Introduction

The efficacy of an inhaled drug is dependent on the propensity for drug particles to deposit throughout the lungs [1]. Deposition of drug particles in the airways is affected by several, often interrelated, factors, including the intrinsic properties of the inhaler, drug formulation and patient characteristics [25]. Drug formulation (i.e., dry powder or aerosol), the mechanics of drug aerosolization and dose delivery of the device [6], the particle carrier and required inhalation rate for particle deagglomeration [for dry-powder inhalers (DPIs)] [7, 8], or the characteristics of the propellant and the plume velocity [for pressurized metered-dose inhalers (pMDIs)] [4, 6, 9], as well as the particle size distribution profile of the emitted drug, are all properties inherent to the inhaled product. Patient-related factors include anatomy, disease state, lung function and the way in which an individual uses his inhaler (i.e., device handling, inhalation maneuvers, and inhalation flow rate and volume) [13, 5, 1012].

In vitro particle size profiles of emitted particles correlate with in vivo lung deposition. Particles larger than 5 μm in aerodynamic diameter are prone to being deposited in the mouth or throat, where they do not exert therapeutic effects and may increase the risk of local side effects [5, 13, 14]. Fine particles (<5 μm) are associated with deposition throughout the airways [1, 4, 5, 1518]. The proportion of an emitted aerosolized drug dose that contains fine particles is referred to as the fine particle fraction (FPF). Pharmacokinetic studies and three-dimensional imaging techniques have shown that inhaled corticosteroids (ICSs) with a higher FPF are associated with greater levels of lung deposition than those with a lower FPF [17, 19]. Imaging data have also indicated that fine particles deposit in both the large and small airways [19]. The FPF may therefore be an important attribute for an ICS and long-acting β2-agonist (LABA) combination therapy (ICS/LABA) because the pathophysiology of asthma involves both the proximal and distal airways. The ICS targets inflammation, which affects airways from the large bronchi to the alveoli, while bronchodilation is driven by β2-agonists acting on smooth muscle in the proximal airways [19, 20]. Thus, it follows that the deposition profile of an ICS/LABA combination therapy should facilitate targeting of inflammation throughout the lungs and provision of bronchodilation in the proximal airways.

The ICS fluticasone propionate (FP) and LABA formoterol fumarate (FORM) have been combined in a single, hydrofluoroalkane (HFA)-based, aerosol inhaler [FP/FORM; flutiform® (Jagotec AG, Muttenz, Switzerland)]. FP/FORM has been approved for maintenance treatment of asthma in adults and adolescents at twice-daily doses of 100/10, 250/10 μg and (adults only) 500/20 μg when an ICS/LABA therapy is appropriate. The efficacy and tolerability of FP/FORM across a range of asthma severities have been demonstrated in several studies. FP/FORM has a similar efficacy and tolerability profile to those of fluticasone propionate/salmeterol xinafoate (FP/SAL; pMDI) and budesonide/formoterol (BUD/FORM; DPI), the two most commonly prescribed ICS/LABA combinations [21, 22].

It is widely accepted that DPIs are often associated with flow rate-dependent changes in the FPF; this is largely because the dry-powder formulation needs to be deagglomerated into respirable drug particles, a process that is dependent, to a variable extent, on the inspiratory force at which the patient inhales through the device [7, 8, 23]. With pMDIs, drug particles, either suspended or dissolved in a co-solvent/propellant mixture (suspension or solution formulations, respectively), are aerosolized on actuation of the inhaler [6]. Therefore, it is expected that inhalation flow rate would not have a significant impact on the FPF. In this study (defining fluticasone propionate/formoterol particulate size [DIFFUSE]), the in vitro FPF profile of the FP/FORM aerosol was compared with that of the commercially available FP/SAL and BUD/FORM DPIs. The beclometasone dipropionate (beclometasone)/formoterol (BDP/FORM) aerosol was included as a benchmark comparator.

Materials and Methods

ICS/LABA Combinations

This study compared the aerodynamic particle size distribution (APSD) profiles of FP/FORM (pMDI) with three other single-inhaler ICS/LABA combination products: FP/SAL [Seretide® Accuhaler® (GlaxoOperations UK Ltd, Herts, UK); DPI]; BUD/FORM [Symbicort® Turbohaler® (AstraZeneca AB, Södertälje, Sweden); DPI]; and BDP/FORM [Foster® (Chiesi Farmaceutici S.p.A, Parma, Italy); HFA pMDI]. Drug strengths were selected based on their ability to deliver similar treatment doses (Table 1).

Table 1 Summary of the ICS/LABA combinations tested

Aerodynamic Particle Size Distribution Profiles

APSD tests were performed using the Apparatus D 8-stage Andersen Cascade Impactor (ACI; Copley Scientific Limited, Nottingham, UK), with an induction port and pre-separator for powder formulations, according to the European Pharmacopoeia 7.8, General Chapter 2.9.18: ‘Preparations for inhalations, aerodynamic assessment of fine particles’. Testing was performed at room temperature under ambient humidity conditions. To ensure thermal re-equilibration of the valve, pMDIs were actuated four times with a delay of approximately 2 s between each actuation. For the DPI testing, the plates of the ACI were coated with a surfactant (Tween® 20; Croda France, Trappes Cedex, France) to avoid potential bouncing or re-entrainment effects; a pressure drop (4 kPa) was included at the start of the ‘inhalation’, as recommended for quality testing (to simulate the sharp initiation inhalation recommended for DPIs). APSD profiles were assessed at inhalation flow rates of 28.3 and 60.0 L/min, which are within the clinically effective range of flow rates for the inhalers tested, and are consistent with industry standards for the quality control testing of inhaled medicines. Five inhalers were tested for each product. The cumulative drug mass recovered from the ACI following each test was normalized to the labeled dose for each combination component at both flow rates tested.

Fine Particle Fraction

The fine particle dose (mass of dose comprising particles <5.0 μm) was interpolated from the cumulative distribution for the ACI stages, with cutoff diameters of 3.3, 4.7 and 5.8 μm, and normalized for the labeled dose. The FPF was calculated as a percentage of the labeled dose for the ICS and for the LABA component of each product, at both flow rates. The ratio between FPFs at 60.0 and 28.3 L/min (FPF 60.0/28.3) for both the ICS and LABA component of each ICS/LABA combination tested was also calculated.

Median Mass Aerodynamic Diameter

The median mass aerodynamic diameter (MMAD) was derived from inertial impaction data, assuming a log-normal dispersion of sizes, according to ISO 27427 ‘Anaesthetic and respiratory equipment—nebulising systems and components’. The geometric standard deviation was calculated from the particle size at the 84th percentile (by mass) divided by the MMAD.

Statistical Analysis

The objectives of the study were: (1) to quantify the effect of flow rate on the FPF for each of the four products and (2) to compare the FPF of the comparator products to that of the FP/FORM pMDI at each of the studied flow rates. Results from one inhaler (#3, Seretide batch R502221) tested at 28.3 L/min were atypical and removed from statistical analysis.

The FPF data were normalized to the target metered dose to compare the products. Both objectives were investigated by means of standard equivalence testing; a 90% confidence interval (CI) for the mean TEST/REFERENCE ratio was determined based on log-transformed FPF data, and equivalence was concluded if the CI was contained in the 0.85–1.18 (±15%) acceptance interval recommended by the European Medicines Agency for in vitro equivalence. For objective (1) the reference was 60 L/min, and for objective (2) the reference was FP/FORM. The same analysis was performed independently for both drug components. Means and standard deviations are presented for each product; API and flow rate were calculated in original scale (not using the log transformation).

This article does not contain any studies of human subjects performed by any of the authors.

Results

Fine Particle Fraction

FP/FORM produced a high FPF of approximately 40% for both the ICS and the LABA components at flow rates of 28.3 and 60.0 L/min (Fig. 1). The FPF of FP/FORM was significantly higher than that of the FP/SAL and BUD/FORM DPIs (Table 2; Fig. 1). The FP/SAL DPI produced an FPF of less than 20% for each component at both flow rates, while the FPF of budesonide for the BUD/FORM DPI was 8.2% at 28.3 L/min and 35.0% at 60.0 L/min.

Fig. 1
figure 1

FPFs of the a ICS and b LABA components of the combination therapies at 28.3 and 60.0 L/min. Error bars represent standard error. CI confidence interval, FPF fine particle fraction, ICS inhaled corticosteroid, LABA long-acting β2-agonist

Table 2 Ratios and 90% CIs of the FPFs for the ICS and LABA components of each combination versus those of FP/FORM

The FPF of FP/FORM was consistent across the flow rates tested, as shown by the FPF 60.0/28.3 ratios of 1.06 for fluticasone and 1.08 for formoterol (Table 2), which were within the specified acceptance interval for equivalence. By contrast, the FPFs emitted from the FP/SAL and BUD/FORM DPIs demonstrated marked flow rate dependency. The FPF 60.0/28.3 ratio for the FP/SAL DPI was 1.43 for fluticasone and 1.33 for salmeterol (Table 2). Similarly, the FPF of BUD/FORM increased more than fourfold for the ICS and LABA components between the lower and higher flow rates (Fig. 1), with FPF 60.0/28.3 ratios of 4.38 for budesonide and 4.63 for formoterol (Table 2).

The FPF of BDP/FORM at 28.3 L/min was less than 30%. The BDP/FORM FPF varied with flow rate and was approximately 40% at 60.0 L/min (Fig. 1; Table 2). The variation in the FPF across flow rates was demonstrated by the 60.0/28.3 ratios of 1.51 for beclometasone and 1.52 for formoterol (Table 2).

Aerodynamic Particle Size Distribution Profiles and Median Mass Aerodynamic Diameters

APSD profiles and MMADs were relatively similar for the ICS and LABA components of each combination tested, at both flow rates, indicating consistency between component particle size fractions (Table 3). FP/FORM showed a similar cumulative particle fraction distribution profile and MMAD at both 28.3 and 60.0 L/min, consistent with the similar FPFs observed between flow rates (Fig. 2a; Table 3). The APSD profile of FP/SAL was shallow at both 28.3 and 60.0 L/min, with particles distributed across a broad range of sizes (Fig. 2b); there was relative inconsistency in both the FPF and the MMAD between the flow rates (Tables 2, 3). For BUD/FORM, the APSD profile and MMAD were both noticeably different at 28.3 L/min and at 60.0 L/min, with an increased mass of particles in the range 1.1–10.0 μm at 60.0 L/min (Fig. 2c; Table 3).

Table 3 MMADs for the ICS and LABA components of each combination tested at 28.3 and 60.0 L/min
Fig. 2
figure 2

Aerodynamic particle size distribution profiles of the LABA component for a FP/FORM, b FP/SAL, c BUD/FORM and d BDP/FORM. Error bars represent standard error. Cumulative drug masses recovered from the ACI normalized for labeled dose are shown at 28.3 and 60.0 L/min. ACI Andersen Cascade Impactor, ICS inhaled corticosteroid, LABA long-acting β2-agonist

BDP/FORM also showed flow rate-dependent effects for the cumulative particle fraction (Fig. 2d), with a shift in the profile from 28.3 to 60.0 L/min and an increased mass of particles smaller than 1.1 μm at the higher flow rate, resulting in a decrease in the MMAD from 1.21 μm and 1.24 μm for ICS and LABA, respectively, at 28.3 L/min, to 0.50 and 0.49 μm at 60.0 L/min (Table 3).

Discussion

The primary aim of this in vitro study was to characterize the FPF of the FP/FORM inhalation aerosol relative to those of the two most widely prescribed ICS/LABA DPIs at two different inhalation flow rates that are achievable by patients using both DPI and pMDIs, and that are considered to deliver clinically effective doses for these device types. The BDP/FORM aerosol was originally included as a benchmark pMDI in this study. FP/FORM produced a high (approximately 40%) and consistent FPF at 28.3 and 60.0 L/min, which was greater than the FPF of FP/SAL and BUD/FORM. By contrast, the FPF varied across flow rates for both DPIs. For the FP/SAL DPI, there was a 5.3% absolute increase (42% relative increase) in ICS FPF and a 3.5% absolute increase (31% relative increase) in LABA FPF when the flow rate increased from 28.3 to 60.0 L/min. For the BUD/FORM DPI, the difference in FPF between flow rates was greater: an absolute increase of 26.8% (327% relative increase) in ICS FPF and a 23.5% absolute increase (356% relative increase) in LABA FPF were observed when the flow rate was increased from 28.3 to 60.0 L/min. These findings are consistent with those of previous in vitro investigations [8, 23, 24] and are largely due to the inspiratory force required to deagglomerate and aerosolize the drug dose during inhalation with a DPI. The similarity of the drug mass ratios for each size fraction, for all products tested, demonstrated that ICS and LABA particles generated during the atomization process are consistent in size or are co-associated, and would therefore be expected to co-deposit in the airways. This finding is in line with those of previous studies [25].

Given that MMAD and FPF are interrelated, it was unsurprising that the consistent FPF of FP/FORM correlated with MMAD at both flow rates. In contrast, both DPIs produced higher MMADs at the lower flow rate, which also varied considerably with increasing flow rate. The FP/SAL DPI was associated with relatively large particles (correlating with the flat APSD profile), with an MMAD of 4.89 μm for the LABA component at the lower flow rate. The BUD/FORM DPI produced an MMAD of 3.77 μm for the LABA component at the lower flow rate. These data indicate that at the lower flow rate, a large proportion of the dose of FP/SAL is emitted as particles larger than 5 μm; particles of this size are known to be more prone to being deposited in the mouth and throat than fine particles [5]. This increased risk of oropharyngeal deposition may be associated with local adverse effects [13, 14]. As expected, the magnitude of the flow rate effect on MMAD was greater for BUD/FORM than for FP/SAL, with a shift in MMAD from approximately 3.8 μm (at 28.3 L/min) to approximately 2.5 μm (at 60.0 L/min) for both the LABA and ICS components. These findings highlight that there is greater potential for deposition of large particles in the mouth and throat, rather than in the airways, at a lower inhalation flow rate for both DPIs compared with the pMDIs tested in this study.

Data on the FPF for BDP/FORM have been reported previously by Lewis and colleagues [25] as a proportion of the delivered dose at a flow rate of 28.3 L/min. In the present study, FPFs were calculated as a proportion of the labeled dose, and the FPF for the LABA component of BDP/FORM was 26.0% at 28.3 L/min. When the data reported by Lewis and colleagues were recalculated as a proportion of the labeled dose, rather than the delivered dose, the FPF for the LABA component of BDP/FORM was 31.6%, which is comparable to the finding of the present study. In addition, the MMAD values calculated in the present study for BDP/FORM were similar to those reported by Lewis and colleagues (1.21 μm for the ICS and 1.24 μm for the LABA in the present study, versus 1.4 and 1.5 μm, respectively, as reported by Lewis). The aerosolized drug emitted from a pMDI is influenced by the propellant and the characteristics of the inhaler and, unlike DPIs, is not dependent on inspiratory force to deagglomerate the drug. This can lead to the perception that the FPF generated with pMDIs is also not dependent on inhalation flow rate. However, when flow rate increased from 28.3 to 60.0 L/min, there was an increase in the BDP/FORM FPF (formoterol component, from 26.0% to 39.5%). Similar to the findings for both DPIs, the flow rate dependency of the BDP/FORM FPF correlated with a lower MMAD at the 60.0 L/min flow rate (from 1.21 μm at 28.3 L/min to 0.50 μm at 60.0 L/min for the ICS), demonstrating the shift toward much smaller particles with increasing flow rate.

The cause of the observed effect of flow rate on BDP/FORM is not understood fully and we are not aware of any published data reporting the APSD profile of the BDP/FORM aerosol at 60.0 L/min or higher. Environmental and experimental factors, such as humidity and temperature, can influence aerosolized drug characteristics in vitro [26]. All products tested in the present study, however, were subject to the same environmental and experimental parameters, with only flow rate as a variable. Given that both combinations were analyzed under identical conditions and in accordance with the European Pharmacopoeia methodology, it is highly likely that the observed results relate to an effect of flow rate. There is support for this from a previous study, which demonstrated that the in vitro performance of some ICS/LABAs can be affected by flow rate, leading to increased deposition of particles in the ‘throat’ of the apparatus at low flow rates [27]. Such variance in FPF could be attributed to several factors, including turbulence (created as the aerosol is generated owing to the velocity difference of the impactor air flow and the aerosol plume) and droplet particle evaporation, which result in changes to particle deposition in the impactor throat [28, 29]. Flow rate had little effect on fluticasone/formoterol, but a greater impact on beclometasone/formoterol; hence, the observed flow rate-dependent effects probably relate to differences in the formulations of the two products. There are several differences between the composition of the FP/FORM and the BDP/FORM aerosols, and their respective devices: FP/FORM is a suspension containing the propellant, apaflurane (HFA-227), whereas BDP/FORM is a solution and contains the propellant norflurane (HFA-134a). Furthermore, there are differences in alcohol content (FP/FORM: 1 mg/actuation; BDP/FORM: 7 mg/actuation) and other excipients (FP/FORM: sodium cromoglycate; BDP/FORM: hydrochloric acid). Interestingly, a similar flow rate-dependent effect has been noted for an HFA-based, solution formulation of beclometasone, which has a similar alcohol content to that of the BDP/FORM product tested in this study (8.35 mg/actuation) [24, 27].

Although an optimal particle size for ICS/LABA therapy has not yet been determined [30], it is known that large particles (>5 μm) are more prone to being deposited in the throat and mouth, and that very small extra-fine particles (0.1–1.0 μm; those that deposit via diffusion and sedimentation) are prone to exhalation, especially if the breath-hold is insufficient. The observation that the BDP/FORM aerosol generated an MMAD of approximately 0.50 μm (for the ICS component) at a flow rate of 60.0 L/min may therefore have in vivo implications, because at the higher flow rate there may be increased potential for particle exhalation; however, particle size distribution around the MMAD has a greater influence on drug deposition throughout the lungs than any specific MMAD value [1, 31]. Evidence suggests that fine particles in the range 2–5 μm may be a suitable size for distribution of an ICS/LABA throughout the lungs [18]. Indeed, in vivo imaging techniques have shown that particles in this size range deposit in both the large and small airways [19]. In asthma, the large and small airways associated with bronchoconstriction and inflammation must be treated [18]. A significant proportion of the emitted FP/FORM dose, at both 28.3 and 60.0 L/min, falls in this size range (2–5 μm), which may facilitate drug deposition throughout the airways.

As with any in vitro investigation, caution must be applied when translating findings to the in vivo setting [1]; however, the European Pharmacopoeia methodology used in this study is well established and can be used to predict the behavior of drug particles in the human respiratory tract [1]. In addition to the properties inherent to the inhaler and the drug formulation, a patient’s inspiratory maneuvers will also affect the potential for deposition of the drug in the airways. Patients often make errors when taking their ICS/LABA, emphasizing the need for regular training and inhaler-use assessments [32]. Furthermore, a patient’s inhalation flow rate is not constant during an inhalation maneuver and may vary from breath to breath, and therefore from dose to dose; the flow rate may also vary from patient to patient [24, 33]. Inhalation flow rate impacts significantly on total and regional patterns of deposition of drug particles in the airways [19]. Therefore, an ICS/LABA with a relatively flow rate independent, consistent FPF might facilitate more uniform dosing in the lungs, compensating for variations in inspiratory flow rates among patients, and variations in patients’ day-to-day or successive inhalation maneuvers.

In 12-week, randomized clinical trials, FP/FORM has a similar efficacy and tolerability profile to FP/SAL pMDI (in patients ≥18 years with 125/5 and 50/5 μg FP/FORM doses) and BUD/FORM DPI (in patients ≥12 years with 125/5 μg FP/FORM dose) [21, 22]. Patients need to demonstrate optimal inhalation maneuvers and undergo repeated training to maintain their inhaler technique for inclusion in such studies; hence, it would be unlikely that a difference between two inhalers would be observed. There is, however, evidence to suggest that some modern HFA-based pMDIs, which have greater FPFs than current DPIs and the discontinued chlorofluorocarbon (CFC)-based pMDIs, may provide better control of asthma symptoms. A recent retrospective, observational study compared asthma outcomes in 3,134 patients who were receiving FP/SAL, administered via either a DPI or a pMDI. The findings of this study demonstrated that in practice, patients using a pMDI achieved better asthma control than those using a DPI [34]. The authors suggested that these results could reflect differences in FPF between the pMDI and DPI formulations of FP/SAL. Notably, the FPF of the FP/SAL DPI in the present study is markedly lower than that previously reported for the pMDI formulation (30–33%, as percentage of the metered dose), when assessed using an ACI without a spacer, as per the present study [35, 36]. Similarly, an HFA-based formulation of beclometasone provided better asthma outcomes than a CFC-based version in another observational study. Improved asthma control with the HFA-based formulation may be related to the higher FPF.

Conclusion

In conclusion, in vitro FP/FORM emits a high and consistent FPF (approximately 40%) at 28.3 and 60.0 L/min. The FPF of FP/FORM was less affected by changes in flow rate than the other products tested. Consistent FPFs may be associated with potential in vivo benefits.