Research on magnetic center measurement of quadrupole and sextupole using vibrating wire alignment technique in HEPS-TF

  • Lei WuEmail author
  • Xiao-long Wang
  • Chun-hua Li
  • Hua-min Qu
  • Zi-hao Wang
  • Min-xian Li
  • Ling-ling Gong
Original Paper
  • 116 Downloads

Abstract

Purpose

In order to meet the extremely low emittance requirement, the magnets in the storage ring of high-energy photon source (HEPS) need to have a stable support and precise positioning. In HEPS-TF, the key and difficult technologies of HEPS should be researched and developed. Vibrating wire alignment technique is one important project of HEPS-TF. It can be used to pre-align the quadrupoles and sextupoles on one girder with high precision. A vibrating wire measurement system was set up and tested to verify the magnetic center measurement precision and the magnet adjustment error.

Methods

There are one sextupole and one quadrupole installed on a multipole girder. Vibrating wire is stretched through mechanical center of the magnet apertures and supported by the test benches on the two sides. A single conducting wire is stretched through the magnet aperture and electrified by alternating current. The wire will vibrate for a period of Lorentz force. By matching the current frequency to one mode of natural frequency of the wire, the vibrating amplitude will be enhanced. And by measuring the vibrating amplitude, the magnetic field at the wire position can be got. Moving the wire across the magnet aperture in the transversal or vertical direction, the distribution of magnetic field and magnetic center position can be measured. According to the magnetic center position error to adjust the magnet. Measure the magnetic center of all magnets installed on the multipole girder one by one, and adjust their magnetic center to a line.

Results

The magnetic center measurement precision is better than ±3 μm, and the magnet adjustment error is less than 6 μm.

Conclusion

The vibrating wire system design and a series of magnetic center measurement experiments have gained good achievements. It has been proved the vibrating wire system is designed reasonable and using the vibrating wire to align the magnets installed on a multipole girder is feasible and can reach a high precision.

Keywords

Vibrating wire Magnetic center Measurement precision Magnet adjustment 

PACS

29.20.db 

Notes

Acknowledgements

This work was supported by High Energy Photon Source Test Facility (HEPS-TF). We sincerely thank the help of Professor Jain Animesh from Brookhaven National Laboratory.

References

  1. 1.
    X.M. Jiang, J.Q. Wang, Q. Qin et al., The Chinese High-Energy Photon Source and its R&D project. Synchrotron Radiat. News 27(6), 27–31 (2014).  https://doi.org/10.1080/08940886.2014.970938 CrossRefGoogle Scholar
  2. 2.
    A. Temnykh, Vibrating wire field-measuring technique. Nucl. Instrum. Methods Phys. Res. A 399, 185–194 (1997).  https://doi.org/10.1016/S0168-9002(97)00972-8 ADSCrossRefGoogle Scholar
  3. 3.
    Temnykh A, The magnetic center finding using vibrating wire technique, in 11th International Magnetic Measurement Workshop, 1999Google Scholar
  4. 4.
    X.L. Wang, L. Dong, L. Wu et al., Analysis and experimental concepts of the vibrating wire alignment technique. Chin. Phys. C 38(11), 117010 (2014).  https://doi.org/10.1088/1674-1137/38/11/117010 CrossRefGoogle Scholar
  5. 5.
    L. Wu, X.L. Wang, C.H. Li et al., Theory and research overview of vibrating wire technique. High Power Laser Part. Beams 25(10), 2480–2486 (2013).  https://doi.org/10.3788/HPLPB20132510.2479 CrossRefGoogle Scholar
  6. 6.
    L. Wu, X.L. Wang, C.H. Li et al., Design of vibrating wire alignment system for HEPS-TF. High Power Laser Part. Beams 27(9), 095102 (2014).  https://doi.org/10.11884/hplpb201527.095102 CrossRefGoogle Scholar
  7. 7.
    A. Jain, Precision alignment of multipoles on a girder for NSLS-II, in 17th International Magnetic Measurement Workshop, Barcelona, 2011Google Scholar
  8. 8.
    A. Jain, Vibrating wire R&D for magnet alignment, in NSLS-II Magnet Workshop, 2012Google Scholar
  9. 9.
    A. Jain, Results from vibrating wire R&D for alignment of multipoles in NSLS-II, in 16th International Magnetic Measurement Workshop, Zurich, 2009Google Scholar

Copyright information

© Institute of High Energy Physics, Chinese Academy of Sciences; Nuclear Electronics and Nuclear Detection Society and Springer Nature Singapore Pte Ltd. 2018

Authors and Affiliations

  1. 1.Institute of High Energy PhysicsChinese Academy of SciencesBeijingChina
  2. 2.Dongguan Institute of Neutron SciencesDongguanChina
  3. 3.University of Chinese Academy of SciencesBeijingChina

Personalised recommendations