Evaluation of a miniature CdTe detector for monitoring left ventricular function

  • Published:
European Journal of Nuclear Medicine Aims and scope Submit manuscript

Abstract

A miniature CdTe probe interfaced to a microcomputer was used to measure left ventricular ejection fraction (LVEF) in 25 patients. LVEF obtained with the CdTe module, in the beat-to-beat mode, or the integrated gated mode agreed well with LVEF obtained with a gamma camera (r=0.80; r=0.82 respectively). Similarly, LVEF by CdTe probe agreed with LVEF obtained by gated equilibrium studies performed with a computerized NaI probe. The CdTe probe can provide comparable measurement of LVEF at a fraction of the cost of a cameracomputer system and, being small and lightweight, the CdTe probe is adaptable for monitoring patients in intensive care facilities.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

References

  1. Berger H, Hoffer P, Steidley F (1980) Miniaturized cardiac module for continuous monitoring of left ventricular performance using a semiconductor gamma detector: development of a protoype and initial clinical evaluation. (Abst) Am J Cardiol 45:408

    Google Scholar 

  2. Berger HJ, Davies PA, Bastford WP (1981) Beat to beat left ventricular performance assessed from the equilibrium cardiac blood pool using a computerized nuclear probe. Circulation 63:133–142

    Google Scholar 

  3. Hoffer P, Berger HJ, and Steidley F (1981) A miniature cadmium telluride module for continuous monitoring of left ventricular function. Radiology 138:477–481

    Google Scholar 

  4. Wagner HN Jr, Wake R, Nickoloff E (1976) The nuclear stethoscoped: a simple device for generation of left ventricular volume curves. Am J Cardiol 38:747–750

    Google Scholar 

  5. Wagner HN Jr, Rigo P, Baxter RH (1979) Monitoring ventricular function at rest and during excerise with a non imaging nuclear detector. Am J Cardiol 43:975–979

    Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Additional information

This work supported in part by USPHS Grant #GM 10548

Rights and permissions

Reprints and permissions

About this article

Cite this article

Harrison, K.S., Liu, X., Han, ST. et al. Evaluation of a miniature CdTe detector for monitoring left ventricular function. Eur J Nucl Med 7, 204–206 (1982). https://doi.org/10.1007/BF00256464

Download citation

  • Received:

  • Issue Date:

  • DOI: https://doi.org/10.1007/BF00256464

Keywords

Navigation