The disturbing artefacts from fillings in teeth were experimentally studied. Pins of amalgam and composite with different diameters were scanned in 10 mm and 5 mm slices. Amalgam gave five different artefacts: (1) blackout, (2) tuning fork artefact, (3) star-burst, (4) collimation artefact in the narrow slice and (5) connecting line artefact. The composite material with an attenuation of about 5000 HU gave no blackout and mild degree of the other artefacts except for the collimation artefact, which was equal to that with amalgam.
Similar content being viewed by others
References
ADA specification No. 27 for direct filling material. J. Am. Dent. Assoc. 94, 1191–1194 (1977)
ADA specification No. 1. J. Am. Dent. Assoc. 95, 614–617 (1977)
Aronow, S.: Studies on physical phantoms in the EMI scanner. J. Can. Assoc. Radiol. 27, 149–156 (1976)
Christensen, E. E., Curry III, T. S., Dowdey, J. E.: An introduction to the physics of diagnostic radiology, p. 69. Philadelphia: Lea & Febiger 1978
EMI Medical 1010 System Manual (C.S.D. Technical Publications Department, EMI Medical Limited, Slough, England). Side 3–30, 1978
Frisén, L., Kjällman, L., Svendsen, P.: Computed frontal tomography (CFT) of the orbit using the EMI CT 1010 head scanner. Neuroradiology, 15, 263–265 (1978)
Kingsley, D. P. E.: Computer tomographic artefacts using the CT 1000. In: The first European seminar on computerized axial tomography in clinical practice pp. 36–42. (eds. du Boulay G. H., Moseley I. F.) Berlin, Heidelberg, New York: Springer 1977
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Svendsen, P., Quiding, L. & Landahl, I. Blackout and other artefacts in computed tomography caused by fillings in teeth. Neuroradiology 19, 229–234 (1980). https://doi.org/10.1007/BF00347800
Received:
Issue Date:
DOI: https://doi.org/10.1007/BF00347800