Advertisement

Performance of Tc-99m Generators in Terms of Molybdenum Contents Measured in NORIN Nawabshah, Pakistan (5 years’ Experience)

  • Muhammad WaqarEmail author
  • Muhammad Shahban
  • Quratulain Soomro
  • Muhammad Ejaz Khan
Research Paper
  • 18 Downloads

Abstract

Technetium-99m is extensively used radiopharmaceutical in nuclear medicine centres throughout the world because of its optimum half-life and photon energy. 99mTc is obtained from 99mTc-99Mo generator. 99Mo contents may appear in 99mTc elutes due to certain reasons. These contents are needed to be checked before injecting it to the patient. In Pakistan, these generators are manufactured in Pakistan Institute of Nuclear Science and Technology (PINSTECH), Islamabad, who names its generators as PAKGEN. 99mTc is eluted on site from generator and used with pharmaceutical kit. Eluted 99mTc may contain some impurities which need to be contained within predefined allowed limits. One of the impurities is molybdenum. For molybdenum, there are two protocols being used worldwide. European pharmacopeia declares the allowed activity of Molly in 99mTc elute to be less than 1 µCi of 99Mo for every mCi of 99mTc, while American pharmacopeia defines it to be less than 0.15 µCi/mCi. PINSTECH is using European standards for its PAKGEN generators. NORIN is using 99mTc-99Mo generators since 2012, purchased from PINSTECH Islamabad. A total of 216 generators were purchased till date. Current study is an assessment of Molly assay tests performed on the first elution of every generators before its administration to patients for diagnostic procedures. Results of these tests are compared with both European and American pharmacopeia standards and found out to be well within both limits. Only 1.4% of total 216 generators were found to be out of the limit of 0.15 µCi/mCi (being practiced in our medical centre), while no one exceeded the limit of 1 µCi/mCi (being implemented by the manufacturer of generators). It is concluded that PAKGEN generators are reliable source of technetium-99m in terms of molybdenum contents.

Keywords

PAKGEN 99mTc 99Mo-99mTc generator MBT Molly assay 

Notes

Acknowledgements

The authors are grateful to Mr. Syed Raza Haider (Ex-RPO), Mr. Adeel Ahmed, Mr. Muhib Ali and Ms. AndleepNaz (Technicians) NORIN Nawabshah, for their assistance and help for collection data and analysis of data.

References

  1. Ann. ICRP 23 (1993) ICRP publication 67-age-dependent doses to members of the public from intake of radionuclides—part 2 ingestion dose coefficientsGoogle Scholar
  2. Ahmad M, Pervez S, Hussain S, Mirza JA, Khan MM, Asif M, Siddique MU, Khalid U, Khan B, Muhammad Khalid (2012) Evaluation of Pakgen 99mTc generators loaded with indigenous fission 99Mo. Radiochim Acta 100(10):792–800Google Scholar
  3. Capintec (2010) CRC-25R-owner’s manual, stock no. 9250-0122Google Scholar
  4. Dantas BM, Dantas ALA, Marques FLN, Bertelli L (2005) Determination of 99Mo contamination in a nuclear medicine patient submitted to a diagnostic procedure with 99mTc. Braz Arch Biol Technol 48(SPECIL ISS 2):215–220CrossRefGoogle Scholar
  5. Khan AK, Khan G, Zaman MU, Fatima N, Naqvi SM, Khan S (2012) Assay for molybdenum-99 breakthrough in PAKGEN, PINSTECH generators: a clinical audit at Akuh, Karachi. Pak J Radiol 22(June):44–45Google Scholar
  6. Memon SA, Laghari NA, Qureshi ST, Mehdi F (2014) Assessment of molybdenum breakthrough levels in molybdenum-99/technetium-99m generators: one year experience at NIMRA Jamshoro Pakistan. Int J Radiat Res 12(4):6–9Google Scholar
  7. Momennezhad M, Zakavi SR, Sadeghi R (2010) Determination of 99 Mo contamination in 99m Tc elute obtained from 99 Mo/99m Tc- generator. Iran J Radiat Res 8(1):31–35Google Scholar
  8. NRC (2007) Code of federal regulations, part 35, section 35.204 (10 CFR 35.204) permissible molybdenum-99, strontium-82, and strontium-85 concentrationsGoogle Scholar
  9. Reese IC, Mishkin FS (1968) A simple method of reducing the molybdenum content of the eluate from a molybdenum 99-technetium 99m generator. Am J Roentgenol Radium Ther Nucl Med 103(4):896CrossRefGoogle Scholar
  10. Shah SSSA, Khan HA, Khan SU (2009) Comparison of molybdenum 99-technitium 99m generator eluates from two manufacturers and its purification for nuclear medicine imaging. Pak J Radiol 19(2):46–49Google Scholar

Copyright information

© Shiraz University 2018

Authors and Affiliations

  • Muhammad Waqar
    • 1
    Email author
  • Muhammad Shahban
    • 1
  • Quratulain Soomro
    • 1
  • Muhammad Ejaz Khan
    • 1
  1. 1.Nuclear Medicine Oncology and Radiotherapy Institute Nawabshah (NORIN)Shaheed BenazirabadPakistan