Journal of The American Society for Mass Spectrometry

, Volume 27, Issue 9, pp 1531–1538 | Cite as

2D FT-ICR MS of Calmodulin: A Top-Down and Bottom-Up Approach

  • Federico Floris
  • Maria van Agthoven
  • Lionel Chiron
  • Andrew J. Soulby
  • Christopher A. Wootton
  • Yuko P. Y. Lam
  • Mark P. Barrow
  • Marc-André Delsuc
  • Peter B. O’Connor
Research Article

Abstract

Two-dimensional Fourier transform ion cyclotron resonance mass spectrometry (2D FT-ICR MS) allows data-independent fragmentation of all ions in a sample and correlation of fragment ions to their precursors through the modulation of precursor ion cyclotron radii prior to fragmentation. Previous results show that implementation of 2D FT-ICR MS with infrared multi-photon dissociation (IRMPD) and electron capture dissociation (ECD) has turned this method into a useful analytical tool. In this work, IRMPD tandem mass spectrometry of calmodulin (CaM) has been performed both in one-dimensional and two-dimensional FT-ICR MS using a top-down and bottom-up approach. 2D IRMPD FT-ICR MS is used to achieve extensive inter-residue bond cleavage and assignment for CaM, using its unique features for fragment identification in a less time- and sample-consuming experiment than doing the same thing using sequential MS/MS experiments.

Graphical Abstract

Keywords

Tandem mass spectrometry FTICR mass spectrometry 2-Dimensional mass spectrometry 

Introduction

Fourier transform ion cyclotron resonance mass spectrometry (FT-ICR MS) provides the highest resolving power and mass measurement accuracy among all kinds of mass spectrometry (MS) [1]. In an FT-ICR MS experiment, the raw signal is measured as a function of time providing time-domain data, or “transient.” The Fourier transform (FT) is then applied, along with a series of operations/algorithms such as zero-filling, convolution [2], and apodization [3] to improve the quality of data [4], transforming the time-domain data into the frequency domain. A frequency spectrum is obtained and converted to a mass (m/z) spectrum through a calibration function [5]. In order to obtain more structural information, analytes are often isolated and fragmented (tandem mass spectrometry, or MS/MS). Tandem MS is influenced by the resolution that can be obtained during isolation and by the various fragmentation techniques, which produce different and often complementary fragmentation [6].

Two-Dimensional (2D) FT-ICR MS allows data-independent fragmentation of all ions in a sample and correlation of precursor and fragment ions without prior isolation [7, 8, 9, 10]. This correlation is obtained through the modulation of precursor ions’ cyclotron radii prior to fragmentation using a series of rf-pulses. Pfandler et al. developed the first pulse sequence used for 2D MS experiments [7, 8]. A different pulse sequence called stored-waveform ion modulation (SWIM) was developed and applied by Ross et al. [11] and by van der Rest and Marshall [12]. Figure 1 shows the rf-pulses used in a typical 2D FT-ICR MS experiment. The ions enter the cell ideally at the center, and are then excited by an excitation pulse, P1. The ions rotate inside the cell according to their cyclotron frequency (ωc) for an encoding time (t1), to accumulate a phase of ωct1. An encoding pulse (P2), equal to P1, is then applied, and the ions are either excited further or de-excited, depending on their instantaneous phase relative to P2. At the end of P2, the ions’ cyclotron radii are modulated according to t1 and ωc [9]. P2 is followed by a fragmentation period (τm), in which the ions are subjected to a radius-dependent fragmentation, and produce fragment ions whose abundances depend on t1 and the precursor ions’ cyclotron frequency (ωc). After τm, P3 excites both precursors and fragment ions, they can then be detected, as per a normal FT-ICR MS experiment, producing a time domain transient (t2). A series of transients are recorded at regular increments of t1, and a 2D Fourier transform along t1 and t2 is performed: the output is a 2D map showing the correlations between fragments and precursors (Figure 2).
Figure 1

Typical pulse programme used for 2D FT-ICR MS experiments. In this work, IRMPD has been used for fragmentation during τm

Figure 2

Interpretation of a 2D mass spectrum. The autocorrelation line (\( y=x \)) shows the correlation of the precursor ion signals with their own cyclotron radius. Horizontal fragment ion spectra (\( y={m}_{\mathrm{precursor}} \)) show the fragmentation patterns of each precursor ion. Vertical precursor ion spectra (\( x={m}_{\mathrm{fragment}} \)) show the precursor ions of each fragment ion. Neutral-loss lines (\( y=x+{m}_{\mathrm{neutral}} \)) show the loss of neutrals by precursor ions

Figure 2 shows the schematic representation of a 2D mass spectrum. The y-axis corresponds to the m/z of the precursor ions, whereas the x-axis shows the m/z of the fragment ions. All the signals along the identity line (i.e., the line whose equation is y = x) represent the precursor species fragmented during the experiment in correlation with their own cyclotron frequency: for this reason, this is called autocorrelation line. All the signals along a horizontal line constitute the fragmentation patterns of each precursor. All the signals along a vertical line represent the precursor ions of a given fragment ion. Finally, in a 2D FT-ICR mass spectrum it is possible to extract lines parallel to the autocorrelation line, with a difference in intercept corresponding to the mass of a neutral molecule divided by the charge of the precursor (\( y=x+{m}_{neutral}/z \)). These are the so-called neutral-loss lines, and show the loss of neutrals by precursor ions. Vertical precursor ion spectra and neutral-loss spectra can be easily extracted from 2D maps, constituting information readily available only in 2D MS [13].

2D FT-ICR MS development started in the 1980s [7, 14], but it was hindered by insufficient computational capabilities: the acquisition of an individual transient for each increment of t1 quickly produces large amounts of data, which were challenging for the computers of that time. With recent improvements in computer technology and the improvements in ion fragmentation techniques for FT-ICR MS, an interest to develop 2D FT-ICR MS into an efficient analytical method was renewed [13]. The original 2D MS experiments were conducted using in-cell collisional dissociation [8], with a consistent loss of resolution resulting from fragmentation at a high orbital radius leading to magnetron expansion of the fragment cloud and ion loss. These in-cell collisions also required neutral gas pulses and significant pumping delays to return to optimum vacuum pressure prior to detection, limiting analysis. Fragmentation modes that do not involve the use of gases, such as infrared multi-photon dissociation (IRMPD) [15] or electron capture dissociation (ECD) [16], assist in the measurements of high-resolution 2D mass spectra by alleviating some of these limitations [17, 18]. An initial disadvantage 2D-FT-ICR MS, which further hindered development, was represented by the number of artefacts that can arise in 2D spectra. These artefacts can come from harmonics (in both dimensions) because of the 2D Fourier Transform [12] or from the fluctuation in the amplitude or frequency of a signal-scintillation noise, the equivalent of t1 noise in NMR spectroscopy [19]. Different algorithms for the de-noising of 2D mass spectra have been developed [20, 21] and are now applied to all contemporary 2D-FT-ICR MS spectra, and as a result scintillation noise no longer represents a significant problem to 2D-FT-ICR MS spectra. The potential of 2D FT-ICR MS as an analytical method for the structural analysis of small molecules has been recently demonstrated on cholesterol, showing the possibility for the technique to differentiate various fragmentation pathways and interrogate detected species [22]. 2D IRMPD FT-ICR MS has also been used for the analysis of peptides derived from tryptic digestion of commercially available collagen, which was not readily achievable with traditional 1D MS/MS [23] and for an ECD/IRMPD comparative study for proteomics, using digested cytochrome c as a model [24]. In this work, 2D IRMPD FT-ICR MS was used for the structural analysis of calmodulin (CaM) [25, 26], in both top-down and bottom-up protein tandem mass spectrometry approaches. CaM is a ubiquitous 16 kDa protein able to bind up to four calcium(II) atoms in the human body for signaling purposes. CaM has been used as a standard for deamidation [27] and to investigate the behavior of anti-cancer drugs at the atomic level [28, 29]. Its dimensions and relatively simple structure, given the absence of disulphide bonds, make it a good model to perform comparative top-down and bottom-up studies.

Experimental

Materials

Bovine calmodulin (CaM), trypsin (TPCK treated from bovine pancreas), and ammonium bicarbonate ((NH4)HCO3) were purchased from Sigma Aldrich (Dorset, UK). LCMS grade methanol and formic acid (HAc) were obtained from Fisher Scientific (Loughborough, UK). Water was purified by a Millipore Direct-Q purification system (Merck Millipore, Billerica, MA, USA).

Sample Preparation

Salt adducts were removed from CaM through 3 kDa molecular weight cut-off centrifugal filters (Amicon Ultra 0.5 mL; Merck Millipore Ltd., Tullagreen, Ireland). CaM was then passed through 0.22 μm centrifugal filters to remove particulates prior to nano-electrospray ionization. For the top-down approach, CaM (7.5 μM) was dissolved in a 75:25 water/methanol (v/v) solution with 0.3% (v/v) of formic acid. For the bottom-up approach, CaM (aqueous solution, 40 μM) was digested with trypsin in ammonium bicarbonate 100 mM at an enzyme-to-protein ratio of 1:40. A 3 kDa centrifugal filter device was used as a reactor at 37 °C, and the digestion was stopped after 4 h by isolation from trypsin through centrifugation of the peptides through the molecular weight cut-off filter. The tryptic-digest was diluted to an end concentration of ~10 μM using a 75:25 water/methanol (v/v) solution and acidified with 0.3% formic acid (v/v).

Methods

FT-ICR MS was performed on a 12 T Bruker SolariX FT-ICR Mass Spectrometer (Bruker Daltonik GmbH, Bremen, Germany) equipped with a custom nano-electrospray source (nESI). The source used 1.22 mm thin-walled glass capillaries (World Precision Instruments, Hitching, UK) pulled to obtain tips of ~1 μm orifice diameter with a flaming/brown micropipette puller (Sutter Instrument Co., Novato, CA, USA). The flow rate of the capillaries was ~2 μL/h in aqueous solution at a capillary voltage of 1000 V. Each experiment used 10–20 μL of sample. IRMPD fragmentation was achieved using a continuous wave, 25 W, CO2 laser (Synrad Inc., Mukilteo, WA, USA) held at 70% of its power output. IR photons were produced at a wavelength of 10.6 μm and pulsed into the ICR cell for 0.3 s prior to detection. The pulse program used for the 2D MS analysis was optimized in order to have the highest fragmentation efficiency according to previous studies [30].

For the top-down 2D FT-ICR MS, 512 scans (increments of t1) of 4 M (32-bit) data points were acquired over a mass range of m/z 883.5–3000 m/z on the vertical axis and m/z 147.5–3000 m/z on the horizontal axis; total time of acquisition was ~20 min.

For the bottom-up 2D FT-ICR MS, 4096 scans of 512 k data points were acquired over a mass range of m/z 368.2–3000 m/z on the vertical axis and m/z 147.5–3000 m/z on the horizontal axis; total time of acquisition was ~50 min. 2D FT-ICR data were processed with a custom program, SPIKE (Illkirch-Graffenstaden, France) [31], which uses urQRd as a de-noising algorithm [21]. Data for the 2D MS analysis were processed using a HP desktop computer (Hewlett Packard, Bracknell, UK), via a virtual machine that simulated a Ubuntu 64 bit operative system, with 8 GB of dedicated RAM and four core processors. For the top-down analysis, a rank of 6 was used for the urQRd denoising algorithm, whereas this value was equal to 10 for the bottom-up analysis. 1D IRMPD/FT-ICR MS/MS spectra of isolated precursor ions were acquired for comparison to the 2D method. All spectra were internally calibrated using known fragment peaks with a quadratic calibration function, and manually interpreted and assigned. For data interpretation of the 1D mass spectra, Data Analysis v 4.1 (Bruker Daltonik GmbH, Bremen, Germany) was used. The parameters used for the 1D and 2D mass spectrometric analyses of CaM are provided in the Supporting Information (Table S3).

Results

Top-Down Calmodulin (CaM)

Figure 3 shows the top-down 2D spectrum of CaM (Figure 3a), with the extraction of the autocorrelation line (Figure 3b), a precursor ion (vertical) scan (Figure 3c), and a fragment ion (horizontal) scan (Figure 3d). The 2D mass spectrum in Figure 3a shows a very clear autocorrelation line representing the correlation of the precursor ion signal with its own cyclotron radius. Figure 3b shows the extraction of the autocorrelation line (top) in comparison with a 1D full MS of CaM with the same solution conditions (bottom). Figure 3b shows a wide range of CaM charge states, from 8+ to 17+, as expected under the denaturing conditions used, and that 2D FT-ICR MS is able to detect and fragment them. The signal intensity of the autocorrelation line profile cannot be directly compared with that of the correlating 1D spectrum (Figure 3b, bottom, signal averaged for 100 scans) because in 2D FT-ICR MS intensities are not only proportional to the abundance of detected ions; they are also dependent on the change in precursor ion intensity during the fragmentation event. Lower intensity sodium/calcium adducted peaks are also visible in Figure 3b, showing an increase in intensity of the sodium ion-adducted peaks in the 2D FT-ICR MS compared with the 1D FT-ICR MS spectrum. As these two spectra were acquired on different days with different samples, it is important not to over-interpret the variation in sodium-adducted peak intensities, but it could indicate there is a greater relative variation (i.e., fragmentation) of salt-adducted peaks compared with the purely protonated species. Furthermore, because the 2D FT-ICR MS experiment requires a longer timeframe, it is possible that Na+ ions leach into solution from the glass capillary, which would increase the signal of the sodiated species in solution with time. Figure 3c shows a vertical precursor ion scan for the y 3 ion of CaM, revealing the precursor ions that generate the y 3 ion at m/z 319.198. In 2D FT-ICR MS, such a precursor ion scan is simultaneously detected and available for all fragments. Figure 3d shows the fragmentation pattern of the 14+ charge state of CaM. It can be seen that the spectrum is very dense, but 2D IRMPD/FT-ICR MS is able to isotopically resolve the fragment ion peaks. The unassigned peaks in spectrum Figure 3d may derive from secondary fragments and even “sequence scrambled” ions, as is commonly observed in CAD and/or IRMPD MS/MS [32]. Horizontal profiles have been extracted and analyzed for each charge state, giving a cleavage coverage of 23% for the top-down analysis at the specified conditions. Detailed peak assignment tables are included in the Supporting Information (Table S1). The dynamic range observed for the reported single scan 2D mass spectrum is ~200. Its value in 1D MS is ~1200, with 200 scans per mass spectrum.
Figure 3

Top-down analysis of CaM in denaturing conditions. The 2D mass spectrum (a) has been acquired with 512 scans of 4 M data points over a mass range of m/z 883.5–3000 on the vertical axis and m/z 147.5–3000 on the horizontal axis. Three ion scans, in different dimensions, are highlighted: (b) the autocorrelation line, in comparison with the full 1D mass spectrum of CaM at the same conditions; (c) a precursor ion (vertical) scan of the ion m/z 319.197597, indicating all the precursors generating the selected ion (information available only in 2D); (d) a fragment ion (horizontal) scan of the ion at charge state 14+, in comparison with an IRMPD FT-ICR MS/MS of the isolated ion with the same m/z. The spectrum (d) is divided in three spaces for clarity

Bottom-Up Calmodulin (CaM)

Figure 4a shows the 2D IRMPD-FT-ICR MS spectrum of the tryptic digest peptide mixture from CaM, with the extraction of the autocorrelation line (Figure 4b), a neutral-loss scan (Figure 4c), and a fragment ion (horizontal) scan (Figure 4d). Figure 4b shows the autocorrelation line profile (top) in comparison with the full 1D mass spectrum of digested CaM (signal averaged over 100 scans) (bottom). It can be seen from this comparison that 2D FT-ICR MS shows the same fragments present in the digest with few exceptions. A list of the assigned peaks is provided in the Supporting Information for the 2D analysis (Table S21). Figure 4c shows an example of neutral loss scan, the spectrum shows all the precursors that during fragmentation lose a neutral molecule whose absence would cause a decrease in m/z by 9. According to the neutral loss line equation, y = x + m neutral /z, for doubly charged ions, this loss would correspond to loss of a water molecule [M – H2O]2+. Figure 4c is therefore showing all the doubly charged ions in the mixture, which lose a water molecule during IRMPD fragmentation. A neutral-loss line showing water loss from doubly charged precursors is chosen as an example the capability of 2D-FT-ICR MS techniques: the same type of profile can be applied to other neutral losses, for example phosphate groups or glycans, respectively, from phosphopeptides and glycopeptides, and it constitutes information available readily in 2D FT-ICR MS. Figure 4d shows a fragment ion scan for the ion at m/z 782.380. The horizontal profile (top) is in direct comparison with the corresponding 1D IRMPD FT-ICR MS/MS of the same peptide (bottom). The figures show that results obtained with 2D IRMPD-FT-ICR MS are generally comparable with the 1D approach in terms of fragments shown and sequence coverage, but do not generate the same signal intensities for the different ions. The differences in ion abundances between the shown fragment ion scan and its homologous MS/MS spectrum derive from the different tuning of the two spectra. Even with the same tuning conditions, though, characteristics such as misalignment of the IR laser, the magnetron expansion of the ion cloud, and its own expansion due to modulation in 2D MS make that 2D spectra result different in abundances compared with 1D spectra tuned in a similar way. Furthermore, the parameters for 1D MS are chosen in order to obtain an optimal fragmentation of a single precursor ion at a time, assuming that the ions are constantly inside the laser beam before excitation for detection. In 2D MS, the pulse program is optimized in a way that the ion packets are modulated in order to be fragmented selectively with the variation of t 1 , causing their oscillation in and out of the laser beam area, and resulting in different fragmentation pattern intensities. The total sequence coverage of CaM from the detected and fragmented peptide obtained with the 2D IRMPD FT-ICR MS of tryptic-digested CaM was 22% (see Supporting Information Tables S21 to S26). The dynamic range observed for the reported single scan 2D mass spectrum is ~1000. The dynamic range observed in 1D MS for a 200 scans spectrum is ~300.
Figure 4

Bottom-up analysis of trypsin-digested CaM. The 2D mass spectrum (a) has been acquired with 4096 scans of 512 k data points over a mass range of m/z 368.2–3000 on the vertical axis and m/z 147.5–3000 on the horizontal axis. Three ion scans, in different dimensions, are highlighted: (b) the autocorrelation line, in comparison with the full 1D mass spectrum of CaM digest at the same conditions; (c) a neutral-loss line (information available only in 2D); (d) a fragment ion (horizontal) scan of the ion m/z 782.380488 in comparison with an IRMPD/FT-ICR MS/MS of the isolated ion with the same m/z

Discussion

2D IRMPD/FT-ICR MS spectra of CaM were acquired using a top-down and a bottom-up approach. The approaches differ greatly with respect to the type and amount of information they provide and need to be addressed individually when finding the optimal conditions for acquiring 1D and 2D FT-ICR mass spectra.

In every mass spectrometric technique, resolution is influenced, among other factors, by the number of dataset size during the acquisition. In 2D FT-ICR MS, the output is a 2D map with m/z along both the axis, and as a result it is possible to define the resolution for both dimensions, vertical and horizontal. Since the y-axis shows the m/z of ions in the mixture prior to fragmentation (i.e., the precursor ions), the vertical resolution is analogous to the ability to separate two precursor ion peaks, whereas since the x-axis shows the m/z of the fragment ions generated by each precursor, the horizontal resolution is related to the separation of narrow fragment ion peaks. Due to the nature of the 2D Fourier transform, the horizontal resolution depends on the number of data points used to acquire t 2 , whereas the vertical resolution depends on the number of iterations in t 1 (scans). As it can be seen in Figure 1, t 2 is the length of the transient acquired after the fragmentation occurs. Thus, t2 is dictated by the number of data points used for each acquisition and the sampling frequency required, and is easily variable, as for all FTMS acquisitions; t 1 , as described before, is the delay between the identical initial pulses P 1 and P 2 . The starting value of t 1 is dictated by the precursor ions of interest, and the number of increments is dictated by the vertical resolution needed to effectively separate those precursors during the subsequent 2D-FT-ICR MS experiment.

In the case of the bottom-up procedure, the tryptic digest solution presents a complex mixture of peptides, whose precursor m/z peaks must be resolved in order to be able to distinguish different precursor peptides and overlapping distributions (Figure 4b). The fragmentation of peptides does not generate spectra as dense in peaks as those generated by the fragmentation of whole proteins. As can be seen in Figure 4c, it is possible to resolve the generated peptide fragment peaks with a relatively low horizontal resolution. However, the increase in sample complexity requires a higher vertical resolution in order to resolve overlapping peaks. In this case, 4096 data points were used for the acquisition in the vertical dimension, eight times the amount of data points used for the same acquisition in the top-down approach (see below). Since the resolution is proportional to the number of data points used for the acquisition, and that in the vertical dimension this value is determined by the number of t 1 time steps, the acquisition of the 2D mass spectrum of digested/complex mixtures would take more time than the top-down experiment because of increased sample complexity. The mass resolving powers for the bottom-up 2D mass spectrum of CaM were RPy = 58766 and an RPx = 1191 at m/z 400.

The top-down analysis of CaM showed an initial full MS (Figure 3b) presenting peaks of different charge states clearly separated in m/z, with no overlapping distributions between charge states. Therefore, vertical resolution needed to be able to separate two narrow peaks in this case would be relatively low, especially compared with the bottom-up procedure (above). For this reason, only 512 t 1 increments (time steps/data points) were required to ensure the effective separation of precursors. The converse applies for the horizontal (MS/MS) scans: after IRMPD fragmentation, the density of the spectrum (Figure 3c) increases significantly compared with the peptide spectra above, requiring a higher horizontal resolution to separate different fragments and provide accurate assignments. In this case, 4 M data points were used for the acquired transient, corresponding to a horizontal resolution of 420,000 at 400 m/z. Generation of very dense spectra is typical of the top-down approach, which benefits from the resolution and mass accuracy of FT-ICR MS for accurate, resolved assignment. The mass resolving power was measured using the FWHM reported to m/z 400. For the top-down 2D mass spectrum of CaM, the results were: RPy (vertical) of 182 and an RPx (horizontal) of 420,000.

Owing to the nature of the 2D map produced in 2D FT-ICR MS, it is possible to define a composite resolution, which takes into account the vertical and horizontal dimensions. In effect the “true resolution” of a 2D MS mass spectrum is the product of the resolution in both dimensions, and is approximately equal to the theoretical peak capacity of the experiment. According to this line of reasoning, the 2D mass resolving powers for the previous spectra are the following: 182 (RPy) × 420,000 (RPx) = RP2D ≈ 77 M for the top-down analysis of CaM and 59,000 (RPy) × 1200 (RPx) RP2D ≈ 70 M for the bottom-up procedure, both at m/z 400.

The results obtained with 2D FT-ICR MS for both the top-down and the bottom-up analyses of CaM are comparable to those obtained with standard 1D FT-ICR MS, with a consistent saving in time and sample consumption. For the bottom-up approach, the 2D FT-ICR MS experiment was remarkably faster than a comparative analysis in 1D-MS would be; isolating, fragmenting, and detecting each individual precursor observed in the 1D spectrum of the tryptic digest of CaM could require several hours in comparison. 1D top-down and bottom-up studies on CaM using IRMPD as fragmentation techniques were performed for comparison to the 2D IRMPD FT-ICR MS analysis, showing cleavage coverages of the protein up to 36% for the top-down approach, and about 25% for the bottom-up approach, for a total cleavage coverage of 52% (see Supporting Information, Figure S1). For the 2D FT-ICR MS experiments, the total cleavage coverage obtained with both the approaches reaches ~40% comparable to those obtained in 1D FT-ICR MS. Since precursors have different fragmentation efficiencies, the limitation of using a set fragmentation parameter combination for all charge states is not ideal and could account for the differences between the 2D FT-ICR MS data and individually optimized 1D data. Similar issues arise for LC-MS/MS experiments.

The experiments performed in this work constitute the first example of 2D FT-ICR MS for top-down proteomics, and show that top-down analysis can be performed in 2D MS. In this case, CaM is used as a model for its structure and size, and experiments performed with such a protein constitute the initial step for top-down analysis of complex mixtures. In fact, although the sample used for this analysis was relatively pure, once the procedure is developed, it is possible to apply top-down 2D MS on non-pure/complex mixture samples. In such samples, depending on the complexity of the mixture, isolating a protein for standard MSn analysis might be difficult if not impossible without contamination from ions close in m/z. As mentioned, in traditional MS/MS the quality of the spectra depends on the resolution obtained with quadrupole isolation of single precursor ions. Narrowing the isolation window in order to include a single ion (e.g., a single charge state of a protein of interest) increases the isolation resolution required, but working with narrow isolation windows in a quadrupole (usually less than 5 Da) inevitably brings substantial signal losses that negatively affect the sensitivity of the final MS/MS spectrum. On the other hand, 2D FT-ICR MS provides the fragmentation patterns of all the ions in the mixture without the necessity of quadrupole isolation, preventing the loss of resolution in the precursor dimension even in the case of complex mixtures, as it is demonstrated in this study for the bottom-up procedure. The resolution in the vertical dimension is higher than the resolution typically obtainable with quadruple isolation, and in 2D FT-ICR MS, the vertical resolution is primarily limited by the number of t1 steps (vertical scans) acquired, and by the computational capacity available.

2D FT-ICR MS presents particular computational challenges. 2D IRMPD FT-ICR MS of CaM resulted in files of 23.5 GB for both approaches, a challenging file size for processing and denoising with standard desktop computers like the one used for this work. 2D NMR, by comparison, usually works with files 103 times smaller [33]. The choice of denoising rank is roughly proportional to the number of precursors in the spectrum, and the lower the rank the more the spectrum is denoised [21]. Denoising a mass spectrum highly increases the data processing time (from about 4 h- to ~12 h for the shown 2D spectra in the described work station system), but it is essential for effective visualization and interpretation. 2D FT-ICR MS development has been hindered until recently because of limited computational capabilities, and computer technology still represents a significant limitation for the technique, but one that is changing rapidly. The current computational approach to processing this data involves parallelizing the 2DFT, denoising, and visualization code, which will greatly alleviate the current computational bottleneck.

Conclusions

2D IRMPD/FT-ICR MS has been used for the top-down and bottom-up analysis of CaM, and fragmentation patterns of CaM have been obtained with both approaches. The results obtained are comparable with 1D FT-ICR MS/MS and the overall procedure is lower time- and sample-consuming than traditional tandem mass spectrometry. In 2D MS more information is available at once through the observation of the neutral-loss lines and the vertical ion scans. 2D FT-ICR MS can be used for the identification of proteins in complex mixtures, although the technique could be improved for top-down analysis. The main limitation is currently represented by computational capacities.

Notes

Acknowledgments

The authors thank Dr. Andrea Clavijo-Lopez, Ms. Hayley Simon, and Ms. Alice M. Lynch for their helpful discussions and support. The authors acknowledge the University of Warwick for the opportunity to do these experiments, and the Engineering and Physical Sciences Research Council for funding (EP/J000302/1).

Supplementary material

13361_2016_1431_MOESM1_ESM.docx (102 kb)
ESM 1 (DOCX 102 kb)

References

  1. 1.
    Marshall, A.G., Hendrickson, C.L., Jackson, G.S.: Fourier transform ion cyclotron resonance mass spectrometry: a primer. Mass Spectrom. Rev. 17, 1–35 (1998)CrossRefGoogle Scholar
  2. 2.
    Bracewell, R.N.: The fourier transform and its applications. McGraw-Hill, New York (2000)Google Scholar
  3. 3.
    Marshall, A.G., Verdun, F.R.: Fourier transforms in NMR, optical, and mass spectrometry a user’s handbook. Elsevier B.V., Amsterdam (1990)Google Scholar
  4. 4.
    Qi, Y., O’Connor, P.B.: Data processing in Fourier transform ion cyclotron resonance mass spectrometry. Mass Spectrom. Rev. 33, 333–352 (2014)CrossRefGoogle Scholar
  5. 5.
    Zhang, L.-K., Rempel, D., Pramanik, B.N., Gross, M.: Accurate mass measurements by Fourier transform mass spectrometry. Mass Spectrom. Rev. 24, 286–309 (2005)CrossRefGoogle Scholar
  6. 6.
    Hogan, J.M., Pitteri, S.J., Chrisman, P.A., Mcluckey, S.A.: Complementary structural information from a tryptic N-linked glycopeptide via electron transfer ion/ion reactions and collision-induced dissociation. J. Proteome Res. 4, 628–632 (2005)CrossRefGoogle Scholar
  7. 7.
    Pfandler, P., Bodenhausen, G., Rapin, J., Houriet, R., Gaumann, T.: Two-dimensional Fourier transform ion cyclotron resonance mass spectrometry. Chem. Phys. Lett. 138, 195–200 (1987)CrossRefGoogle Scholar
  8. 8.
    Pfandler, P., Gaumann, T.: Broad-band two-dimensional fourier transform ion cyclotron resonance. J. Am. Chem. Soc. 110, 5625–5628 (1988)CrossRefGoogle Scholar
  9. 9.
    Guan, S., Jones, P.: A theory for two-dimensional Fourier-transform ion cyclotron resonance mass spectrometry. J. Chem. Phys. 91, 5291–5295 (1989)CrossRefGoogle Scholar
  10. 10.
    Bensimon, M.: A method to generate phase continuity in two-dimensional Fourier transform ion cyclotron resonance mass spectrometry. Chem. Phys. Lett. 157, 97–100 (1989)CrossRefGoogle Scholar
  11. 11.
    Ross, C.W., Guan, S., Grosshans, P.B., Ricca, T.L., Marshall, A.G.: Two-dimensional Fourier transform ion cyclotron resonance mass spectrometry/mass spectrometry with stored-waveform ion radius modulation. J. Am. Chem. Soc. 115, 7854–7861 (1993)CrossRefGoogle Scholar
  12. 12.
    van der Rest, G., Marshall, A.G.: Noise analysis for 2D tandem Fourier transform ion cyclotron resonance mass spectrometry. Int. J. Mass Spectrom. 210/211, 101–111 (2001)CrossRefGoogle Scholar
  13. 13.
    van Agthoven, M.A., Delsuc, M.-A., Bodenhausen, G., Rolando, C.: Towards analytically useful two-dimensional Fourier transform ion cyclotron resonance mass spectrometry. Anal. Bioanal. Chem. 405, 51–61 (2013)CrossRefGoogle Scholar
  14. 14.
    Marshall, A.G., Wang, L.I.N., Lebatuan, T.: Ion cyclotron resonance excitation/de-excitation: a basis for stochastic Fourier transform ion cyclotron resonance mass spectrometry. Chem. Phys. Lett. 105, 233–236 (1984)CrossRefGoogle Scholar
  15. 15.
    Little, D.P., Speir, J.P., Senko, M.W., O’Connor, P.B., McLafferty, F.W.: Infrared multiphoton dissociation of large multiply charged ions for biomolecule sequencing. Anal. Chem. 66, 2809–2815 (1994)CrossRefGoogle Scholar
  16. 16.
    Zubarev, R., Kelleher, N.L., McLafferty, F.W.: Electron capture dissociation of multiply charged protein cations. A nonergodic process. J. Am. Chem. Soc. 120, 3265–3266 (1998)CrossRefGoogle Scholar
  17. 17.
    van Agthoven, M.A., Delsuc, M.-A., Rolando, C.: Two-dimensional FT-ICR/MS with IRMPD as fragmentation mode. Int. J. Mass Spectrom. 306, 196–203 (2011)CrossRefGoogle Scholar
  18. 18.
    van Agthoven, M.A., Chiron, L., Coutouly, M.-A., Delsuc, M.-A., Rolando, C.: Two-dimensional ECD FT-ICR mass spectrometry of peptides and glycopeptides. Anal. Chem. 84, 5589–5595 (2012)CrossRefGoogle Scholar
  19. 19.
    Morris, G.A.: Systematic sources of signal irreproducibility and t1 noise in high-field NMR spectrometers. J. Magn. Reson. 100, 316–328 (1992)Google Scholar
  20. 20.
    van Agthoven, M.A., Coutouly, M.-A., Rolando, C., Delsuc, M.-A.: Two-dimensional Fourier transform ion cyclotron resonance mass spectrometry: reduction of scintillation noise using Cadzow data processing. Rapid Commun. Mass Spectrom. 25, 1609–1616 (2011)CrossRefGoogle Scholar
  21. 21.
    Chiron, L., van Agthoven, M.A., Kieffer, B., Rolando, C., Delsuc, M.-A.: Efficient denoising algorithms for large experimental datasets and their applications in Fourier transform ion cyclotron resonance mass spectrometry. Proc. Natl. Acad. Sci. U. S. A. 111, 1385–1390 (2014)CrossRefGoogle Scholar
  22. 22.
    van Agthoven, M.A., Barrow, M.P., Chiron, L., Coutouly, M.-A., Kilgour, D., Wootton, C.A., Wei, J., Soulby, A., Delsuc, M.-A., Rolando, C., O’Connor, P.B.: Differentiating fragmentation pathways of cholesterol by two-dimensional fourier transform ion cyclotron resonance mass spectrometry. J. Am. Soc. Mass Spectrom. 26, 2105–2114 (2015)CrossRefGoogle Scholar
  23. 23.
    Simon, H., van Agthoven, M.A., Lam, P.Y., Floris, F., Chiron, L., Delsuc, M.-A., Rolando, C., Barrow, M., O’Connor, P.: Uncoiling Collagen: a multidimensional mass spectrometry study. Analyst (2015). doi: 10.1039/C5AN01757B Google Scholar
  24. 24.
    van Agthoven, M.A., Wootton, C.A., Chiron, L., Coutouly, M., Soulby, A., Wei, J., Barrow, M.P., Delsuc, M., Rolando, C., O'Connor, P.B.: Two-dimensional mass spectrometry for proteomics, a comparative study with cytochrome c. Anal. Chem. 88, 4409–4417 (2016)CrossRefGoogle Scholar
  25. 25.
    Stevens, F.C.: Calmodulin: an introduction. Can. J. Biochem. Cell Biol. 61, 906–910 (1982)CrossRefGoogle Scholar
  26. 26.
    Means, A.D.: Calmodulin: an intracellular calcium receptor. Nature 285, 73–77 (1980)CrossRefGoogle Scholar
  27. 27.
    Soulby, A.J., Heal, J.W., Barrow, M.P., Roemer, R.A., O’Connor, P.B.: Does deamidation cause protein unfolding? A top-down tandem mass spectrometry study. Protein Sci. 24, 850–860 (2015)CrossRefGoogle Scholar
  28. 28.
    Li, H., Lin, T.Y., Van Orden, S.L., Zhao, Y., Barrow, M.P., Pizarro, A.M., Qi, Y., Sadler, P.J., O’Connor, P.B.: Use of top-down and bottom-up fourier transform ion cyclotron resonance mass spectrometry for mapping calmodulin sites modified by platinum anticancer drugs. Anal. Chem. 83, 9507–9515 (2011)CrossRefGoogle Scholar
  29. 29.
    Li, H., Wells, S.A., Jimenez-Roldan, J.E., Römer, R.A., Zhao, Y., Sadler, P.J., O’Connor, P.B.: Protein flexibility is key to cisplatin crosslinking in calmodulin. Protein Sci. 21, 1269–1279 (2012)CrossRefGoogle Scholar
  30. 30.
    van Agthoven, M.A., Chiron, L., Coutouly, M.-A., Sehgal, A.A., Pelupessy, P., Delsuc, M.-A., Rolando, C.: Optimization of the discrete pulse sequence for two-dimensional FT-ICR mass spectrometry using infrared multiphoton dissociation. Int. J. Mass Spectrom. 370, 114–124 (2014)CrossRefGoogle Scholar
  31. 31.
    SPIKE: Spectrometry Processing Innovative KErnel. Available at: https://bitbucket.org/delsuc/spike. Accessed May 2015
  32. 32.
    Paizs, B., Suhai, S.: Fragmentation pathways of protonated peptides. Mass Spectrom. Rev. 24, 508–548 (2005)CrossRefGoogle Scholar
  33. 33.
    Kumar, A., Ernst, R.R., Wüthrich, K.: A two-dimensional nuclear Overhauser enhancement (2D NOE) experiment for the elucidation of complete proton-proton cross-relaxation networks in biological macromolecules. Biochem. Biophys. Res. Commun. 95, 1–6 (1980)CrossRefGoogle Scholar

Copyright information

© American Society for Mass Spectrometry 2016

Authors and Affiliations

  • Federico Floris
    • 1
  • Maria van Agthoven
    • 1
  • Lionel Chiron
    • 2
  • Andrew J. Soulby
    • 1
  • Christopher A. Wootton
    • 1
  • Yuko P. Y. Lam
    • 1
  • Mark P. Barrow
    • 1
  • Marc-André Delsuc
    • 2
    • 3
  • Peter B. O’Connor
    • 1
  1. 1.University of WarwickCoventryUK
  2. 2.CASC4DEIllkirch-GraffenstadenFrance
  3. 3.IGBMCIllkirch-GraffenstadenFrance

Personalised recommendations