Global Water Flow Measurement and Calibration Facilities: Review of Methods and Instrumentations

  • Shiv Kumar Jaiswal
  • Sanjay Yadav
  • A. K. Bandyopadhyay
  • Ravinder Agarwal
Review Paper

Abstract

Water Flow Calibration Facilities (WFCFs) are widely used for the calibration of water flowmeters all over the world. Last two decades have witnessed significant research contributions in the development of instrumentation and techniques with special emphasis on reducing the measurement errors and improving the measurement uncertainty in flow calibration. It is difficult to find a consolidated report/review of such developments for the new entrants/researchers/metrologists in this field. Therefore, authors considered that it is high time to review the literature and compile a concise report to this effect. Present paper is an attempt to publish terse review on the historical background, recent developments and current status of WFCF and related works. The work going on at National Physical Laboratory, India, in this direction is also briefly described.

Keywords

Water flow calibration facility (WFCF) Weighing method Volumetric method Measurement error and uncertainty Diverter Calibration and measurement capability (CMC) 

References

  1. [1]
    J.F. Ruff, Flow Measurement: History, in Encyclopedia of Water Science by S.W. Trimble, 2nd Ed., CRC Press, Boca Raton, (2008) pp 388–393.Google Scholar
  2. [2]
    ISO 4185, Measurement of Liquid Flow in Closed Conduits—Weighing Methods, 1st Ed., ISO, Charlottesville, (1980) pp. 1–21.Google Scholar
  3. [3]
    ISO 9368-1, Measurement of Liquid Flow in Closed Conduits by Weighing Methods—Procedure for Checking the Installations, Part 1: Static Weighing System, 1st Ed., ISO, Charlottesville, (1990) pp. 1–22.Google Scholar
  4. [4]
    ISO 5168, Measurement of Liquid Flow in Closed Conduits—Procedure for the Evaluation of Uncertainties, 2nd Ed., ISO, Charlottesville, (2005) pp. 1–65.Google Scholar
  5. [5]
    JCGM 100, Evaluation of measurement data – Guide to the expression of uncertainty in measurement, 1st Ed., JCGM, Viana Do Castelo, (2008) pp. 1–134.Google Scholar
  6. [6]
    ISO 4064, Measurement of Water Flow in Fully Charged Closed Conduits—Meters for Cold Potable Water and Hot Water—Part 1: Specifications, 3rd Ed., ISO, Charlottesville, (2005) pp. 1–42.Google Scholar
  7. [7]
    ISO 4064, Measurement of Water Flow in Fully Charged Closed Conduits—Meters for Cold Potable Water and Hot Water—Part 2: Installation Requirements, 3rd Ed., ISO, Charlottesville, (2005) pp. 1–10.Google Scholar
  8. [8]
    ISO 4064, Measurement of Water Flow in Fully Charged Closed Conduits—Meters for Cold Potable Water and Hot Water—Part 3: Test Methods and Equipment, 3rd Ed., ISO, Charlottesville, (2005) pp. 1–71.Google Scholar
  9. [9]
    M. Benkova, S. Makovnik, I. Mikulecky and V. Zamecnik, Bell Prover—Calibration and Monitoring of Time Stability, MAPAN J. Metrol. Soc. India, 26 (2011) 165–171.Google Scholar
  10. [10]
    C.M. Su, W.T. Lin and S. Masri, Establishment and Verification of a Primary Low-Pressure Gas Flow Standard at NIMT, MAPAN-J. Metrol. Soc. India, 26 (2011) 173–186.Google Scholar
  11. [11]
    A.N. Johnson, C.J. Crowley and T.T. Yeh, Uncertainty Analysis of NIST’s 20 Liter Hydrocarbon Liquid Flow Standard, MAPAN-J. Metrol. Soc. India, 26 (2011) 187–202.Google Scholar
  12. [12]
    C. Li and A.N. Johnson, Bilateral Comparison between NIM’s And NIST’s Gas Flow Standards, MAPAN-J. Metrol. Soc. India, 26 (2011) 211–224.Google Scholar
  13. [13]
    M. Carter, W. Johansen and C. Britton, Performance of a Gas Flow Meter Calibration System Utilizing Critical Flow Venturi Standards, MAPAN-J. Metrol. Soc. India, 26 (2011) 247–254.Google Scholar
  14. [14]
    M. Bair, The Dissemination of Gravimetric Gas Flow Measurements through an LFE Calibration Chain, Proceedings of NCSLI 1999 Workshop and Symposium, Charlotte, July 13, 1999.Google Scholar
  15. [15]
  16. [16]
  17. [17]
  18. [18]
  19. [19]
  20. [20]
  21. [21]
    C. David and P. Claudel, Novel Water Flow Facility in France Range Extension to Low Flow Rates (10 000 ml/h down to 1 ml/h), MAPAN-J. Metrol. Soc. India, 26 (2011) 203–209.Google Scholar
  22. [22]
  23. [23]
    NIST Special Publication 250, NIST Calibration Services for Water Flowmeters, Water Flow Calibration Facility, NIST, Gaithersburg, August 2006.Google Scholar
  24. [24]
    T.T. Yeh, N.P. Yende and P.I. Espina, Theoretical Self-Error Cancelling Diverters for Liquid Flow Calibration Facilities, Proceedings of the 11th International Conference on Flow Measurement (FLOMEKO 2003), Groningen, May 12–14, 2003.Google Scholar
  25. [25]
    R. Engel and U. Klages, A novel approach to improve diverter performance in liquid flow calibration facilities, Proceedings of the 10th International Conference on Flow Measurement (FLOMEKO 2000), Salvador, June 5–8, 2000.Google Scholar
  26. [26]
    R. Engel, Dynamic Weighing—Improvements in Gravimetric Liquid Flowmeter Calibration, Proceedings of the 5th International Symposium on Fluid Flow Measurement, Arlington, April 2002.Google Scholar
  27. [27]
    I.I. Shinder and M.R. Moldover, Accurately Measuring Unsteady Water Flow Using a Dynamic Standard, Proceedings of Measurement Science Conference (MSC-2009), Anaheim, March 23–27, 2009.Google Scholar
  28. [28]
    I.I. Shinder and M.R. Moldover, Dynamic Gravitational Standard for Liquid Flow: Model and Measurements, Proceedings of the 7th International Symposium for Fluid Flow Measurement, Fairbanks, August 10–14, 2009.Google Scholar
  29. [29]
    I.I. Shinder and M.R. Moldover, Feasibility of an Accurate Dynamic Standard for Water Flow, Flow Meas. Instrum., 21 (2010) 128–133.CrossRefGoogle Scholar
  30. [30]
    W. Poeschel and R. Engel, The Concept of a New Primary Standard for Liquid Flow Measurement at PTB, Braunschweig, Proceedings of the 9th International Conference on Flow Measurement (FLOMEKO ‘98), Lund, June 15–17, 1998.Google Scholar
  31. [31]
    W. Poeschel, R. Engel, D. Dopheide, H.J. Baade, H.J. Kecke, R. Praetor, N. Weist and E. Kurras, A Unique Fluid Diverter Design for Water Flow Calibration Facilities,” Proceedings of the 10th International Conference on Flow Measurement (FLOMEKO 2000), Salvador, June 5–8, 2000.Google Scholar
  32. [32]
    R. Engel, H.J. Baade, and A. Rubel, Performance Improvement of Liquid Flow Calibrators by Applying Special Measurement and Control Strategies, Proceedings of the 11th International Conference on Flow Measurement (FLOMEKO 2003), Groningen, May 12–14, 2003.Google Scholar
  33. [33]
    R. Engel and H.J. Baade, New-Design Dual-Balance Gravimetric Reference Systems with PTB’s New Hydrodynamic Test Field, Proceedings of the 11th International Conference on Flow Measurement (FLOMEKO 2003), Groningen, May 12–14, 2003.Google Scholar
  34. [34]
    R. Engel, PTB’s Hydrodynamic Test Field—Investigations to Verify the Measurement Uncertainty, Proceedings of the 12th International Conference on Flow Measurement (FLOMEKO 2004), Guilin, September 14–17, 2004.Google Scholar
  35. [35]
    R. Engel, Modeling the Uncertainty in Liquid Flowmeters Calibration and Application—Requirements and Their Technical Realization for PTB’s National Water Flow Standard, Proceedings of SENSOR Conference 2007, Nurnburg, May 22-24, 2007.Google Scholar
  36. [36]
    R. Engel and H.J. Baade, Model-Based Flow Diverter Analysis for an Improved Uncertainty Determination in Liquid Flow Calibration Facilities, Meas. Sci. Technol., 21 (2010) 1–11.CrossRefGoogle Scholar
  37. [37]
    V. Gowda, T.T. Yeh and P.I. Espina, The New NIST Water Flow Calibration Facility, Proceedings of the 11th International Conference on Flow Measurement (FLOMEKO 2003), Groningen, May 12–14, 2003.Google Scholar
  38. [38]
    N.P. Yende, T.T. Yeh and P.I. Espina, Performance of an Error Free Liquid Flow Diverter Design, Proceedings of the 11th International Conference on Flow Measurement (FLOMEKO 2003), Groningen, May 12–14, 2003.Google Scholar
  39. [39]
    I. Marfenko, T.T. Yeh and J. Wright, Diverter Uncertainty Less Than 0.01 % for Water Flow Calibrations, Proceedings of the 6th International Symposium for Fluid Flow, Querétario, May 16–18, 2006.Google Scholar
  40. [40]
    T. Shimada, S. Oda, Y. Terao and M. Takamoto, Development of a New Diverter System for Liquid Flow Calibration Facilities, Flow Meas. Instrum., 14 (2003) 89–96.CrossRefGoogle Scholar
  41. [41]
    R. Doihara, T. Shimada, Y. Terao and M. Takamoto, “Development of Weighing Tank System Employing Rotating Double Wing Diverter,” Flow Meas. Instrum., 17 (2006) 141–152.CrossRefGoogle Scholar
  42. [42]
    J.N. Som, R. Singh, V. Babu and I.S. Taak, Study on the Error Analysis and Uncertainty Parameters Involved on the Flow Rate Measurement with the NPL Primary Water Flow Measurement Facility, The 4th International Conference on Advances in Metrology (AdMet-2003), Delhi, Feburary 5–7, 2003.Google Scholar
  43. [43]
    J.N. Som, R. Singh, V. Babu and I.S. Taak, Environmental Conditions Required for Fluid Flow Measurement Laboratory with Special Reference to Water Flow Measurement Set-Up at NPL, MAPAN-J. Metrol. Soc. India, 13 (1998) 3–8.Google Scholar
  44. [44]
    S.K. Jaiswal, I.S. Taak, D. Sharma, C. Singh and A.K. Bandyopadhyay, The Fluid Flow Measurement Standard at National Physical Laboratory, New Delhi, Proceedings of NCSLI 2010 Workshop and Symposium, Providence, July 25–29, 2010.Google Scholar

Copyright information

© Metrology Society of India 2012

Authors and Affiliations

  • Shiv Kumar Jaiswal
    • 1
  • Sanjay Yadav
    • 1
  • A. K. Bandyopadhyay
    • 1
  • Ravinder Agarwal
    • 2
  1. 1.CSIR-National Physical Laboratory, India (CSIR-NPLI)New DelhiIndia
  2. 2.Thapar UniversityPatialaIndia

Personalised recommendations