Correct operation of a scintillation camera in cerebral scintigraphy

Bon fonctionnement d'une caméra à scintillation en scintigraphie cérébrale

Korrekte Ergebnisse einer Szintilations-Camera in der cerebralen Szintigraphie

  • Originals
  • Published:
Neuroradiology Aims and scope Submit manuscript

Summary

In a direct clinical comparison of 137 patients submitted to cerebral scintigraphy with both a scintillation camera and a rectilinear scanner, a complete coincidence of results was obtained — 42 positives and 7 false-negatives with both machines. Subjective comparisons of the images obtained from the 42 positive cases indicated that the camera provided a marginally better yield of diagnostic information.

Résumé

L'auteur a comparé les résultats obtenus chez 137 patients soumis à la scintigraphie cérébrale, à l'aide d'une caméra à scintillation d'une part et d'un scanner linéaire d'autre part. Dans les deux cas, les résultats coincidaient, soit 42 positifs et 7 faux négatifs. La comparaison des deux images obtenues dans les 42 cas positifs montrait que la caméra a un rendement légèrement supérieur.

Zusammenfassung

Bei einem klinischen Vergleich von 137 Patienten, bei denen Isotopen-Untersuchungen sowohl mit einer Szintilations-Camera als auch mit einem rectilinearen Scanner durchgeführt wurden, ergab sich eine vollständige Übereinstimmung der Ergebnisse (42 positive und 7 falsch-negative Untersuchungen mit beiden Methoden). Der subjektive Vergleich der Untersuchungen der 42 positiven Fälle zeigt, daß die Camera eine bessere diagnostische Information liefert.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

Reference

  • Burrows, E. H.: Brit. med. J. 1, 473 (1972a).

    Google Scholar 

  • Burrows, E. H.: Progress in nuclear medicine 1, 287 (1972b).

    Google Scholar 

  • Collica, C. J., Robinson, T., Hayt, D. B.: Amer. J. Roentgenol. 100, 761 (1967).

    Google Scholar 

  • Dowding, D. V., Kimber, P. M.: Radiography 38, 255, 1972.

    Google Scholar 

  • Gottschalk, A., Anger, H. O.: J. Amer. med. Ass. 192, 448 (1965).

    Google Scholar 

  • Keeling, D. K.: Proc. roy. Soc. Med. 64, 340 (1971).

    Google Scholar 

  • Loken, M. K., Wigdahl, L. O., Gilson, J. M., Staab, E. V.: J. Nucl. Med. 7, 209 (1966).

    Google Scholar 

  • McCready, V. R.: Erstes Heidelberger Symposium über Kameraszintigraphie, 177 (1968).

  • Pedersen, M., Haase, J.: Acta radiol. Diagn. 10, 534 (1970).

    Google Scholar 

  • Pollycove, M., Fish, M. B., Khentigan, A.. J. Nucl. Med. 8, 321 (1967).

    Google Scholar 

  • Schenk, P., Penzholz, W., Pietrowski, W., Tornow, W.: Erstes Heidelberger Symposium über Kameraszintigraphie, 119 (1958).

  • Telander, G. T., Loken, M. K.: J. Nucl. Med. 8, 481 (1967).

    Google Scholar 

  • Van Eck, J. H. M., Penning, L.: Neuroradiology 1, 107 (1970)

    Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Rights and permissions

Reprints and permissions

About this article

Cite this article

Burrows, E.H. Correct operation of a scintillation camera in cerebral scintigraphy. Neuroradiology 5, 77–81 (1973). https://doi.org/10.1007/BF00337487

Download citation

  • Issue Date:

  • DOI: https://doi.org/10.1007/BF00337487

Keywords

Navigation