Summary
A high pressure xenon arc lamp XBO 2500 W is used as a radiation source for the production of quantum fluxes of the order of 6 μEinstein/sec at λ=700 nm and of 2 μEinstein/sec at λ=400 nm. Wavelength selection between 300 and 900 nm is made by tandem interference filters of the band pass type (band width at half peak transmission approximately 16 nm), with a total transmission outside of passband<0.08%. The radiation load on the filters is reduced by the use of interchangeable selective mirrors. The front surface of the interference filters is cooled with de-ionized water (metal interference filters absorb more than 30% of the impinging visible radiation). Homogenity of irradiance is achieved by means of a “honeycomb condensor”. A simple cascaded feedback system sensitive to changes in lamp current and irradiance keeps radiation intensity constant with time. Technical details including safety devices are given.
Zusammenfassung
Mit einer Lampe XBO 2500 W können “monochromatische” Quantenflüsse von der Größenordnung von 2 μE s -1 bei λ=400 nm und 6 μE s -1 bei λ=700 nm erzielt werden. Die Wellenlängenauswahl zwischen 300 und 900 nm erfolgt mit doppelten Interferenz-Bandfiltern (DAL), deren gesamter spektraler Untergrund unter 0,08% der durchgelassenen Strahlung bleibt. Die Strahlungsbelastung der Filter wird durch Vorschalten auswechselbarer selektiver Spiegel gemindert. Zusätzlich müssen die Filter an ihren Frontflächen mit Wasser gekühlt werden. Die Homogenität des Bestrahlungsfeldes wird durch einen Wabenkondensor erreicht. Eine Kaskadenregelung ermöglicht die zeitliche Konstanz der Bestrahlungsstärke. Technische Details, die Sicherheitsschaltungen einschließen, werden beschrieben.
Similar content being viewed by others
Literatur
Ito, T.: Ultraviolet irradiation apparatus with grating monochromator. Jap. J. Genetics 39, 151–154 (1964).
Jacques, R., R. Chabbal, P. Chouard et P. Jacquinot: Mise au point d'un illuminateur spectral à usage biologique. C.R. Acad. Sci. (Paris) 259, 1581–1584 (1964).
Johns, H. E., A. M. Rauth: Theory and design of high intensity u.v. monochromators for photobiology and photochemistry. Photochem. Photobiol. 4, 673–692 (1965).
Mohr, H., u. G. Schoser: Eine Interferenzfilter-Monochromatoranlage für photobiologische Zwecke. Planta (Berl.) 53, 1–17 (1959).
——: Eine mit Xenonbögen ausgerüstete Interferenzfilter-Monochromatoranlage für kurzwellige sichtbare und langwellige ultraviolette Strahlung. Planta (Berl.) 55, 143–152 (1960).
Osram GmbH: Xenon-Hochdrucklampen für Lichtwurfzwecke. Firmenschrift 1964.
Osram GmbH: Hinweise zur Ausgestaltung von Geräten für Xenon-Hochdrucklampen XBO. Firmenschrift, o.J. 1964.
Schläfer, R.: Anwendung monochromatischer Filter in den Naturwissenschaften. Druckschrift Nr. 2514 des Jenaer Glaswerkes Schott & Gen., Mainz, o.J. 1964.
Withrow, R. B.: An interference-filter monochromator system for the irradiation of biological material. Plant. Physiol. 32, 355–360 (1957).
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Raschke, K. Eine Anlage zur monochromatischen Bestrahlung biologischer Objekte, ausgerüstet mit Interferenzfiltern und einer elektronisch geregelten 2,5 kW-Xenonlampe. Planta 75, 55–72 (1967). https://doi.org/10.1007/BF00380839
Received:
Issue Date:
DOI: https://doi.org/10.1007/BF00380839