Summary
Pulsed field gradients used together with trim pulses may cause artifacts in NMR spectra that originate from partial refocusing of dephased magnetization. These effects can reduce the efficiency of solvent suppression. The duration of the trim and PFG pulses should be in the range in which refocusing is negligible. Background gradients due to bad shimming also interfere with the B0 field gradient pulses, producing gradient-recalled echoes that reduce the receiver gain for NMR experiments. The shim settings can be optimized using simple experiments, as described in this paper. Eddy currents that cannot be completely compensated by adjustments of preemphasis induce phase shifts in NMR signals. The decay constants for a given spectrometer setup can easily be measured. If the experiment does not allow for proper compensation delays, the phase of the pulses must be adjusted to compensate for these phase shifts.
Similar content being viewed by others
References
Bax, A. and Davis, D.G. (1985) J. Magn. Reson., 65, 355–360.
Bax, A. and Pochapsky, S.S. (1992) J. Magn. Reson., 99, 638–643.
Bowtell, R. and Peters, A. (1995) J. Magn. Reson. Ser. A, 115, 55–59.
Crozier, S., Eccles, C.D., Beckey, F.A., Field, J. and Doddrell, D.M. (1992) J. Magn. Reson., 97, 661–665.
Crozier, S., Beckey, F.A., Eccles, C.D., Field, J. and Doddrell, D.M. (1994) J. Magn. Reson. Ser. B, 103, 115–119.
Davis, A.L., Keeler, J., Laue, E.D. and Moskau, D. (1992) J. Magn. Reson., 98, 207–216.
Dötsch, V., Wider, G. and Wüthrich, K. (1994) J. Magn. Reson. Ser. A. 109, 263–264.
Eccles, C.D., Crozier, S., Westphal, M. and Doddrell, D.M. (1993) J. Magn. Reson. Ser. A, 103, 135–141.
Ernst, R.R., Bodenhausen, G. and Wokaun, A. (1987) Principles of Nuclear Magnetic Resonance in One and Two Dimensions, Oxford University Press, Oxford, U.K.
Hull, W.E. (1994) In Two-dimensional NMR Spectroscopy (Eds., Croasmun, W.R. and Carlson, R.M.K.), VCH, New York, NY, pp. 186–193.
Hurd, R.E. and John, B.K. (1991) J. Magn. Reson., 91, 648–653.
Karlicek Jr., R.F. and Lowe, I.J. (1980) J. Magn. Reson., 37, 75–91.
Keeler, J., Clowes, R.T., Davis, A.L. and Laue, E.D. (1994) Methods Enzymol., 239, 145–206.
Majors, P.D., Blackley, J.L., Altobelli, S.A., Caprihan, A. and Fukushima, E. (1990) J. Magn. Reson., 87, 548–553.
Moonen, C.T.W. and Van, Zijl, P.C.M. (1990) J. Magn. Reson., 88, 28–41.
Moonen, C.T.W., Sobering, G., Van, Zijl, P.C.M., Gillen, J., von, Kienlin, M. and Bizzi, A. (1992) J. Magn. Reson., 98, 556–575.
Redfield, A.G. and Gupta, R.K. (1971) Adv. Magn. Reson., 5, 81–113.
Ross, A., Czisch, M., Cieslar, C. and Holak, T.A. (1993) J. Biomol. NMR, 3, 215–224.
Ruiz-Cabello, J., Vuister, G.W., Moonen, C.T.W. Van, Gelderen, P., Cohen, J.S. and Van, Zijl, P.C.M. (1992) J. Magn. Reson., 100, 282–302.
Sattler, M., Schwendinger, M.G., Schleucher, J. and Griesinger, C. (1995) J. Biomol. NMR, 6, 11–22.
Shaka, A.J., Barker, P.B. and Freeman, R. (1985) J. Magn. Reson., 64, 547–552.
Stonehouse, J., Clowes, R.T., Shaw, G.L., Keeler, J. and Laue, E.D. (1995) J. Biomol. NMR, 5, 226–232.
Tyburn, J.-M., Brereton, I.M. and Doddrell, D.M. (1992) J. Magn. Reson., 97, 305–312.
Van, Zijl, P.C.M., Chesnick, A.S., DesPres, D., Moonen, C.T.W., Ruiz-Cabello, J. and Van, Gelderen, P. (1993) Magn. Reson. Med., 30, 544–551.
Van, Zijl, P.C.M., Sukumar, S., O'Neil Johnson, M., Webb, P. and Hurd, R.E. (1994) J. Magn. Reson. Ser. A, 111, 203–207.
Wider, G., Dötsch, V. and Wüthrick, K. (1994) J. Magn. Reson. Ser. A, 108, 255–258.
Yamazaki, T., Lee, W., Arrowsmith, C.H., Muhandiram, D.R. and Kay, L.E. (1994) J. Am. Chem. Soc., 116, 11655–11666.
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Czisch, M., Ross, A., Cieslar, C. et al. Some practical aspects of B0 gradient pulses. J Biomol NMR 7, 121–130 (1996). https://doi.org/10.1007/BF00203822
Received:
Accepted:
Issue Date:
DOI: https://doi.org/10.1007/BF00203822