Evaluating NeuroSENSE for assessing depth of hypnosis during desflurane anesthesia: an adaptive, randomized-controlled trial

  • Matthias GörgesEmail author
  • Nicholas C. West
  • Erin M. Cooke
  • Shanshan Pi
  • Rollin F. Brant
  • Guy A. Dumont
  • J. Mark Ansermino
  • Richard N. Merchant
Reports of Original Investigations

Abstract

Purpose

Processed electroencephalography (EEG) monitors support depth-of-hypnosis assessment during anesthesia. This randomized-controlled trial investigated the performance of the NeuroSENSE electroencephalography (EEG) monitor to determine whether its wavelet anesthetic value for central nervous system (WAVCNS) index distinguishes consciousness from unconsciousness during induction of anesthesia (as assessed by the anesthesiologist) and emergence from anesthesia (indicated by patient responsiveness), and whether it correlates with changes in desflurane minimum alveolar concentration (MAC) during maintenance of anesthesia.

Methods

EEG was collected using a fronto-temporal bilateral montage. The WAVCNS was continuously recorded by the NeuroSENSE monitor, to which the anesthesiologist was blinded. Anesthesia was induced with propofol/remifentanil and maintained with desflurane, with randomized changes of −0.4, 0, or +0.4 MAC every 7.5 min within the 0.8–1.6 MAC range, if clinically acceptable to the anesthesiologist. During emergence from anesthesia, desflurane was stepped down by 0.2 MAC every five minutes.

Results

Data from 75 patients with a median [interquartile range] age of 41[35-52] yr were obtained. The WAVCNS distinguished consciousness from unconsciousness as assessed by the anesthesiologist, with area under the receiver operating characteristic curve of 99.5% (95% confidence interval [CI], 98.5 to 100.0) at loss of consciousness and 99.4% (95% CI, 98.5 to 100.0) at return of consciousness. Bilateral WAVCNS changes correlated with desflurane concentrations, with −8.0 and −8.6 WAVCNS units, respectively, per 1 MAC change in the 0.8–1.6 MAC range during maintenance of anesthesia and −10.0 and −10.5 WAVCNS units, respectively, in the 0.4–1.6 MAC range including emergence from anesthesia.

Conclusion

The NeuroSENSE monitor can reliably discriminate between consciousness and unconsciousness, as assessed by the anesthesiologist, during induction of anesthesia and with a lower level of reliability during emergence from anesthesia. The WAVCNS correlates with desflurane concentration but plateaus at higher concentrations, similar to other EEG monitors, which suggests limited utility to titrate higher concentrations of anesthetic vapour.

Trial registration

clinicaltrials.gov, NCT02088671; registered 17 March, 2014.

Une évaluation du moniteur NeuroSENSE pour mesurer la profondeur de l’hypnose pendant une anesthésie au desflurane : une étude randomisée contrôlée adaptable

Résumé

Objectif

Cette étude randomisée contrôlée a évalué la performance du moniteur d’électroencéphalographie (EEG) NeuroSENSE afin de déterminer si sa valeur anesthésique d’ondelette pour l’indice du système nerveux central (wavelet anesthetic value for central nervous system, WAVSNC) était capable de distinguer la conscience de l’inconscience pendant l’induction de l’anesthésie (telle qu’évaluée par l’anesthésiologiste) et le réveil de l’anesthésie (tel qu’indiqué par la réactivité du patient), et si cette valeur était corrélée aux modifications de la concentration alvéolaire minimale (MAC) de desflurane pendant le maintien de l’anesthésie.

Méthode

Les données électro-encéphalographiques ont été récoltées à l’aide d’un monitorage fronto-temporal bilatéral. La WAVSNC a été enregistrée en continu par le moniteur NeuroSENSE, dont les données étaient tenues cachées de l’anesthésiologiste. L’anesthésie a été induite à l’aide de propofol / rémifentanil et maintenue avec du desflurane, avec des modifications randomisées de -0,4, 0 ou +0,4 MAC toutes les 7,5 min dans une plage de MAC de 0,8-1,6. Pendant le réveil de l’anesthésie, le desflurane a été réduit de 0,2 MAC toutes les cinq minutes.

Résultats

Les données de 75 patients d’un âge médian [écart interquartile] de 41 [35-52] ans ont été obtenues. La WAVSNC a différencié la conscience de l’inconscience avec une surface sous la courbe de fonction d’efficacité de l’observateur de 99,5 % (intervalle de confiance [IC] 95 %, 98,5 à 100,0) au moment de la perte de conscience et de 99,4 % (IC 95 %, 98,5 à 100,0) au rétablissement de la conscience. Les modifications bilatérales de la WAVSNC étaient corrélées aux concentrations de desflurane, avec des unités de WAVSNC de -8,0 et -8,6, respectivement par unité de MAC dans la plage de MAC de 0,8-1,6 pendant le maintien de l’anesthésie et de -10,0 et -10,5 unités WAVSNC, respectivement, dans la plage de MAC de 0,4-1,6, incluant le réveil de l’anesthésie.

Conclusion

Le moniteur NeuroSENSE peut distinguer de façon fiable un état de conscience d’un état d’inconscience pendant l’induction de l’anesthésie et, à un degré de fiabilité moindre, pendant le réveil de l’anesthésie. La WAVSNC est corrélée à la concentration de desflurane mais atteint un plateau à des concentrations plus élevées, tout comme les autres moniteurs EEG, ce qui suggère une utilité limitée pour titrer de plus fortes concentrations de gaz anesthésiques.

Enregistrement de l’étude

clinicaltrials.gov, NCT02088671; enregistrée le 17 mars 2014.

Notes

Author contributions

Matthias Görges conducted the literature review, designed the study, obtained the ethical approval to conduct the research, analyzed and interpreted the findings, and drafted the manuscript. Nicholas C. West collected the data and drafted the manuscript. Erin M. Cooke conducted the literature review, designed the study, obtained the ethical approval to conduct the research, and collected the data. Shanshan Pi analyzed and interpreted the findings and drafted the manuscript. Rollin F. Brant analyzed and interpreted the findings. Guy A. Dumont conducted the literature review, designed the study, and obtained the ethical approval to conduct the research. J. Mark Ansermino conducted the literature review, designed the study, and obtained the ethical approval to conduct the research. Richard N. Merchant conducted the literature review, designed the study, and obtained the ethical approval to conduct the research. All authors critically reviewed the manuscript.

Acknowledgements

The authors wish to thank all the participating patients, anesthesiologists, the surgical and nursing teams in the operating theatre, as well as Aryannah Rollinson and Sonia Brodie for their help with the data collection.

Declaration of interests

Guy A. Dumont is a co-inventor of the NeuroSENSE monitor (NeuroWave Systems Inc., Cleveland Heights, OH, USA). J. Mark Ansermino and Guy A. Dumont have consulted for NeuroWave Systems Inc. Matthias Görges, J. Mark Ansermino, Guy A. Dumont, and Richard N. Merchant are party to a licensing agreement between NeuroWave Systems Inc. and the University of British Columbia for control technology. The remaining authors have no disclosures.

Funding

This study was funded by NeuroWave Systems Inc. (Cleveland Heights, OH, USA), under contract from the US Army (W81XWH-06-C-0016).

Editorial responsibility

This submission was handled by Dr. Gregory L. Bryson, Deputy Editor-in-Chief, Canadian Journal of Anesthesia.

References

  1. 1.
    Russell IF. Midazolam-alfentanil: an anaesthetic? An investigation using the isolated forearm technique. Br J Anaesth 1993; 70: 42-6.CrossRefGoogle Scholar
  2. 2.
    Morgan GE Jr, Mikhail MS, Murray MJ. Clinical Anesthesiology. 4th ed. NY: McGraw-Hill Medical; 2005 .Google Scholar
  3. 3.
    Ferreira AL, Nunes C, Mendes JG, Amorim P. Do we have today a reliable method to detect the moment of loss of consciousness during induction of general anaesthesia? Rev Esp Anestesiol Reanim 2019; 66: 93-103.CrossRefGoogle Scholar
  4. 4.
    Bibian S, Dumont GA, Zikov T. Dynamic behavior of BIS, M-entropy and neuroSENSE brain function monitors. J Clin Monit Comput 2011; 25: 81-7.CrossRefGoogle Scholar
  5. 5.
    Marchant N, Sanders R, Sleigh J, et al. How electroencephalography serves the anesthesiologist. Clin EEG Neurosci 2014; 45: 22-32.CrossRefGoogle Scholar
  6. 6.
    Shander A, Lobel GP, Mathews DM. Brain monitoring and the depth of anesthesia: another Goldilocks dilemma. Anesth Analg 2018; 126: 705-9.CrossRefGoogle Scholar
  7. 7.
    Fahy BG, Chau DF. The technology of processed electroencephalogram monitoring devices for assessment of depth of anesthesia. Anesth Analg 2018; 126: 111-7.CrossRefGoogle Scholar
  8. 8.
    Musialowicz T, Lahtinen P. Current status of EEG-based depth-of-consciousness monitoring during general anesthesia. Curr Anesthesiol Rep 2014; 4: 251-60.CrossRefGoogle Scholar
  9. 9.
    Escallier KE, Nadelson MR, Zhou D, Avidan MS. Monitoring the brain: processed electroencephalogram and peri-operative outcomes. Anaesthesia 2014; 69: 899-910.CrossRefGoogle Scholar
  10. 10.
    Zorrilla-Vaca A, Healy RJ, Wu CL, Grant MC. Relation between bispectral index measurements of anesthetic depth and postoperative mortality: a meta-analysis of observational studies. Can J Anesth 2017; 64: 597-607.CrossRefGoogle Scholar
  11. 11.
    Zikov T, Bibian S, Dumont GA, Huzmezan M, Ries CR. Quantifying cortical activity during general anesthesia using wavelet analysis. IEEE Trans Biomed Eng 2006; 53: 617-32.CrossRefGoogle Scholar
  12. 12.
    Agrawal G, Bibian S, Zikov T. Recommended clinical range for WAVCNS index during general anesthesia. Proceedings of the 2010 Annual Meeting of the American Society of Anesthesiologists. San Diego, CA: American Society of Anesthesiologists; 2010: A1347 (abstract).Google Scholar
  13. 13.
    Bibian S, Zikov T. NeuroSENSE® Monitor with WAVCNS Cortical Quantifier: a Deterministic Approach to EEG Analysis. Cleveland Heights, OH: September 2018. Available from URL: http://www.neurowavesystems.com/pdfs/NeuroSENSE_White_Paper_September_2018.pdf (accessed September 2019).
  14. 14.
    West N, Dumont GA, van Heusden K, et al. Robust closed-loop control of induction and maintenance of propofol anesthesia in children. Paediatr Anaesth 2013; 23: 712-9.CrossRefGoogle Scholar
  15. 15.
    West N, van Heusden K, Görges M, et al. Design and evaluation of a closed-loop anesthesia system with robust control and safety system. Anesth Analg 2018; 127: 883-94.CrossRefGoogle Scholar
  16. 16.
    Bresson J, Gayat E, Agrawal G, et al. A randomized controlled trial comparison of NeuroSENSE and bispectral brain monitors during propofol-based versus sevoflurane-based general anesthesia. Anesth Analg 2015; 121: 1194-201.CrossRefGoogle Scholar
  17. 17.
    Johansen JW. Update on bispectral index monitoring. Best Pract Res Clin Anaesthesiol 2006; 20: 81-99.CrossRefGoogle Scholar
  18. 18.
    Zikov T, Bibian S, Ries CR, Jaromczyk BS, Voltz DM. Performance of BIS and WAVCNS to predict loss of consciousness during routine propofol induction. Proceedings of the 2006 Annual Meeting of the American Society Anesthesiologists. American Society of Anesthesiologists; 2006: A1030 (abstract).Google Scholar
  19. 19.
    Kreuer S, Bruhn J, Larsen R, Bialas P, Wilhelm W. Comparability of Narcotrend index and bispectral index during propofol anaesthesia. Br J Anaesth 2004; 93: 235-40.CrossRefGoogle Scholar
  20. 20.
    Robin X, Turck N, Hainard A, et al. pROC: an open-source package for R and S+ to analyze and compare ROC curves. BMC Bioinformatics 2011;  https://doi.org/10.1186/1471-2105-12-77.CrossRefPubMedPubMedCentralGoogle Scholar
  21. 21.
    Carpenter J, Bithell J. Bootstrap confidence intervals: when, which, what? A practical guide for medical statisticians. Stat Med 2000; 19: 1141-64.CrossRefGoogle Scholar
  22. 22.
    Carmichael FJ, Haas DA. General anesthetics. In: Kalant H, Roschlau WH (Eds). Principles of Medical Pharmacology. Oxford University Press; 1998: 278-92.Google Scholar
  23. 23.
    Liu J, Colditz GA. Optimal design of longitudinal data analysis using generalized estimating equation models. Biom J 2017; 59: 315-30.CrossRefGoogle Scholar
  24. 24.
    Liang KY, Zeger SL. Longitudinal data analysis using generalized linear models. Biometrika 1986; 73: 13-22.CrossRefGoogle Scholar
  25. 25.
    Kreuer S, Bruhn J, Ellerkmann R, Ziegeler S, Kubulus D, Wilhelm W. Failure of two commercial indexes and spectral parameters to reflect the pharmacodynamic effect of desflurane on EEG. J Clin Monit Comput 2008; 22: 149-58.CrossRefGoogle Scholar
  26. 26.
    Han DW, Linares-Perdomo OJ, Lee JS, Kim JH, Kern SE. Comparison between cerebral state index and bispectral index as measures of electroencephalographic effects of sevoflurane using combined sigmoidal E(max) model. Acta Pharmacol Sin 2011; 32: 1208-14.CrossRefGoogle Scholar
  27. 27.
    Whitlock EL, Villafranca AJ, Lin N, et al. Relationship between bispectral index values and volatile anesthetic concentrations during the maintenance phase of anesthesia in the B-Unaware trial. Anesthesiology 2011; 115: 1209-18.PubMedPubMedCentralGoogle Scholar
  28. 28.
    Kanazawa S, Oda Y, Maeda C, Okutani R. Electroencephalographic effect of age-adjusted 1 MAC desflurane and sevoflurane in young, middle-aged, and elderly patients. J Anesth 2017; 31: 744-50.CrossRefGoogle Scholar
  29. 29.
    Gupta M, Shri I, Sakia P, Govil D. Comparison of equi-minimum alveolar concentration of sevoflurane and isoflurane on bispectral index values during both wash in and wash out phases: a prospective randomised study. Indian J Anaesth 2015; 59: 79-84.CrossRefGoogle Scholar
  30. 30.
    Purdon PL, Sampson A, Pavone KJ, Brown EN. Clinical electroencephalography for anesthesiologists: part I: background and basic signatures. Anesthesiology 2015; 123: 937-60.CrossRefGoogle Scholar
  31. 31.
    White PF, Ma H, Tang J, Wender RH, Sloninsky A, Kariger R. Does the use of electroencephalographic bispectral index or auditory evoked potential index monitoring facilitate recovery after desflurane anesthesia in the ambulatory setting? Anesthesiology 2004; 100: 811-7.CrossRefGoogle Scholar
  32. 32.
    Ibrahim TH, Yousef GT, Hasan AM, Eldesuky HI. Effect of bispectral index monitoring on desflurane consumption and recovery time in morbidly obese patients undergoing laparoscopic sleeve gastrectomy. Anesth Essays Res 2013; 7: 89-93.CrossRefGoogle Scholar
  33. 33.
    Kreuer S, Bruhn J, Stracke C, et al. Narcotrend or bispectral index monitoring during desflurane-remifentanil anesthesia: a comparison with a standard practice protocol. Anesth Analg 2005; 101: 427-34.CrossRefGoogle Scholar
  34. 34.
    Luginbühl M, Wüthrich S, Petersen-Felix S, Zbinden AM, Schnider TW. Different benefit of bispectal index (BIS) in desflurane and propofol anesthesia. Acta Anaesthesiol Scand 2003; 47: 165-73.CrossRefGoogle Scholar
  35. 35.
    Sahinovic MM, Eleveld DJ, Kalmar AF, et al. Accuracy of the composite variability index as a measure of the balance between nociception and antinociception during anesthesia. Anesth Analg 2014; 119: 288-301.CrossRefGoogle Scholar
  36. 36.
    Aho AJ, Yli-Hankala A, Lyytikäinen LP, Jäntti V. Facial muscle activity, Response Entropy, and State Entropy indices during noxious stimuli in propofol-nitrous oxide or propofol-nitrous oxide-remifentanil anaesthesia without neuromuscular block. Br J Anaesth 2009; 102: 227-33.CrossRefGoogle Scholar
  37. 37.
    Mathews DM, Clark L, Johansen J, Matute E, Seshagiri CV. Increases in electroencephalogram and electromyogram variability are associated with an increased incidence of intraoperative somatic response. Anesth Analg 2012; 114: 759-70.CrossRefGoogle Scholar
  38. 38.
    Brodie SM, Görges M, Ansermino JM, Dumont GA, Merchant RN. Closed-loop control of total intravenous anesthesia during significant intraoperative blood loss: a case report. A A Case Rep 2017; 9: 239-43.CrossRefGoogle Scholar

Copyright information

© Canadian Anesthesiologists' Society 2019

Authors and Affiliations

  1. 1.Department of Anesthesiology, Pharmacology & TherapeuticsUniversity of British ColumbiaVancouverCanada
  2. 2.BC Children’s Hospital Research InstituteVancouverCanada
  3. 3.Department of StatisticsUniversity of British ColumbiaVancouverCanada
  4. 4.Department of Electrical EngineeringUniversity of British ColumbiaVancouverCanada
  5. 5.Department of AnesthesiaRoyal Columbian Hospital, Fraser Health AuthorityNew WestminsterCanada

Personalised recommendations