Optimal preprocessing Butterworth-Wiener filter for Tl-201 myocardial SPECT

  • Published:
European Journal of Nuclear Medicine Aims and scope Submit manuscript

Abstract

The optimal frequency characteristic of Butter-worth-Wiener filters (BWF) for improving the image quality of 201 Tl myocardial SPECT was determined by a phantom experiment. Thirty two projection images of the phantom containing 11.1 MBq of 201Tl with 4 different cold lesions were collected during a 180° arc of a gamma camera. A set of the projection images were processed with each of 27 different BWFs, and SPECT images were reconstructed by Shepp-Logan filtered backprojection. The SPECT images were evaluated for their diagnostic ability to visibly detect the cold lesions by four nuclear medicine physicians. The lesion contrasts were used as an adjunctive tool to determine the optimum filter. The optimal combination of the parameters determining BWF characteristics (for the data of about 100 count/pixel at the myocardium) is: 1) cutoff of 0.25/pixel, 2) FWHM of 3 pixels, 3) noise/signal ratio of 0.02. FWHM and noise/signal ratio affected lesion contrast much less than cutoff frequency. Clinical myocardial SPECT images processed with the optimal BWF showed less noise and sharper delineation of the myocardium.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

References

  • Gonzalez RC, Wintz P (1977) Digital image processing. Addison-Wesley, London

    Google Scholar 

  • King MA, Doherty PB, Schiwinger RB, Jacobs DA, Kider RE, Miller TM (1983) Fast count-dependent digital filtering of nuclear medicine images: concise communication. J Nucl Med 24:1039–1045

    Google Scholar 

  • King MA, Ronald BS, Doherty PW, Penny BC (1984) Two-dimensional filtering of SPECT images using the Metz and Wiener filters. J Nucl Med 27:1327–1336

    Google Scholar 

  • King MA, Ronald BS, Penny BC, Doherty PW, Bianco JA (1986) Digital restoration of Indium-111 and Iodine-123 SPECT images with optimized Metz filters. J Nucl Med 27:1327–1336

    Google Scholar 

  • Madsen MT, Park CH (1985) Enhancement of SPECT images by Fourier filtering the projection image set. J Nucl Med 26:395–402

    Google Scholar 

  • Rosenfeld A, Kak AC (1976) Digital Picture Processing, Academic Press, New York

    Google Scholar 

  • Shepp LA, Logan BF (1974) The Fourier reconstruction of a head section. IEEE Trans Nucl Sci 21:21–43

    Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Rights and permissions

Reprints and permissions

About this article

Cite this article

Honda, N., Machida, K., Tsukada, J. et al. Optimal preprocessing Butterworth-Wiener filter for Tl-201 myocardial SPECT. Eur J Nucl Med 13, 404–407 (1987). https://doi.org/10.1007/BF00292492

Download citation

  • Received:

  • Issue Date:

  • DOI: https://doi.org/10.1007/BF00292492

Key words

Navigation