Introduction

Bacterial cellulose (BC) represents an interesting emerging biocompatible nanomaterial which can be used in a wide range of biomedical applications, such as production of artificial blood vessels or microvessels and artificial skin [1]. Microbial cellulose is biosynthesized by the Gram-negative bacterium named Gluconacetobacter xylinus, formerly Acetobacter xylinum [2]. Bacterial cellulose is susceptible to modifications that significantly improve its properties and/or functionality. As a result, new BC-based nanocomposites are obtained, for example, BC/chitosan, BC/collagen, BC/gelatin, or BC/fibroin [1]. These nanocomposites are a combination of the good properties of BC with specific features of other materials for specific applications [3]. Chitosan modification is an example of improving the properties of bacterial cellulose. Chitosan-based biomaterials are considered antimicrobial, functional, renewable, nontoxic, biocompatible, bioabsorbable, and biodegradable biopolymer agents [4]. Many factors and qualities of chitosan derivatives make the BC/chitosan nanoparticles adequate materials for implantation. Chitosan is also characterized by good mechanical properties in wet state, high ability to maintain moisture, high bioaffinity, bacteriostatic, and bactericidal activity against bacteria Escherichia coli G(−) or Staphylococcus aureus G(+). Moreover, BC/chitosan molecules show susceptibility to specific hydrolytic enzymes, for example, lysozyme. During the enzymatic degradation of chitosan, bioactive mono- and oligosaccharides are being released, which stimulate angiogenesis and tissue regeneration [5]. These types of nanoparticles are promising to advance strategies to treat vascular disease. It has been proven that MBC provides a good scaffold for the host cells and promotes the regeneration of tissues by the body. It has been found that due to its mechanical and biological properties, bacterial cellulose modified with chitosan is a promising material for replacing small-diameter vessels grafts. In recent years, bioengineering of artificial blood vessels is focused mainly on discovery and development of new artificial vascular substitutes, alternative materials that integrate with the patient’s native tissue to restore physiological function [6]. The main challenge for tissue engineering is to develop new biologically active small-diameter grafts. Currently expanded-polytetrafluoroethylene (ePTFE; synthetic fluorocarbon polymer), polyethylene terephthalate (Dacron; polymer containing ester functional group), and polyurethanes (formed through a reaction between isocyanates and polyol) are the most commonly used synthetic graft materials for artificial blood vessels [6]. However, these commercial prosthesis-materials are unsuitable for microsurgical requirements, because they have larger diameter than 3 mm. Therefore, they cannot be used in microvessels surgeries. Tissue-engineered blood vessels (TEBVs) represent an innovative approach for overcoming reconstructive problems associated with vascular diseases by providing small-caliber vascular grafts. Moreover they confirm that BC has high mechanical strength, promotes in situ vascular tissue regeneration, can be shaped into three-dimensional structures, and does not require pre-treatment in the form of cell seeding. The ideal vascular prosthesis should be resistant to thrombosis, inflammation, and excessive thickening of the intima. Numerous publications suggest that BC seems to be the best candidate among those possible to construct novel vascular prosthesis and substitutes of small blood vessels [7, 8].

In this study, a novel vascular prostheses made of bacterial nanocellulose modified with chitosan (MBC) will be tested for their potential toxicity and adverse sensitization effects. The toxicology study of nanocellulose and nanocellulose-based biocomposites is still restricted at a very preliminary stage (mainly on the level of cytotoxicity) [9]. Although previous studies have not shown the toxicity of nanocellulose, we want to check in vivo whether our product will be safe for patients in the future. Long-term stability and suitability to replace small-caliber blood vessels without significant thrombogenicity must be investigated. Study of compliance between the graft and the surrounding native vessels and tissues, together with the postoperative complications, is now warranted and is the subject of current study.

Experimental Section—Materials

In Vitro Testing

The assessment of cytotoxic activity in vitro was conducted in the Cell Culture Laboratory, Department of Histology and Embryology of the Wrocław Medical University. The study was conducted on a reference cell line, mouse fibroblasts 3T3/Balb obtained from the Tissue Bank, Institute of Immunology and Experimental Therapy of the Polish Academy of Sciences in Wrocław. The cells were stored in liquid nitrogen, with the addition of about 10% DMSO. After thawing, the cells were passed through twice with a solution of 0.05% trypsin + 0.02% EDTA in PBS, pH 7.2 (provided by SIGMA, USA).

The following composition was used for fibroblast cell culture: MEM Eagle (SIGMA, USA); 10 % FCS (SIGMA, USA); glutamine, 1 mM/ml (SIGMA, USA); penicillin, 100 μg/ml; streptomycin, 100 μg/ml.

The cells were cultured under the conditions recommended for tissue culture in an atmosphere of 5% CO2, at 37°C, with constant chamber moisturization.

In vivo Testing

The present study was conducted in accordance with the prior consent of the No. 9 Local Ethics Committee for animal experiments in Lodz and the Ethics Committee of the Medical University of Lodz, Poland; Resolution No. 49/LB 483/2009 on September 21, 2009. The work was supported by Development of Biological Vascular Prostheses of Modified Bacterial Cellulose Research Project grant (no. N205 068 31/3102); patent PL 190961 “Method of obtaining modified bacterial cellulose” [10, 12, 15].

Animals

Forty rats (outbred stock Han Wistar Imp:WIST rats)—healthy, both sexes, aged approximately 2 to 3 months, body mass about 300 g

Three New Zealand albino rabbits (outbred stock Imp:BN)—healthy, both sexes, approximately 3 months old, body mass 240–350 g

Fifteen guinea pigs (outbred stock Dunkin-Hartley Imp:D-H)—healthy, both sexes, aged approximately 2 to 3 months, body mass 240–350 g

All laboratory animals were provided by the Institute of Occupational Medicine, where they were bred. The animals were housed as one animal per cage at 19–21.5°C, 55–60% relative humidity, and /12h:12h/ light/dark cycle. Animals had access to standard feed and water ad libitum.

Reagents

For the sensitization test Freund’s adjuvant, sodium dodecyl sulfate and cottonseed oil were supplied by Sigma-Aldrich Corporation (Poznan, Poland), while saline solution (0.9% NaCl) was provided by Baxter Manufacturing Sp. z o.o. (Lublin, Poland).

For the sensitization, aqua pro injection was provided by Baxter Manufacturing Sp. z o.o. (Lublin, Poland) and sterile gauze compress 17 filiform was supplied by TZMO SA (Torun, Poland). To validate the skin sensitization assessment, 85% α-hexyl cinnamaldehyde was supplied by Sigma-Aldrich Corporation (Poznan, Poland).

Tested Materials

The research material was modified bacterial cellulose in tubular form (BCTube). Samples for the study were provided by the Institute of Biopolymers and Chemical Fibres, Lodz, Poland. The BC was synthesized using an acetic strain of bacteria, Acetobacter sp. (CCM 2360), from the Czech Collection of Microorganisms, Masaryk University, Brno, Czech Republic. Bacterial cellulose was modified with chitosan at the stage of biosynthesis—by adding chitosan oligomers to the culture medium. As presented previously [5], chitosan chains are physically incorporated into the cellulose network. Modified Hestrin-Schramm medium was used, composed of the following (on 1000 cm3): glucose 20.0 g, yeast extract 5.0 g, soy peptone 5.0 g. disodium phosphate 2.7g, acetic acid 1.2 g, ethanol 20.0 cm3, and chitosan oligomers (ChitoOligo-100, Aminogen, Korea) 2.0 g. Modificator-ChitoOligo-100 is a water-soluble oligomer of β-1,4-linked D-glucosamine, produced from crab chitosan by an enzymatic process using a chitosanase, developed by the Aminogen company (Korean Patent No. 0227040). Product characteristics are as follows: molecular weight >5kDa, degree of acetylation 92%, molecular weight distribution of 1 to 1.5, an intrinsic viscosity of 0.020 to 0.250 g/D-sec, a moisture content >1%, and an inorganic content of 0 to 1%. Samples of the BC tubes were analyzed by Fourier transform infrared spectroscopy (FTIR). BC tubes were obtained by biosynthesis for 7 days at 30 °C in an incubator (Fig. 1.1). Bacterial cellulose in tubular form, BC Tube (Fig. 1.2), was obtained by biosynthesis in single silicone tubes. The silicone tube (internal diameter 10 mm, wall thickness 1.4 mm) was filled with HS culture medium modified with chitosan oligomers; the ends of the tube were tightly closed with plastic plugs. The external diameter of the BC tube depended on the internal diameter of the silicone tube. For testing biological properties, BC tubes with an external diameter of 10 mm and wall thickness of 1.7–1.8 mm were used (Fig. 1.3) On Fig. 2a, SEM morphology of internal wall surface of BCTube was shown. On Fig. 2b–c, 3 cross-section and layered growth of bacterial cellulose were presented.

Fig. 1
figure 1

Biosynthesis and morphology of BC tubes. Bacterial cellulose in tubular form was obtained by biosynthesis in silicone tubes (1.1). BC tubes used for testing with an external diameter of 10 mm and wall thickness of 1.7–1.8 mm (1.2 and 1.3).

Fig. 2
figure 2

SEM image presenting the structure of BC tube produced by the static method; (2.1) internal wall surface of BCTube, magnification ×50,000; (2.2) cross-section and layered growth of BCTube, magnification ×20,000; (2.3) cross-section and layered growth of BC tube, magnification ×40,000

The obtained BC tubes were washed with distilled water until the complete removal of culture medium components. Next, the tubes were soaked in 1% NaOH, autoclaved at 121 °C for 15 min, and washed with distilled water until reaching neutral pH and conductivity <20 μS.

Hemocompatibility of BC Tubes

The hemocompatibility tests of a modified bacterial cellulose blood vessel prosthesis were prepared on human blood O Rh+ taken into CPD preservative (sodium citrate, citric acid, glucose dihydrogen phosphate). The scope of work included the following:

1. Hemolytic activity study:

(a) After temporary contact with blood (plasma Hb determination)

(b) After temporary contact of the extract with the red blood cell suspension (determination of the hemolysis rate)

2. Tests of the plasma clotting system after temporary contact of full blood with the prosthesis:

(a) Determination of partial thromboplastin time after activation

(b) The indication of the prothrombin time

(c) Indication of thrombin time

(d) Determination of fibrinogen

Experimental Section—Methods

In vitro Testing

Mouse fibroblast cells were cultured on NUNC 12-well plates in density of 0.5·106. After 24 h, the cells adhered to the plate’s surface and divided, covering about 50% of the plate surface. After this time, the culture medium was removed and a new medium was added to each well. Samples of 5×5 mm in size were applied to the cell cultures (the samples covered about 10% of the culture surface) and plates were incubated at 37°C, in 5% CO2 atmosphere. Quantitative and morphological changes, after contact with the examined materials, were evaluated after 24, 48, and 72 h under a reversed contrast-phase microscope. Trypan blue staining was used to determine the number of dead cells. The degree of toxicity of the materials was assessed on the basis of changes in cell morphology, their survival, and ability to proliferate according to the criteria given in Table 1.

Table 1 Degrees of direct contact toxicity test

In vivo Testing

Based on European Pharmacopoeia 6th edition and ISO 10993-1:2010 Biological evaluation of medical devices-Part 1: Evaluation and testing within a risk management process (ISO 10993-1:2009) guidelines for biological evaluation of MBC vascular prostheses, we have conducted the following tests:

  • Acute oral toxicity test

  • Intradermal reactivity test

  • Guinea pig maximization test (GPMT)

  • Histopathological evaluation of biocompatibility after implantation of BC tubes in animal body

Toxicity Test

The toxicological test was carried out in accordance with the requirements of OECD Guidelines for the testing of chemicals, section 4 Health Effects Test No. 425: acute oral toxicity: up-and-down procedure. The limit test is carried out on 5 animals, which are given a test substance of 2000 mg/kg. If at least 3/5 of the animals survive within 14 days, the LD50 is considered to be above 2000 mg/kg. If 3/5 of the animals die, the main test is carried out. If at least 3/5 of the animals survive for 14 days, the LD50 is considered to be above 2000 mg/kg. If 3/5 of the animals die, a further study is performed—the main test, consisting of administration of the following sequence of doses 1.75, 5.5, 17.5, 55, 175, 550, and 2000 mg/kg (growth factor of successive doses 3.2) to individual animals at 24-h intervals. If the first animal survives, the next one is given a higher dose. If after the next dose the animal dies or is in agonal condition, a lower dose is administered, between the lethal and non-lethal dose. Statistical evaluation of the test results is performed using the “Statistical Programme-AOT 425 StatPgm, Version: 1.” The study used a suspension of fragmented lyophilizate of MBC. Biological-modified bacterial cellulose vascular prostheses were homogenized in distilled water, freeze-dried to remove water, ground, and then dried for 20 h at 105 °C. The material for testing was placed in polypropylene packages and secured against re-absorption of moisture by the material with parafilm. Thus prepared prostheses for testing were suspended in water for injections, thus forming a dense, gelatinous suspension. The suspension intended for administration with a stomach probe was obtained after adding up to 1000 mg of cellulose 12.5 ml of water. Injection water “aqua pro injection” was used for the doses. (500ml, Series No.: 0907404, Manufacturer Baxter Manufacturing Sp. z o.o.). The study was conducted on ten white Imp:WIST rats (five males and five females). Prior to the application of doses, the body mass of the animals was accurately determined. Bacterial cellulose suspension at a dose of 2 g/kg b.w was administered through a probe into the stomach. Due to the fact that the quantity of the administered suspension exceeded 2 ml/100 g b.w. (OECD guidelines), the substance was administered in two equally divided portions at an interval of four hours (0900 and 1300). The total amount of administered cellulose dispersion was 2.5 ml/100 g b.w. = 0.2 g/100 g m.c, which corresponds to a dose of 2 g/kg b.w. Consecutive observations were performed twice a day. The body weight of rats was determined 3, 7, and 14 days after treatment, as stated in OECD guidelines No. 425. The scope of animal observations was as follows: autonomic and central nervous system, respiratory system, and circulatory system, as well as the skin, fur, eyes, and mucous membranes. Attention was drawn to the occurrence of tremors, convulsions, salivation, diarrhea, lethargy, and coma. Somatomotor activity and behavior were also evaluated.

Intradermal Reactivity Test

The study used three rabbits Imp:BN. Two extracts were prepared: nonpolar in cottonseed oil and polar in isotonic sodium chloride solution (which are substances of choice in our lab for this procedure), according to ISO 10993-12:2012.

Intracutaneous injections with 0.2 ml of the extract obtained with polar or non-polar solvent at five sites were performed on the left part of each rabbit. Similarly, the right side was injected with polar or non-polar solvent control as shown in Fig. 4. Observations were performed immediately after injection and at 24, 48, and 72 h, as well as 4 and 5 days after injection.

Guinea Pig Maximization Test (GPMT)

The maximization test was conducted according to the OECD Guidelines for the Testing of Chemicals, Section 4 Health Effects Test No. 406: Skin Sensitization of Magnusson and Kligman and based on ISO 10993: 10; 2011 [12]. On the day before investigation, the fur was removed from dorsal area of the guinea pigs on both sites of the trunk, without injuring the skin. Then, on the first day of the induction period, three pairs of intradermal injections (0.1 ml/site) were performed in the scapular area (Fig. 5). On day 8 of the experiment, 0.5 g of 10% sodium lauryl sulfate in vaseline was applied on the right site of all guinea pigs to evoke local inflammation, and these areas were washed after 24 h. Next, a sterile gauze compress with 0.5-ml analyzed water extract was applied to the exposed skin, stabilized with a bandage and left for 48 h. On day 22 of the study, a sensitization reaction was induced in all guinea pigs.

Checking test for sensitization was performed after 24 h. The effects of sensitization were evaluated according to the Magnusson-Kligman Classification 48, 72, and 96 h following the sensitization reaction.

Implantation

The examination was carried out in accordance with the requirements of ISO 30993:6 2009. Thirty male white rats were used in the vascular prosthesis testing. With the animals under deep anesthesia (80–100mg/kg ketamine hydrochloride i.p.; Sigma Aldrich Sp. z.o.o, Poland), the tested implants were inserted into the pocket of panniculus camosus muscles along the dorsal midline, and kept there for 1 month (vascular prosthesis). Following this time, a block measuring 2 cm × 2 cm from the point of entry of the implant (cervical-interscapular region) was resected. This sample covered the entire thickness of the back muscles.

Histopathological Evaluation After Implantation of Vascular Prosthesis of Chitosan-Modified Bacterial Cellulose

The examination was carried out in accordance with the requirements of ISO 30993:6; 2009. The muscle sections were dehydrated in increasing concentrations of alcohol and imbued with paraffin in an RVG/1 computer tissue processor (Belair Instrument Company). Microscopic samples of 4–6μm thickness were prepared on an MH-325 rotary microtome (Thermo Scientific™) and stained with hematoxylin and eosin in a Varistain Gemini staining machine (Varistain™ Gemini ES Automated Slide Stainer A78000014). In addition, samples were stained in acidic fuchsin in picric acid in order to visualize collagen fibers.

After removal of the implants, two assessments of the rats’ entrails were conducted: evaluation of the internal organs and their blood supply and a macroscopic examination of the muscles surrounding the implants and of the stomach, liver, and kidneys.

Statistical Analysis

The statistical analysis of in vitro testing results was performed using the Student’s t-test. The correlation coefficient was assumed to be significant at *p <0.05, **p<0.01, ***p<0.001.

In comparing the results obtained by in vivo testing, the chi-square test with Fisher’s exact test were used to determine whether the differences in scores between the tested animals or sites and respective controls were significant. The differences in the number of responses of the test site/group versus the control site/group were compared using the Mann-Whitney U test. Differences were considered significant for p<0.05. Statistical analysis of the results was performed with Statistica software for Windows (v.12 StatSoft, Inc., Tulsa, OK, USA).

Results

Physical-Mechanical Properties of BC Tubes

Figure 3 shows the spectra for bacterial cellulose without modification; chitosan oligomers used to modify BC tubes and modified bacterial cellulose—BC tube. A wide band at 3400 cm−1, attributed to the stretching vibration of the OH (connected by hydrogen bond) and/or NH group, could be observed in the all spectra. A band within the range 2800–2950 cm−1 corresponding to the stretching vibration of the C-H bond—characteristic for both polysaccharides—also could be observed. Because during biosynthesis the nutrition medium was modified by chitosan oligomers, also the bands characteristic for chitosan oligomers corresponding to the stretching vibration of the C=O bond in primary amides detected in 1630–1660 cm−1, was observed also in BC tubes structure. A band at 1320 cm−1 observed in chitosan oligomers and BC tube corresponds to C-N stretching of amide III. A band at 1550 cm−1 corresponding to N-H bending of amide II was observed in chitosan oligomers, but not in BC tube. Probably it was overlapped by other bands.

Fig. 3
figure 3

Fourier transform infrared spectroscopy (FTIR) analysis of investigated. Samples detecting structures characteristic for saccharide structures and chitosan oligomers

Fig. 4
figure 4

Location of skin application sites for Intradermal Reactivity Test

Fig. 5
figure 5

Schematic illustration of Guinea pig maximization test

Fig. 6
figure 6

Proliferation rates in 3T3Balb/C mouse fibroblast in control cultures and cultures with vascular prosthesis samples from modified bacterial cellulose (MBC).

Fig. 7
figure 7

Histopathological examination of BC tubes implants. (7.1) Shows the implant inserted underneath the epimysium, (7.2) implant on the muscle surface, and (7.3) implant inserted intramuscularly. The implants are surrounded by a thick connective tissue bag, from which vascular and connective tissue bands grow into the implant. (7.4) Intramuscular implant. The muscle is sharply separated from the implant by a bag made of connective tissue fibers, fibroblasts, and histiocytes. Numerous spindle cells (fibroblasts) and mononuclear cells penetrate from the capsule to the implant. The muscle has a normal structure without infiltration from the reagent cells. (7.5) Implant inserted underneath the epimysium. The muscle is separated from the implant by a bag made of connective tissue fibers, fibroblasts, and histiocytes. Preparations stained with hematoxylin and eosin. Numerous spindle cells (fibroblasts) and mononuclear cells penetrate from the bag to the implant. The muscle has a normal structure without infiltration from the reagent cells. (7.6) Intramuscular implant. The muscle is sharply demarcated from the implant by a thick bag made up of connective tissue fibers, fibroblasts, and histiocytes. Strands of connective tissue with vessels grow into the implant. Numerous fibroblasts penetrate from the capsule and connective tissue band. (7.7) Intramuscular implant. The end of the connective tissue band in the center of the implant. Proliferating capillaries are visible. (7.8) Implant inserted intramuscularly. Visible groups of blood capillaries and numerous spindle cells (fibroblasts). No form of reactive inflammatory cells was observed in the muscle area

Fig. 8
figure 8

Histopathological examination of implanted BC tubes. Preparations dyed Van Gieson. The implant inserted intramuscularly. The muscle is sharply demarcated from the implant by a bag made of collagen fibers and the implant inserted under the muscle. The muscle from the implant is sharply demarcated by a thick collagen fiber bag. A vascular and connective tissue band containing collagen fibers grows into the implant. No signs of muscle interstitial fibrosis

The BC tubes were found to possess the following physical-mechanical properties:

  • Length 10–12 cm

  • Surface density of about 10 g/m2 (of dry cellulose)

  • Selected molecular parameters of bacterial cellulose (GPC/SEC analysis according to the Turbak procedure [10, 11]): number average molar mass Mn 150 kD, mass average molar mass Mw 310 kDa, average polymerization degree equal to 1954

  • Peripheral tenacity of approximately 14.5 cN/mm, (according ISO 7198:1998)

  • A longitudinal tenacity of approximately 930 cN, (according to ISO 7198:1998)

  • A suture retention strength of 50 cN, (according to ISO 7198:1998)

  • Water permeability below 1 ml/min cm2 at a pressure of 120 mm Hg (according to ISO 7198)

  • Content of nitrogen: 0,16mg/g BCTube (Vario Macro Cube analyzer).

  • Endotoxin content—1.25 EU/product—semi-quantitative gel test LAL (limulus amebocyte lysate) (according to US Pharmacopeia 27). The permissible level of endotoxin content according to USP 27 is 20 EU/ product

Biocompatibility of BC Tubes

In Vitro Testing of Cytotoxic Action: fControl Cell Cultures

After 24, 48, and 72 h in control cultures, the cells adhered to the medium and had proper morphological features. No agglutination, vacuolization, separation from the medium, or lysis of cell membranes was found. Cell proliferation was normal. The cells formed colonies, which covered the whole plate. After 72 h in control cultures, 1% of dead cells were found.

Cultures with Vascular Prosthesis Samples from Modified Bacterial Cellulose (MBC)

In cultures with modified cellulose samples, at all testing times, the cells adhered to the substrate and had correct morphological characteristics. No agglutination, vacuolization, separation from the background, or lysis of cell membranes was found. Cell proliferation after 24, 48, and 72h was insignificantly lower compared to cell proliferation in control cultures. After 48h in cultures with samples 2% of the cells were dead and after 72h 5% of the cells were dead (Table 2, Fig. 6).

Table 2 Cytotoxic changes in 3T3Balb/C mouse fibroblast in control cultures and cultures with vascular prosthesis samples from modified bacterial cellulose (MBC)

Hemocompatibility

In the study of hemolytic activity using concentrated red blood cells, which come into contact with both the material under evaluation and its extract in physiological saline, the average value of hemolysis percentage was 0.25 ± 0.06%. The obtained value did not exceed the value allowed by the standard, i.e., 3%; in full blood hemolytic activity tests (CPD), having direct contact with the surface of the material, the mean value of hemolysis percentage was 0.14± 0.03%. The obtained value did not exceed the value acceptable by the standard, i.e., 1%; in tests of plasma clotting system, significant (p<0.05) shortening of prothrombin time (PT) with unchanged values of partial thromboplastin time after activation (APTT) was found within 2 h of blood contact with the material. However, after 24 h, APTT and PT were prolonged. The thrombin time (TT) values were prolonged and fibrinogen concentration decreased. It follows that the grafts made of modified bacterial cellulose activates to some extent the blood clotting process. However, the observed changes in the values of the determined clotting system parameters do not exceed the range of reference values. These changes may be due to the effect of material components on the coagulation factors of both the intra- and extra-transmitted system as well as to the conversion of fibrinogen into fibrin. The activation of the clotting system by the material under assessment is evidenced by shortened clotting time of whole blood after calcification.

Toxicity Test

No pathological signs were observed in the fur, orifices, eyes, muscle tension, or behavior of three males and three females immediately after administration of modified BC suspension. After 1 day, female no. 65 was less mobile; there was a clear tilt of the head to the right and increased tension in the neck muscles. After 2 days, the tested female died. Based on clinical observations and the results of the post-mortem examination, it was assumed that the cause of rat’s death was a lesion of the esophagus wall during administration of the test substance. Death was then assessed not be related to the toxic effects of tested implants. All of the remaining animals survived the period of 14 days of observation (five males and four females). Constant weight gain was found (Table 3). No signs of pathology were found regarding animal behavior, evaluation of hair in natural orifices of the body, the appearance of the eyes, or muscle tension. The autopsy found the structure and blood supply to the internal organs to be normal. The macroscopic image of the gastric area and small and large bowel was within normal limits. No pathological signs regarding animal behavior, hair, orifices, the eyes, or muscle tension were noted within 14 days. Detailed results of sectional studies was presented in Table 4.

Table 3 Body weight measurements of rats after intragastric administration of the suspension of fragmented lyophilizate of biological vascular prostheses from chitosan-modified bacterial cellulose (MBC) for acute oral toxicity testing
Table 4 Results of sectional studies

Intradermal Reactivity Test

The results are presented in Table 5. The non-polar and polar extracts of modified bacterial cellulose implants revealed mild erythema in one out of three rabbits after 24 hours of examination time after 48 hours, this erythema disappeared. No significant changes in intradermal reactivity were observed. The Primary Irritation Index was found to be 0.03: The product is not irritating.

Table 5 Total summary of results of the study of reactivity intradermal extract of polar and of non-polar extract of biological vascular prostheses from chitosan-modified bacterial cellulose

Guinea Pig Maximization Test (GPMT)

The detailed results are presented in Table 6. According to Chapter 3.4, Respiratory or Skin Sensitization, GHS guidelines, and substances which test maximize Magnusson and Kligman cause an allergic reaction less than 30% of the test animals should not pose a risk of allergic reactions in humans and should not be classified as substances with sensitizing properties (Global System for harmonization of classification and labeling of chemicals, United Nations, New York and Geneva, 2003). Therefore, it is considered that medical devices—biological vascular prostheses with modified bacterial cellulose—should not pose a risk for allergic reactions in humans.

Table 6 Skin sensitization test results

Histopathology

Histopathological studies showed that the implanted vascular implants induced an initial inflammatory response which developed into a chronic inflammatory immune response known as host foreign body reaction (FBR). The classic host response is generally divided into several stages, from the formation of the wound during implant placement to the wound healing period. The final host response reaches a relative impasse with most implants, unresolved in chronic inflammation as long as the implant remains. This final condition is defined by the long-term presence of both activated macrophages and the associated giant foreign body cells (FBGC) at the tissue/implant interface and the formation of a dense fibrous collagen capsule around the implant. These two characteristics—the presence of FBGC and the thickness of the fibrous capsule—are characteristic for the evaluation of the FBR. Chronic local condition remains for the whole period of implantation. The implant morphology and biomaterial as well as location of the implant determine the host’s response to the implant which is not dependent on the chemical composition of the material, the size of the implant, and method of introduction. The study was conducted on outbred Stock Han Wistar Imp:WIST rats due to the possibility of extrapolating the results to human populations.

In microscopic view, implants of modified bacterial cellulose were visible as an amorphous, fibrillar structure. The implants were placed between the muscles of the back, the epimysium, and intramuscularly (Fig. 7.1, 7.2, and 7.3). In all animals, the implants were surrounded by connective capsule (Fig. 7.4, 7.5, and 7.6 and Fig. 8.1 and 8.2) formed by histiocytes and fibroblasts (Fig. 7.4 and 7.5) and collagen fibers (Fig. 8.1 and 7.8). The implants placed under the epimysium produced a significant thickening of the connective capsule over the implant (Fig. 7.1). In all cases, vascular-connective tissue strands grew from the inside of the capsule into the implants (Fig. 7.1, 7.2, 7.3, and 7.6). These bands were made of collagen fibers (Fig. 8.2), blood vessels, histiocytes and fibroblasts (Fig. 7.6 and Fig. 7.7). In the central parts of the implant, new capillary loops were visible (Fig. 7.7 and 7.8).

Numerous fibroblasts surrounded the vascular-connective bands growing between the capsule and the implant (Fig. 7.4 and 7.6). The implants introduced into the intramuscular area were surrounded with a connective capsule, well-separated from the structure of skeletal muscle. No visible change in the form of reactive inflammatory cells was observed in the muscle area (Fig. 7.4, 7.5, and 7.6). There were no apparent signs of muscle interstitial fibrosis (Fig. 8.1 and 8.2). The macroscopic examination of the muscles surrounding the implants and of the stomach, liver, and kidneys confirmed the correct structure of internal organs and the degree of their blood supply. Macroscopically evaluated appearance of gastric and small and large intestine mucosa was within normal limits.

Discussion

Constantly rising expectancy of life and progress in the implant technology causes continuous increase in number of implantations in human medicine. Thanks to advance in the field of microscopy, suture materials and microsurgical instruments microsurgical techniques have developed to make surgical interventions at the finest structures of organism possible. In case of vascular prostheses, autogenous vessels are still most widely used due to their biocompatibility; however, this procedure requires additional operation and there are situations in which usage of this kind of prostheses is not possible. Moreover, the number of autologous grafts for coronary artery disease and for peripheral artery disease is limited. Due to this fact, scientists from all around the world started looking for alternatives. The ideal implant material should integrate without complications with tissue of the patient and maintain close connection over the time. Clinical problems such as damage of tissue around the implants, loosening of the implants, pain, and surgical revisions could be avoided. Particularly materials that will have contact with blood must comply with certain requirements such as mechanical strength, biocompatibility, blood compatibility, and sterilizability. Moreover, the inner and outer surface of artificial blood vessel must have different properties: outer should be overgrown with connective tissue whereas inner should be covered with endothelial cells and not stimulate adhesion of blood cells and components [13]. The essential requirement for biomedical materials is biocompatibility, referred as the ability of a foreign material implanted in the body to exist in harmony with tissue without causing deleterious changes. Generally, studies provide various results due to the different methodologies, workflow, and sample preparations [9]. Nowadays the most common clinically used synthetic grafts in peripheral artery disease remain expanded polytetrafluoroethylene, Dacron, and polyurethane [14]. Because of the unsatisfactory patency rates of common clinically used synthetic grafts, risk of stasis and thrombosis and rigidity, the researchers focus mainly on new grafts as small-diameter vessel substitutes [14]. The main challenge for tissue engineering is to develop new biologically active small-diameter grafts that would not cause serious side effects. Bacterial nanocellulose (BNC) is one of such example which has demonstrated a tempting prospect for applications in substitute of small blood vessels [15]. Bacterial cellulose (BC) synthesized by Gluconacetobacter xylinus has recently been studied as a biocompatible material due to its unique features such as high purity, crystallinity, biodegradability, and tensile strength. Promising mechanical properties and good biocompatibility of BC promote its research and development as substitute/medical biomaterial, such as the replacement of blood vessel—vascular grafts [9]. Currently, BC has very high potential to be used as biomaterial on a larger scale. In our study, new material made from bacterial nanocellulose modified with chitosan used in future as biological vascular prostheses was investigated. In our study, this novel vascular prostheses made of bacterial nanocellulose modified with chitosan (MBC) was tested for their potential toxicity and adverse sensitization effects. The presented results and previously conducted research are consistent that in most of the experimental studies, the lack of toxicity or low toxicity of nanocellulose was observed. However, nanocellulose used as biomedical material in humans should be subject to more specific toxicological control. Chitosan is affirmed as a non-dangerous and biocompatible polymer. It is thought that chitosan-based biomaterials are antimicrobial, functional, renewable, nontoxic, bioabsorbable, and biodegradable biopolymer agents [4]. CS alone is considered to be suitable for oral organization, but its properties may change totally upon modification and combination with nanocellulose [16]. Moreover, it is understood that the pharmacokinetic properties of this kind of nanomaterials can significantly change when incorporated into a nanoparticulate system [16]. It should be emphasized that toxicology studies of nanocellulose and nanocellulose-based biocomposites are still restricted at a very preliminary stage, mainly on the level of cytotoxicity [9]. To confirm the earlier hypothesis about the lack of toxicity of materials based on nanocellulose and chitosan, we performed a number of tests, e.g., Toxicity Test No. 425: acute oral toxicity: up-and-down procedure; intradermal reactivity test; Guinea Pig Maximization Test No. 406: Skin Sensitization of Magnusson and Kligman (GPMT); and histopathological evaluation after implantation of vascular prosthesis of modified bacterial cellulose. Toxicity test showed that modified bacterial cellulose suspension does not exhibit any signs of pathology in animal behavior. Assessment of hair in the body’s natural orifices, eye appearance, and muscle tension did not indicate any signs of toxic effects of the test substance. The autopsy confirmed a normal structure and blood supply to internal organs and orifices of the body. In Intradermal Reactivity Test, the non-polar and polar extract prepared from biological vascular prostheses of modified bacterial cellulose have been evaluated. None of the above showed an irritating effect. The maximization test (GPMT) No. 406: Skin Sensitization of Magnusson and Kligman confirmed lack of sensitization effect. During implantation process, correct structure of internal organs as well as the level of their blood supply was reported. Macroscopic view of gastric and small and big bowel was within normal limits. Histopathology assessment showed that in all animals, the implants were surrounded by a connective capsule mainly consisting of histiocytes, fibroblasts, and collagen fibers. In the muscle area, the reactive inflammatory cells were not found. There were no apparent signs of muscle interstitial fibrosis. Summarizing all obtained results, we can conclude that the study confirmed that biological vascular prosthesis of modified bacterial cellulose does not induce inflammatory response and hyperplasia of muscle tissue surrounding the implant. Previous studies on bacterial synthesized cellulose (BASYC®) have been a great success in the field of implantation and long-term maintenance of carotid arteries among rats and pigs. Another study conducted on BASYC® tube prosthesis for blood vessels has shown that BC has a potential to be used as a scaffold in tissue engineering process and in in vivo tissue-engineered blood vessels as part of cardiovascular surgeries. This work confirms that it will be possible to fabricate BC as in vitro scaffold and implant it as a substitute for a small diameter carotid arteries [17]. Wippermann et al. showed that BC exhibits properties that are promising for its further use in tissue-engineered blood vessels programs. Tissue-engineered blood vessels (TEBVs) represent an innovative approach for overcoming reconstructive problems associated with vascular diseases by providing small-caliber vascular grafts. Moreover they confirm that BC has high mechanical strength, promotes in situ vascular tissue regeneration, can be shaped into three-dimensional structures, and does not require pre-treatment in the form of cell seeding. The new BC-engineering technique they used resulted in a production of new stable vascular conduits [7]. Scherner et al. investigated the biocompatibility, blood compatibility, and biostability of a novel biomaterial of bacterially synthesized cellulose (BC) as a potential scaffold for small-diameter TEBV. BC tubes were implanted to replace the carotid arteries of 10 sheep for a period of 3 months. Analysis of BC grafts in a sheep model revealed the development of a three-layered structure of these grafts similar to those of native arteries with a single layer of endothelium with basement membrane followed by a concentric layer of smooth muscle cells and an outer layer with adjacent tissue with ingrowing capillaries after 3 months. The study confirmed that BC grafts support the physical load imposed by hemodynamic environment and allow remodeling, have minimal inflammatory potential, and are functionally stable. All patent grafts displayed confluent luminal endothelialization without neointima or thrombus formation. Scherner et al. came to similar conclusions during the research on in vivo application of tissue-engineered blood vessels of bacterial cellulose. Conducted study confirmed that BC can be a potential scaffold for small-diameter TEBVs and confirmed the statement that in future, TEBV programs will be used for cardiovascular surgeries [8]. Leitão et al. demonstrated that grafts made of BC which was histologically evaluated do not show any signs of chronic inflammation, infection, foreign body responses, clot formation, cell in growth, and diameter via electron microscopy [18]. BASYC® a novel biomaterial introduced to the market by Lee et al. actually is successfully used for implantation and long-term maintenance of carotid arteries. Moreover BASYC® prostheses are non-toxic and have non-immunoreactive effect of BC implantation. Azevedo et al. obtained similar results to ours; he proved that cellulose and chitosan (CELL:CHIT) polymers are biocompatible, nontoxic, and very abundant, which make them good candidates for biomedical applications [19]. Kim et al. investigated the immunoreactivity of BC in vitro in human umbilical vein endothelial cells (HUVECs) and in vivo using BALB/c mice. BC does not induce apoptosis and necrosis in HUVECs and does not stimulate immune response in both HUVECs and BALB/c mice models. These results suggest that BC may be widely used as a biocompatible biomaterial for tissue engineering.

Conclusions

Biological vascular prosthesis of modified bacterial cellulose does not induce inflammatory response and hyperplasia of muscle tissue surrounding the implant. The tested substances do not cause any allergic, intradermal reactions and finally, do not display acute toxicity.