knochen als auch an den Prothesenteilen ausmeßbare Strukturen gefunden werden. Die konstante Raumgeometrie von Schaftprothese, Schalenprothese und Hüftpfanne ermöglicht die Definition von Fixpunkten an diesen künstlichen Gelenkkörpern, im Bereich des Knochens werden intraoperativ mehrere zylinderförmige Metallstifte, die sich radiologisch darstellen, versenkt und ermöglichen damit die Definition von Fixpunkten im Inplantatbett. Zwischen diesen Fixpunkten werden Abstandsvektoren berechnet, die als Informationsträger für die Relativlage der Prothesenteile in ihrem Implantatlager dienen. Die biometrischen Grundlagen und die programmtechnische Realisierung und Mcßgenauigkeit dieser Lokalisationsdiagnostik werden dargestellt.
Similar content being viewed by others
Literatur
Dreyer, J., Georgi, P., Prevot, H.: Der Aussagewert der Szintigraphie bei Verdacht auf Endoprothesenlockerung. Orthop. Praxis 1/XIII, 45 (1977)
Freeman, M.: ICLH Cemented Double Cup Arthroplasty of the Hip: The results in the first 5 years. 64. Tagung der Deutschen Gesellschaft fur Orthopädie und Traumatologie, Würzburg, 1977
Gajewski, H.: Methoden der Bildbeeinflussung. In: Bildgüte in der Radiologie. Stuttgart: Fischer 1964
Hackenbroch, M. H., Bruns, H., Hobe, R., Lechleuther, X.: Unsere Ergebnisse mit der Totalprothese des Hüftge-lenkes 3-6 Jahre postoperativ. Orthop. Praxis 6/XII, 586 (1976)
Köhnle, H.: Röntgenstereoverfahren. In: Handbuch der medizinischen Radiologie, Band III. Berlin, Heidelberg, New York: Springer 1967
Meiler, J.: Die Begrenzung der Detailerkennbarkeit im Rötgenbild durch den Schwellenkontrast. Fortschr. Röntgenstr. 127, 179 (1977)
Paul, J. P.: Design aspects of Endoprostheses for the lower limb. In: Perspectives in Biomechanical Engineering. R. M. Kenedi (ed.). London: MacMillan 1973
Probst, K. J., Baacke, M.: Röntgenstereoverfahren in der Orthopädie. Orthopädie 5, 195 (1976)
Rosenow, U.: Untersuchungen über Grundlagen und Leistungsfähigkeit einer neuen röntgen-stereo-photogram-metrischen Meßmethode. Diss. Land., Göttingen 1969
Rydell, N.: Forces in the Hip Joint. II. Intravital Studies. In: Biomechanics and Related Bioengineering Topics. R. M. Kenedi (ed.). London: Pergamon 1965
Salvati, E. A., Freiberger, R. H., Wilson, E. D., Jr.: Arthrography for Complications of Total Hip Replacement. J. Bone Jt. Surg. 53 A, 701 (1971)
Schulitz, K. P., Gärtner, B., Hüwel, R.: Schwingungsmechanische Methode zur Erfassung einer Prothesenlockerung. Orthop. Praxis 1/XIII, 42 (1977)
Schumann, Z., Darup, K., Beck, H.: Möglichkeiten und erste Erfahrungen der Ganzkörpertomographie (Delta-Scan) in der Extremitätenchirurgie. Vortrag auf dem 94. Deutschen Chirurgenkongreß, München, 1977
Suezawa, Y., Dietschi, C.: Prothesenwechsel am Huftgelenk. Z. Orthop. 115, 159 (1977)
Ungethüm, M.: Biomechanische Untersuchungen zum Problem von Schaftbrüchen bei Hüftprothesen. 64. Tagung der Deutschen Gesellschaft fur Orthopädie und Traumatologie, Würzburg, 1977
Wagner, H.: Der alloplastische Gelenkflächenersatz am Hüftgelenk. Arch. orthop. Unfall-Chir. 82, 101 (1975)
Weigand, H., Ritter, G.: Die röntgenologische Beurteilung des totalprothetischen Hüftgelenkersatzes und seiner Komplikationen. Dtsch. med. Wschr. 99, 20–46 (1974)
Author information
Authors and Affiliations
Additional information
Mit Unterstützung der Deutschen Forschungsgemeinschaft Herrn OA Dr. med. M. C. Baacke gewidmet
Rights and permissions
About this article
Cite this article
Probst, K.J., Herp, A. & Leidel, W. Stereo-radiologische Lokalisationsdiagnostik von Schalen- und Totalendoprothesen des Hüftgelenkes. Arch. Orth. Traum. Surg. 92, 75–81 (1978). https://doi.org/10.1007/BF00381644
Received:
Issue Date:
DOI: https://doi.org/10.1007/BF00381644