Abstract
The accuracy of white light and laser interferometers in predicting visual acuity after YAG laser capsulotomy was compared. 42 eyes of 41 patients were tested with both a Haag-Streit (Lotmar) white light interferometer and a Rodenstock laser interferometer, and 14 were also tested with a Site white light machine. The laser interferometer predicted a final visual acuity to within one line of that actually achieved in 93%, and to within two lines in 98%, whereas for the Haag-Streit these figures were 64% and 81%, and for the Site 77% and 92%. In patients with poor initial visual acuity, the difference in the relative performance of the two instruments was increased further. When interferometry was repeated after capsulotomy, the values obtained with all instruments agreed closely with Suellen acuity. This difference in predictive accuracy shows that capsular thickening causes a greater degree of optical degradation of the image produced by a white light interferometer than occurs when a laser interferometer is employed.
Similar content being viewed by others
References
Bernth-Petersen P, Naeser K (1982) Clinical evaluation of the Lotmar visuometer for macula testing in cataract patients. Acta Ophthalmol (Copenh) 60:525–532
Campbell FW, Green D (1965) Optical and retinal factors affecting visual resolution. J Physiol (Lond) 181:576–593
Dardenne M, Gerten G, Kokkas K, Kermani O (1989) Retrospective study of retinal detachment following neodymium: YAG laser posterior capsulotomy. J Cataract Refract Surg 15:676–680
Fastenberg DM, Schwartz PL, Lim HZ (1984) Retinal detachment following neodymium: YAG laser capsulomy. Am J Ophthalmol 97:288–291
Faulkner HW (1982) The laser interferometer in predicting efficacy of secondary posterior capsulotomy. Am Intraocul Implant Soc J 8:136–140
Faulkner HW (1983) Laser interferometric prediction of postoperative visual acuity in patients with cataracts. Am J Ophthalmol 95:626–636
Faulkner HW (1986) Macula function testing through opacities. In: Focal points 1986: Clinical modules for ophthalmologists, vol 4, module 2. Am Acad Ophthalmol
Ficker LA, Vickers S, Capon MRC, Mellerio J, Cooling RJ (1987) Retinal detachment following neodymium: YAG laser capsulotomy. Eye 1:86–89
Goldmann H, Chrenkova A, Cornaro S (1980) Retinal visual acuity in cataractous eyes. Arch Ophthalmol 98:1778–1781
Green DG (1970) Testing the vision of cataract patients by means of laser generated interference fringes. Science 168:1240–1242
Halliday BL, Ross JE (1983) Comparison of two interferometers for predicting visual acuity in patients with cataract. Br J Ophthalmol 67:273–277
Hanna IT, Sigurdsson H, Baines PS, Roxburgh STD (1989) The role of white light interferometry in predicting visual acuity following posterior capsulotomy. Eye 3:468–471
Jaffe NS, Clayman HM, Jaffe MS (1982) Cystoid macular edema after intracapsular and extracapsular cataract extraction with and without an intraocular lens. J Am Acad Ophthalmol 89:25–29
Lang TA, Lindstrom RL (1985) Efficacy of laser interferometry in predicting visual result of YAG laser posterior capsulotomy. Am Intraocular Implant See J 11:367–371
Leff SR, Welch JC, Tasman W (1987) Rhegmatogenous retinal detachment after YAG laser capsulotomy. Ophthalmology 94:1222–1224
Lindstrom RL, Harris WS (1980) Management of the posterior capsule following posterior chamber lens implantation. Am Intraocular Implant See J 6:255–258
Lotmar W (1980) Apparatus for the measurement of retinal visual acuity by moire fringes. Invest Ophthalmol Vis Sci 19:393–400
Miami Study Group (1979) Cystoid macular edema in aphakic and pseudophakic eyes. Am J Ophthalmol 88:45–48
Moisseiev J, Bartov E, Schochat A, Blumenthal M (1989) Longterm study of the prevalence of capsular opacification following extracapsular cataract extraction. J Cataract Refract Surg 15:531–533
Percival SPB, Anand V, Das SK (1983) Prevalence of aphakic retinal detachment. Br J Ophthalmol 67:43–45
Selenow A, Ciuffreda KJ, Mozlin R, Rumpf D (1986) Prognostic value of laser interferometric visual acuity in amblyopia therapy. Invest Ophthalmol Vis Sci 27:273–277
Shah GR, Gills JP, Durham DG, Ausmus WH (1986) Three thousand YAG lasers in posterior capsulotomies: an analysis of complications and comparison to polishing and surgical discission. Ophthalmic Surg 17:473–477
Spurney RC, Zaldivar R, Belcher CD, Simmons RJ (1986) Instruments for predicting visual acuity. Arch Ophthalmol 104:196–200
Terry AC, Stark WJ, Maumenee E, Fagadau W (1988) Neodymium: YAG laser for posterior capsulotomy. Am J Ophthalmol 96:716–720
Wilhelmus KR, Emery JM (1980) Posterior capsule opacification following phakoemulsification. Ophthalmic Surg 11:262–267
Author information
Authors and Affiliations
Additional information
Offprint requests to: N. Strong
Rights and permissions
About this article
Cite this article
Strong, N. Interferometer assessment of potential visual acuity before YAG capsulotomy: relative performance of three instruments. Graefe's Arch Clin Exp Ophthalmol 230, 42–46 (1992). https://doi.org/10.1007/BF00166761
Received:
Accepted:
Issue Date:
DOI: https://doi.org/10.1007/BF00166761