The reproducibility of single photon absorptiometry in a clinical setting

  • Published:
European Journal of Nuclear Medicine Aims and scope Submit manuscript

Abstract

The reproducibility of single photon absorptiometry (SPA) results for detection of changes in bone mineral content (BMC) was evaluated in a clinical setting. During a period of 18 months with 4 different sources, the calibration scans of an aluminium standard had a variation of less than 1% unless the activity of the 125I source was low. The calibration procedure was performed weekly and this was sufficient to correct for drift of the system. The short term reproducibility in patients was assessed with 119 duplicate measurements made in direct succession. The best reproducibility (CV=1.35%) was found for fat corrected BMC results expressed in g/cm, obtained at the site proximal to the 8 mm space between the radius and ulna. Analysis of all SPA scans made during 1 year (487 scans) showed a failure of the automatic procedure to detect the space of 8 mm between the forearm bones in 19 scans (3.9%). A space adjacent to the ulnar styloid was taken as the site for the first scan in these examinations. This problem may be recognized and corrected relatively easy. A significant correlation was found between BMC of the lower arm and BMC of the lumbar spine assessed with dual photon absorptiometry. However, the error of estimation of proximal BMC (SEE=20.0%) and distal BMC (SEE=19.4%) made these measurements of little value to predict BMC at the lumbar spine in individuals. The short term reproducibility in patients combined with the long term stability of the equipment in our clinical setting showed that SPA is a reliable technique to assess changes in bone mass at the lower arm of 4% between 2 measurements with a confidence level of 95%.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

References

  • Christiansen C, Rødbro P, Jensen H (1975) Bone mineral content in the forearm measured by photon absorptiometry. Principles and reliability. Scand J Clin Lab Invest 35:323–330

    Google Scholar 

  • Cummings SR, Black D (1986) Should perimenopausal women be screened for osteoporosis. Ann Intern Med 104:817–823

    Google Scholar 

  • Dunn WL, Kan SH, Wahner HW (1987) Errors in longitudinal measurements of bone mineral: effect of source strength in single and dual photon absorptiometry. J Nucl Med 28:1751–1757

    Google Scholar 

  • Geusens P, Dequeker, J, Verstraeten A, Nijs J (1986) Age-, sex-, and menopause-related changes of vertebral and peripheral bone: population study using dual and single photon absorptiometry and radiogrammetry. J Nucl Med 27:1540–1549

    Google Scholar 

  • Krølner B, Pors Nielsen S (1980) Measurement of bone mineral content (BMC) of the lumbar spine, I. Theory and application of a new two-dimensional dual-photon attenuation method. Scand J Clin Lab Invest 40:653–663

    Google Scholar 

  • Krølner B, Pors Nielsen S, Lund B, Lund BJ, Sørensen OH, Uhrenholdt A (1980) Measurement of bone mineral content (BMC) of the lumbar spine, II. Correlation between forearm BMC and lumbar spine BMC. Scand J Clin Lab Invvest 40:665–670

    Google Scholar 

  • Nilas L, Borg, J, Gotfredsen A, Christiansen C (1985) Comparison of single- and dual-photon absorptiometry in postmenopausal bone mineral loss. J Nucl Med 26:1257–1262

    Google Scholar 

  • Nilas L, Pødenphant J, Riis BJ, Gotfredsen A, Christiansen C (1987) Usefulness of regional bone measurements in patients with osteoporotic fractures of the spine and distal forearm. J Nucl Med 28:960–965

    Google Scholar 

  • Ott SM, Kilcoyne RF, Chesnut III CH (1986) Longitudinal changes in bone mass after one year as measured by different techniques in patients with osteoporosis. Calcif Tiss Int 39:133–139

    Google Scholar 

  • Mazess RB, Peppler WW, Chesney RW, Lange TA, Lindgren U, Smith Jr E (1984) Does bone measurement on the radius indicate skeletal status? Concise communication. J Nucl Med 25:281–288

    Google Scholar 

  • Riggs BL (1984) Treatment of osteoporosis with sodium fluoride: An appraisal. In: Peck WA (ed) Bone and Mineral Research, Annual 2: A Yearly survey of Developments in the Field of Bone and Mineral. Elsevier, New York, pp 366–393

    Google Scholar 

  • Riggs BL, Melton III LJ (1986) Involutional osteoporosis. N Engl J Med 314:1676–1686

    Google Scholar 

  • Riggs BL, Wahner HW, Dunn WL, Mazess RB, Offord KP, Melton III LJ (1981) Differential changes in bone mineral density of the appendicular and axial skeleton with aging. Relationship to osteoporosis. J Clin Invest 67:328–335

    Google Scholar 

  • Riggs BL, Wahner HW, Melton III LJ, Richelson LS, Judd HL, Offord KP (1986) Rates of bone loss in the appendicular and axial skeletons of women: evidence of substantial vertebral bone loss before menopause. J Clin Invest 1487–1491

  • Wahner HW, Dunn WL, Riggs BL (1984) Assessment of bone mineral. Part 2. J Nucl Med 25:1241–1253

    Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Rights and permissions

Reprints and permissions

About this article

Cite this article

Valkema, R., Blokland, J.A.K., Papapoulos, S.E. et al. The reproducibility of single photon absorptiometry in a clinical setting. Eur J Nucl Med 15, 269–273 (1989). https://doi.org/10.1007/BF00257547

Download citation

  • Received:

  • Issue Date:

  • DOI: https://doi.org/10.1007/BF00257547

Key words

Navigation