Metabolic phenomena can be studied and measured non-invasively using positron emitting radionuclides and a suitably adapted tomographic system. The choice of a single slice ring camera is justified by its physical performance, which is presented here and discussed. A series of measurements with geometrical phantoms and analytical simulations have been performed to determine the critical characteristics of the system. This has permitted optimization of certain parameters enabling very interesting clinical results to be obtained at SHFJ, particularly in the area of cerebral physiopathology.
In addition, the potential of obtaining absolute quantitative values of regional activity is presented. The calibration of the system, spatial non-stationarities, and attenuation correction, which represent the main sources of error, are considered in detail.
A precision of the order of 10% should be obtainable. Such a quantitation method has been successfully applied to the in vivo study of the regional extraction of cerebral oxygen.
Similar content being viewed by others
References
Baron, J.C., Comar, D., Bousser, M.G., Soussaline, F., Crouzel, C., Kellershohn, C., Castaigne, P.: Etude tomographique chez l'homme, du débit sanguin et de la consommation d'oxygène du cerveau par inhalation continue d'oxygène 15. Rev. Neurol. (Paris) 134, 545–556 (1978)
Berger, G., Mazière, M., Marazano, C., Comar, D.: Carbon 11 labelling of the psychoactive drug O-methyl-bufornine and its distribution in the animal organism. Eur. J. Nucl. Med. 3, 101–104, (1978)
Brownell, G.L., Burnham, C.A., Wilensky, S., Aronow, S., Kazemi, S., Strieder, D.: New Developments in positron scintigraphy and the application of cyclotron produced positron emitters. Medical radioisotope scintigraphy, Proc. Symp. Salzburg (1968) 1, pp. 163–176. Vienna: IAEA 1968
Brownell, G.L., Burnham, C.A.: In: Tomographic imaging in nuclear medicine, Friedman, G.S. (ed), pp 154–164. New York: Society of Nuclear Medicine 1973
Brownell, G.L., Burnham, C.A., Hoop, B.Jr.,: Positron scintigraphy with short-lived cyclotron produced radiopharmaceuticals. In: Medical radioisotope scintigraphy, 1, 313–330, Vienna: IAEA 1973
Budinger, T.F., Derenzo, S.E., Greenberg, W., Gullberg, G.T., Huesman, R.H.: Quantitative potentials of dynamic emission computed tomography. J. Nucl. Med., 19, 309–315 (1978)
Cho, Z.H., Cohen, M.B., Singh, M.: Performance and evaluation of the circular ring transverse axial positron camera (CRTAPC). IEEE Nucl. Sci. NS 24, 530–543, (1977)
Comar, D., Mazière, M., Berger, G., Mestelan, G.: Quelques exemples de l'intérêt des molécules marquées au 11C en recherche médicale. BIST 220, 19–26 (1976)
Crouzel, C., Comar, D., Berger, G., Mazière, M., Soussaline, F., Todd-Pokropek, A.E., Baron, J.C., Guenard, H., Verhas, M.: First clinical results using short lived radioisotopes produced by a medical cyclotron. In: VIIIth Int. Conf. on Cyclotrons and their medical applications. Bloomington: 1978
Derenzo, S.E., Budinger, T.F., Camoon, J.L.: High resolution computed tomography of positron emitters. IEEE Nucl. Sci. NS 24, 544–558 (1977)
Derenzo, S.E., Zaklad, H., Budinger, T.F.: Analytical study of a high resolution positron ring detector system for transaxial reconstruction tomography. J. Nucl. Med. 16, 1166–1173 (1975)
Jones, T., Chesler, D.A., Ter-Pogossian, M.M.: The continuous inhalation of oxygen 15 for assessing regional oxygen extraction in the brain of man. Br. J. Radiol. 49, 339–343 (1976)
Lim, C.K., Chu, D., Kaufman, L.: Initial characterization of a multiwire proportional chamber positron camera. IEEE Nucl. Sci. NS 22, 388–394 (1975)
Muehllehner, G., Buchin, N.P., Dubek, G.H.: Performance parameters of the positron imaging camera. IEEE Nucl. Sci. NS 23, 528–537 (1976)
Phelps, M.E., Hoffman, E.J., Kuhl, D.E.: Physiologic tomography: A new approach to in vivo measure of metabolism and physiological function. In: Medical radionuclide imaging, 1, 233–253, Vienna: IAEA 1977
Phelps, M.E., Hoffman, E.J., Mullani, N.: Design and performance characteristics of a whole body transaxial tomograph. (PETT III). IEEE Nucl. Sci. 23, 516–522 (1976)
Phelps, M.E., Hoffman, E.J., Mullani, N.: Design and performance characteristics of a whole body transaxial tomograph. J. Nucl. Med 17, 493–502 (1976)
Phelps, M.E., Hoffman, E.J., Huang, S.C., Kuhl, D.E.: ECAT a new computerized tomographic imaging system for positronemitting radiopharmaceuticals. J. Nucl. Med. 19, 635–647 (1978)
Shepp, L.A., Logan, B.F.: Some insights into the Fourier reconstruction of a head section. IEEE Nucl. Sci. NS 21, 21–43 (1974)
Syrota, A., Comar, D., Cerf, M., Plummer, D., Mazière, M., Kellershohn, C.: 11C-Methionine pancreatic scanning with positron emission computed tomography. J. Nucl. Med. (in press) (1979)
Ter-Pogossian, M.M., Phelps, M.E., Brownell, G.L.: Reconstruction tomography in diagnostic radiology and nuclear medicine. Baltimore: University Park Press 1977
Ter-Pogossian, M.M., Phelps, M.E., Hoffman, E.J.: A positron emission transaxial tomograph for nuclear medicine imaging (PETT). Radiology 114, 89–98 (1975)
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Soussaline, F., Todd-Pokropek, A.E., Plummer, D. et al. The physical performances of a single slice positron tomographic system and preliminary results in a clinical environment. Eur J Nucl Med 4, 237–249 (1979). https://doi.org/10.1007/BF00304879
Received:
Published:
Issue Date:
DOI: https://doi.org/10.1007/BF00304879