A feasibility study for using ABS plastic and a low-cost 3D printer for patient-specific brachytherapy mould design

  • Benjamin D. Harris
  • Sanna Nilsson
  • Christopher M. Poole
Scientific Paper

Abstract

This feasibility study aims to determine if a low-cost 3D printer (BitsFromBytes 3D Touch) with ABS plastic can print custom mould structures and catheter channels defined in a brachytherapy treatment planning system (Nucletron Oncentra) for patient-specific treatment. Printer accuracy was evaluated through physical measurement, and print quality was investigated by adjusting print parameters (print speed, layer thickness, percentage infill). Catheter positioning and reproducibility were measured over repeated insertions. ABS plastic water equivalency was investigated by comparing Ir-192 HDR source dose distributions, measured with radiochromic film, in ABS plastic and in water. Structures and catheter channels were printed accurately to within 0.5 mm laterally and 1 mm in the vertical print direction. Adjusting print parameters could reduce print time, albeit with reduced print quality. 3.5 mm channel diameters allowed for easy catheter insertion. Catheter positioning was reproducible to within 0.5 mm but, because of catheter flex within the channel, was on average 1 mm offset from defined TPS positions. This offset could be accounted for by repeating the treatment planning CT scan with the printed mould positioned on the patient. Dose attenuation in ABS plastic and in water was equivalent to within the measurement limitations. While clinical uses for this particular low-cost printer and ABS plastic are limited by print size restrictions and non-certification for biocompatibility, it has been demonstrated that a low-cost 3D printer set-up can accurately create custom moulds and catheter channels potentially acceptable for clinical use.

Keywords

3D printing Brachytherapy Customised moulds Patient-specific treatment 

References

  1. 1.
    Gerbaulet A, Pötter R, Mazeron J-J, Meertens H, Van Limbergen E (eds) (2002) The GEC ESTRO handbook of brachytherapy. ACCO, Leuven, pp 313–315Google Scholar
  2. 2.
    De Boeck L, Beliën J, Egyed W (2014) Dose optimization in HDR brachytherapy: a literature review of quantitative models from 1990 to 2010. Oper Res Health Care 3(2):80–90CrossRefGoogle Scholar
  3. 3.
    Shwetha B, Ravikumar M, Katke A, Supe SS, VenkataGiri G, Ramanand N, Pasha T (2010) Dosimetric comparison of various optimization techniques for high dose rate brachytherapy of interstitial cervix implants. J Appl Clin Med Phys 11(3):225–230Google Scholar
  4. 4.
    Gerbaulet A, Pötter R, Mazeron J-J, Meertens H, Van Limbergen E (eds) (2002) The GEC ESTRO handbook of brachytherapy. ACCO, Leuven, pp 316–318Google Scholar
  5. 5.
    Gerbaulet A, Pötter R, Mazeron J-J, Meertens H, Van Limbergen E (eds) (2002) The GEC ESTRO handbook of brachytherapy. ACCO, Leuven, p 578Google Scholar
  6. 6.
    Guix B, Finestres F, Tello JI, Palma C, Martinez A, Guix JR, Guix R (2000) Treatment of skin carcinomas of the face by high-dose-rate brachytherapy and custom-made molds. Int J Radiat Oncol Biol Phys 47(1):95–102CrossRefPubMedGoogle Scholar
  7. 7.
    The Economist (2013) 3D printing scales up. The Economist Newspaper.http://www.economist.com/news/technology-quarterly/21584447-digital-manufacturing-there-lot-hype-around-3d-printing-it-fast.. Accessed 22 Sept 2014
  8. 8.
    Cumming I, Joshi CP, Lasso A, Rankin A, Falkson C, John Schreiner L, Fichtinger G (2014) SU-E-T-04: 3D printed patient-specific surface mould applicators for brachytherapy treatment of superficial lesions. Med Phys 41:222CrossRefGoogle Scholar
  9. 9.
    Cunha J, Sethi R, Mellis K, Siauw T, Sudhyadhom A, Hsu I, Pouliot J (2014) WE-F-16A-01: commissioning and clinical use of PC-ISO for customized, 3D printed, gynecological brachytherapy applicators. Med Phys 41:514CrossRefGoogle Scholar
  10. 10.
    Cunha JAM, Mellis K, Sethi R, Siauw T, Sudhyadhom A, Garg A, Goldberg K, Hsu I-C, Pouliot J (2015) Evaluation of PC-ISO for customised, 3D printed, gynecologic 192IR HDR brachytherapy applicators. JACMP 16(1):246–253Google Scholar
  11. 11.
    Poulin E, Gardi L, Fenster A, Pouliot J, Beaulieu L (2014) A novel approach for real-time, personalized breast HDR brachytherapy treatment using 3D printing technology. Brachytherapy 13(S1):S18CrossRefGoogle Scholar
  12. 12.
    Garg A, Patil S, Siauw T, Cunha AM, Hsu IC, Abbeel P, Pouliot J, Goldberg K (2013) An algorithm for computing customised 3D printed implants with curvature constrained channels for enhancing intracavitary brachytherapy radiation delivery. IEEE Int Conf Autom Sci Eng (CASE) 2013:466–473Google Scholar
  13. 13.
    Stratasys Inc. (2012) Form 10-K: Annual Report.http://investors.stratasys.com/secfiling.cfm?filingID=1206774-12-1018&CIK=915735. Accessed 14 Feb 2015
  14. 14.
    BitsFromBytes/3D Systems Inc. (2011) 3D Touch: Make personal colour manufacturing a reality.http://cubify.s3.amazonaws.com/public/bfb/0911_touch_uk_3ds.pdf. 14 Feb 2015
  15. 15.
    Gibson I, Rosen DW, Stucker B (2010) Additive manufacturing technologies: rapid prototyping to direct digital manufacturing. Springer, New York, pp 143–156CrossRefGoogle Scholar
  16. 16.
    Chilson, L (2013) The difference between ABS and PLA for 3D printing. ProtoParadigm,http://www.protoparadigm.com/news-updates/the-difference-between-abs-and-pla-for-3d-printing/. Accessed 10 Feb 2015
  17. 17.
    RapMan Australia (n.d.) Materials: ABS (Acrylonitrile Butadiene Styrene).http://www.rap-man.com.au/Material_files/ABSMaterial.html. Accessed 10 Feb 2015
  18. 18.
    White, D (n.d.) Health and Safety and sample risk assessment for RapMan and BfB3000 3D printers.http://www.rap-man.com.au/Downloads/Generic_Health_and_Safety_V3.pdf. Accessed 10 Feb 2015
  19. 19.
    3D Systems (2013) Safety data sheet: cube ABS plastic/CubeX ABS plastic.http://cubify.s3.amazonaws.com/public/sds/cube_cubex_sds_abs_english.pdf. Accessed 10 Feb 2015
  20. 20.
    Turk, G (1994) The PLY polygon file format. The Board of Trustees of The Leland Stanford Junior University.http://www.dcs.ed.ac.uk/teaching/cs4/www/graphics/Web/ply.html. Accessed 3 Feb 2015
  21. 21.
    Gibson I, Rosen DW, Stucker B (2010) Additive manufacturing technologies: rapid prototyping to direct digital manufacturing. Springer, New York, pp 341–343CrossRefGoogle Scholar
  22. 22.
    3DVerkstan (2015). A visual ultimaker troubleshooting guide.http://support.3dverkstan.se/article/23-a-visual-ultimaker-troubleshooting-guide. Accessed 14 Feb 2015
  23. 23.
    3D Systems Inc. (2011) 3D Touch.http://cubify.s3.amazonaws.com/public/bfb/0911_touch_uk_3ds.pdf. Accessed 14 Feb 2015
  24. 24.
    Stratasys Inc. (2009) FORTUS 360mc/400mc ACCURACY STUDY.http://www.stratasys.com/resources/~/media/748AD69721184448AF46F0C37BBD205B.pdf. Accessed 14 Feb 2015
  25. 25.
    3DVerkstan (2015) Getting better prints.http://support.3dverkstan.se/article/30-getting-better-prints. Accessed 14 Feb 2015
  26. 26.
    Airwolf 3D Printers (2015) Airwolf 3D FAQ. http://airwolf3d.com/faq/.Accessed 14 Feb 2015
  27. 27.
    Materialise (n.d.) Magics: Premier Software for Additive Manufacturing Professionals. http://software.materialise.com/sites/default/files/public/SAM/Products/Magics/broch_magics_3fold_a4_eng.pdf. Accessed 14 Feb 2015
  28. 28.
    Kumar R, Sharma SD, Despande S, Ghadi Y, Shaiju VS, Amols HI, Mayya YS (2009) Acrylonitrile butadiene styrene (ABS) plastic-based low cost tissue equivalent phantom for verification dosimetry in IMRT. JACMP 11(1):24–32Google Scholar
  29. 29.
    Rivard MJ, Coursey BM, DeWerd LA, Hanson WF, Saiful Huq M, Ibbott GS, Mitch MG, Nath R, Williamson JF (2004) Update of AAPM task group No. 43 report: a revised AAPM protocol for brachytherapy dose calculations. Med Phys 31(3):633–674Google Scholar
  30. 30.
    IC3D, LLC. (2014) 3D Printing Tips & Tricks.http://www.ic3dprinters.com/3d-printing-tips-tricks/. Accessed 14 Feb 2015
  31. 31.
    BitsFromBytes/3D Systems Inc. (2011) 3D touch: set up and operations manual: 1.5.5 Print Material.http://cubify.s3.amazonaws.com/public/bfb/3dtouch_setup_operations_manual.pdf. Accessed 14 Feb 2015
  32. 32.
    Frisch Labor Systems (n.d.) Online Shop: Impression technique: Fricotan A + B. http://www.frisch-labor-systems.de/index.php?201010-FRICOTAN-A—B5062f30df3807. Accessed 14 Feb 2015
  33. 33.
    Stratasys Inc. (2015) PC-ISO: Create biocompatible parts with superior strength.http://www.stratasys.com/materials/fdm/pc-iso. Accessed 14 Feb 2015
  34. 34.
    Stratasys Inc. (2015) ABS-M30i: 3D print biocompatible, sterilizable parts.http://www.stratasys.com/materials/fdm/abs-m30i. Accessed 14 Feb 2015

Copyright information

© Australasian College of Physical Scientists and Engineers in Medicine 2015

Authors and Affiliations

  • Benjamin D. Harris
    • 1
  • Sanna Nilsson
    • 2
  • Christopher M. Poole
    • 3
  1. 1.Radiation OncologyPrincess Alexandra HospitalBrisbaneAustralia
  2. 2.Cancer Care ServicesRoyal Brisbane and Women’s HospitalBrisbaneAustralia
  3. 3.Department of Obstetrics & GynaecologyUniversity of Melbourne, Royal Women’s HospitalMelbourneAustralia

Personalised recommendations