Summary
We have followed up 227 total hip replacements using a cementless minimadreporic prosthesis for an average of 9 1/2 years. Special attention was paid to the acetabular component, of which 2 types were used; 77 incorporated screw rings and 134 were stabilised by spikes. The screw prostheses were evaluated at an average of 8 1/2 years from operation; one third showed radiological evidence of migration, and revision because of this was required in 12.6% of cases. The spiked cups were followed up for an average of 10 years; 14.8% showed radiological evidence of migration and 2.3% needed revision. We observed no radiological features to explain this difference. The quality of the bone and the mechanical characteristics of the bone/implant contrast may influence the result. We feel that an uncemented spiked acetabular implants is a worthwhile concept.
Résumé
Sur 277 prothèses totales de hanche sans ciment minimadréporiques, 223 ont pu être étudiées avec un recul moyen de 9 ans et 6 mois comportant, sur le plan cotyloïdien, 134 cotyles à pointes et 77 anneaux vissés; l'étude séparée du devenir de l'implant cotyloïdien vise à préciser sa part de responsabilité dans les échecs et à proposer un concept de cotyle non cimenté. Au recul moyen étudié (8 ans 6 mois) 33,3% des anneaux vissés présentent une migration radiologique responsable d'un taux de réintervention de 12,64%. En comparaison et avec un recul moyen plus long (10 ans) le cotyle à pointes ne présente que 14,8% de signes radiologiques de migration avec un taux de reprise de 2,27%. Aucun des facteurs radiocliniques envisagés ne permet d'expliquer cette différence. Celle-ci apparaît liée aux propriétés biomécaniques du cotyle osseux et au comportement mécanique du couple composite osimplant. Cette étude permet aux auteurs de proposer le concept du cotyle à pointes comme une réponse possible à la stabilité à long terme des implants cotyloïdiens non cimentés.
Similar content being viewed by others
Bibliographie
Anderson, Freeman M, Swanson (1972) Loosening of the cemented actabular cup in total hip replacement. J Bone Joint Surg [Br] 54:590–599
Ayerza M, Pierson R, Sheinkop M, Rosenberg A, Landon G, Galente J (1988) A clinical review of a conical screw-in acetabular design. Orthop Trans 12:662
Baldursson H, Hansson LI, Olsson TH, Selvik G (1980) Migration of acetabular socket after total hip replacement determined by roentgen stereophotogrammetry. Acta Orthop Scand 51:535–540
Bancel P (1982) Un composant acétabulaire sans ciment d'arthroplastie totale de hanche: l'anneau vissé. Etude expérimentale et résultats des 376 premiers cas. Thèse médecine, St Antoine, Paris
Bigard O (1990) La prothèse universelle minimadréporique de D Schmitt. Evaluation radioclinique. Thèse médecine, Nancy
Bobyn J, Engh C, Glassman A (1988) Radiography and histology of the threaded acetabular implant. J Bone Joint Surg [Br] 70:302–304
Bowden FP, Tabor D (1954) The friction and lubrication of solids. Clarendon Press, Oxford
Braun E (1985) La prothèse universelle minimadréporique de D Schmitt. Thèse médecine, Nancy
Duparc J, Massin P (1991) Prothèses totales de hanche avec des anneaux vissés. Résultats à 2 ans et plus d'une série de 198 prothèses consécutives. Rev Chir Orthop 77:221–231
Engh C, Griffin W, Marx C (1990) Cementless acetabular components: a comparison of porous coated and threaded designs. J Bone Joint Surg [Br] 72:53–59
Favard L, Bonnard C, Burdin P, Rosset P, Bertrand P, Queguiner A (1992) Etude expérimentale de la fixation primaire des cotyles vissés. Influence du taraudage préalable. Rev Chir Orthop 78:223–227
Huiskes R (1987) Finite element analysis of acetabular reconstruction: non cemented threaded cups. Acta Orthop Scand 58:620–625
Kody M, Kabo J, Markolf K, Delaunay C, Amstutz H (1988) Initial fixation strength of screw ring acetabular components. Orthop Trans 12:465–466
Kody M, Kabo JM, Markolf KL, Dorey FJ, Delaunay C, Amstutz HC (1990) Fixation mécanique initiale des anneaux vissés. Communication à la 64e réunion annuelle de la SOFCOT. Rec Chir Orthop 76 [suppl 1]: 83–84
Lord G, Bancel P (1983) The madreporic cementless arthroplasty: new experimental data and a seven year clinical follow-up. Clin Orthop 176:164–173
Massin P, Schmidt L, Engh CA (1989) Evaluation of cementeless acetabular components migration. J Arthroplasty 4:245–251
Mjoberg B, Hansson LI, Servik G (1984) Instability of total hip prostheses at rational stress. A roentgen stereophotogrammettric study. Acta Orthop Scand 55:504–506
Mjoberg B, Servik G, Hansson LI, Rosenquist R, Onnerfalt R (1986) Mechanical loosening of total hip prostheses. A radiographic and roentgen stereophotogrammettric study. J Bone Joint Surg [Br] 68:770–774
Morscher E, Masar Z (1988) Development and first experience with an uncemented press-fit cup. Clin Orthop 232:96–103
Morscher E, Moulin P (1985) Le cotyle de hanche non scellé dans l'arthroplastie totale de hanche. Acta Orthop Belg 51:345–357
Schmitt D, Braun E (1987) Conception technique et surveillance radiologique de la prothèse totale de hanche non cimentée minimadréporique de D Schmitt. Bilan de 275 arthroplasties avec un recul supérieur à 6 ans pour les cas les plus anciens. Rev Chir Orthop 73:219–230
Snorasson F, Karrholm J (1990) Primary migration of fully-threaded acetabular prostheses. J Bone Joint Surg [Br] 72:647–651
Teinturier P, Terver S, Jaramillo CV, Besse JP (1984) La biomécanique du cotyle. Rev Chir Orthop 70:41–46
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Schmitt, D., Bigard, O., Bresler, F. et al. Résultats à long terme de deux composants cotyloïdiens non cimentés dans les prothèses totales de hanche. International Orthopaedics 17, 357–364 (1993). https://doi.org/10.1007/BF00180454
Issue Date: