Patient prosthesis mismatch and its impact on left ventricular regression following aortic valve replacement in aortic stenosis patients

  • Abid IqbalEmail author
  • Varghese Thomas Panicker
  • Jayakumar Karunakaran
Original Article

Abstract

Introduction

Patient prosthesis mismatch is known to alter post-operative remodeling of left ventricle adversely in aortic stenosis patients. An indexed orifice area of 0.85 is considered as conventional cutoff for patient prosthesis mismatch based on hemodynamic principles. Many patients have smaller annulus and annulus enlargement techniques may be required to avoid this benchmark which complicates the surgery.

Aims and objectives

To determine incidence of patient prosthesis mismatch (PPM) in our population, to assess left ventricle (LV) regression and impact of indexed effective orifice area on LV regression, and to determine a minimum acceptable cut off indexed orifice area that will result in LV regression.

Materials and methods

A single-center retrospective observational study of all patients who underwent aortic valve replacement surgery for severe aortic stenosis between July 2015 and December 2015 was conducted. Patients who underwent concomitant revascularization or other valve surgery were excluded. Data regarding conventional risk factors for PPM, valve type, and indexed orifice area as well as pre-operative and late post-operative LV mass were collected.

Observations

Thirty-seven of the 91 patients satisfied inclusion criteria. All patients had significant regression in LV mass. The incidence of PPM was 68%. Regression of indexed left ventricular mass showed a positive correlation of with indexed effective orifice area of the prosthetic valve. The correlation coefficient is + 0.48 (95% CI 0.18–0.698). Left ventricular mass regression was significantly higher in patients with indexed effective orifice area more than 0.75 (68.89 ± 29 vs 122.55 ± 58.84, p = 0.028). Relative left ventricular regression also was significantly higher in patients with an indexed effective orifice area more than 0.75 (39.53 ± 13.13 vs 49.73 ± 15.56, p = 0.022). There is a positive correlation between left ventricular regression and the reduction in mean gradient achieved by valve replacement. The correlation coefficient was + 0.35 (95% CI 0.03–0.61).

Conclusions

The incidence of PPM was found to be comparable to literature. There is a positive correlation between left ventricular regression and indexed effective orifice area of the prosthetic valve. Left ventricular regression was significantly higher in patients with an indexed effective area more than 0.75. This can be considered as criteria for significant PPM in study population.

Keywords

Aortic valve surgery Patient prosthesis mismatch Prosthesis 

Notes

Acknowledgements

The authors thank Mr. Kulwant Singh Kappor, MSc (Biostatistics) for his help with the statistical analysis.

Compliance with ethical standards

Conflict of interest

The authors declare that they have no conflict of interest.

Ethical issues

Ethical approval with waiver of consent was obtained from Institutional Ethics Committee prior to commencement of study.

References

  1. 1.
    Nathaniel S, Saligram S, Innasimuthu AL. Aortic stenosis: an update. World J Cardiol. 2010;2:135–9.CrossRefGoogle Scholar
  2. 2.
    Manjunath CN, Srinivas P, Ravindranath KS, Dhanalakshmi C. Incidence and patterns of valvular heart disease in a tertiary care high-volume cardiac center: a single center experience. Indian Heart J. 2014;66:320–6.CrossRefGoogle Scholar
  3. 3.
    Apostolakis E, Baikoussis NG, Papakonstantinou NA, Goudevenos J. Patient-prosthesis mismatch and strategies to prevent it during aortic valve replacement. Hell J Cardiol. 2011;52:41–51.Google Scholar
  4. 4.
    Horstkotte D, Loogen F. The natural history of aortic valve stenosis. Eur Heart J. 1988;9:E57–64.CrossRefGoogle Scholar
  5. 5.
    Levy D, Garrison RJ, Savage DD, Kannel WB, Castelli WP. Left ventricular mass and incidence of coronary heart disease in an elderly cohort: the Framingham Heart Study. Ann Intern Med. 1989;110:101–7.CrossRefGoogle Scholar
  6. 6.
    Mohty D, Dumesnil JG, Echahidi N, et al. Impact of prosthesis-patient mismatch on long-term survival after aortic valve replacement: influence of age, obesity and left ventricular dysfunction. J Am Coll Cardiol. 2009;53:39–47.CrossRefGoogle Scholar
  7. 7.
    Rahimtoola SH. The problem of valve prosthesis-patient mismatch. Circulation. 1978;58:20–4.CrossRefGoogle Scholar
  8. 8.
    Pibarot P, Dumesnil JG. Prosthesis-patient mismatch: definition, clinical impact, and prevention. Heart. 2006;92:1022–9.CrossRefGoogle Scholar
  9. 9.
    Dumesnil JG, Honos GN, Lemieux M, Beauchemin J. Validation and applications of indexed aortic prosthetic valve areas calculated by Doppler echocardiography. J Am Coll Cardiol. 1990;16:637–43.CrossRefGoogle Scholar
  10. 10.
    Dumesnil JG, Yoganathan AP. Valve prosthesis hemodynamics and the problem of high transprosthetic pressure gradients. Eur J Cardiothorac Surg. 1992;6:S34–8.CrossRefGoogle Scholar
  11. 11.
    Bonderman D, Graf A, Kammerlander AA, et al. Factors determining patient-prosthesis mismatch after aortic valve replacement - a prospective cohort study. PLoS One. 2013;8:e81940.CrossRefGoogle Scholar
  12. 12.
    Devereux RB, Alonso DR, Lutas EM, et al. Echocardiographic assessment of left ventricular hypertrophy: comparison to necropsy findings. Am J Cardiol. 1986;57:450–8.CrossRefGoogle Scholar
  13. 13.
    Lang RM, Bierig M, Devereux RB, et al. Recommendations for chamber quantification: a report from the American Society of Echocardiography’s guidelines and standards committee and the chamber quantification writing group, developed in conjunction with the European Association of Echocardiography, a branch of the European Society of Cardiology. J Am Soc Echocardiogr. 2005;18:1440–63.CrossRefGoogle Scholar
  14. 14.
    Pibarot P, Dumesnil JG. The relevance of prosthesis-patient mismatch after aortic valve replacement. Nat Clin Pract Cardiovasc Med. 2008;5:764–5.CrossRefGoogle Scholar
  15. 15.
    Pibarot P, Dumesnil JG. Hemodynamic and clinical impact of prosthesis-patient mismatch in the aortic valve position and its prevention. J Am Coll Cardiol. 2000;36:1131–41.CrossRefGoogle Scholar
  16. 16.
    Walther T, Rastan A, Falk V, et al. Patient prosthesis mismatch affects short- and long-term outcomes after aortic valve replacement. Eur J Cardiothorac Surg. 2006;30:15–9.CrossRefGoogle Scholar
  17. 17.
    Kohsaka S, Mohan S, Virani S, et al. Prosthesis-patient mismatch affects long-term survival after mechanical valve replacement. J Thorac Cardiovasc Surg. 2008;135:1076–80.CrossRefGoogle Scholar
  18. 18.
    Howell NJ, Keogh BE, Ray D, et al. Patient-prosthesis mismatch in patients with aortic stenosis undergoing isolated aortic valve replacement does not affect survival. Ann Thorac Surg. 2010;89:60–4.CrossRefGoogle Scholar
  19. 19.
    Howell NJ, Keogh BE, Barnet V, et al. Patient-prosthesis mismatch does not affect survival following aortic valve replacement. Eur J Cardiothorac Surg. 2006;30:10–4.CrossRefGoogle Scholar
  20. 20.
    Hong S, Yi GJ, Youn YN, Lee S, Yoo KJ, Chang BC. Effect of the prosthesis-patient mismatch on long-term clinical outcomes after isolated aortic valve replacement for aortic stenosis: a prospective observational study. J Thorac Cardiov Surg. 2013;146:1098–104.CrossRefGoogle Scholar
  21. 21.
    Fuster RG, Montero Argudo JA, Albarova OG, et al. Patient-prosthesis mismatch in aortic valve replacement: really tolerable? Eur J Cardiothorac Surg. 2005;27:441–9.CrossRefGoogle Scholar
  22. 22.
    Ruel M, Al-Faleh H, Kulik A, Chan KL, Mesana TG, Burwash IG. Prosthesis-patient mismatch after aortic valve replacement predominantly affects patients with preexisting left ventricular dysfunction: effect on survival, freedom from heart failure, and left ventricular mass regression. J Thorac Cardiovasc Surg. 2006;131:1036–44.CrossRefGoogle Scholar
  23. 23.
    Kaminishi Y, Misawa Y, Kobayashi J, et al. Patient-prosthesis mismatch in patients with aortic valve replacement. Gen Thorac Cardiovasc Surg. 2013;61:274–9.CrossRefGoogle Scholar
  24. 24.
    Medalion B, Blackstone EH, Lytle BW, White J, Arnold JH, Cosgrove DM. Aortic valve replacement: is valve size important? J Thorac Cardiovasc Surg. 2000;119:963–74.CrossRefGoogle Scholar
  25. 25.
    LaPar DJ, Ailawadi G, Bhamidipati CM, et al. Small prosthesis size in aortic valve replacement does not affect mortality. Ann Thorac Surg. 2011;92:880–8.CrossRefGoogle Scholar
  26. 26.
    Sportelli E, Regesta T, Salsano A, et al. Does patient-prosthesis mismatch after aortic valve replacement affect survival and quality of life in elderly patients? J Cardiovasc Med. 2016;17:137–43.CrossRefGoogle Scholar
  27. 27.
    Hernández-Vaquero D, Llosa JC, Díaz R, et al. Impact of patient-prosthesis mismatch on 30-day outcomes in young and middle-aged patients undergoing aortic valve replacement. J Cardiothorac Surg. 2012;7:46.CrossRefGoogle Scholar
  28. 28.
    Hanayama N, Christakis GT, Mallidi HR, et al. Patient prosthesis mismatch is rare after aortic valve replacement: valve size may be irrelevant. Ann Thorac Surg. 2002;73:1822–9.CrossRefGoogle Scholar
  29. 29.
    Kato Y, Suehiro S, Shibata T, Sasaki Y, Hirai H. Impact of valve prosthesis-patient mismatch on long-term survival and left ventricular mass regression after aortic valve replacement for aortic stenosis. J Card Surg. 2007;22:314–9.CrossRefGoogle Scholar
  30. 30.
    Guo L, Zheng J, Chen L, et al. Impact of prosthesis–patient mismatch on short-term outcomes after aortic valve replacement: a retrospective analysis in East China. J Cardiothorac Surg. 2017;12:42.CrossRefGoogle Scholar
  31. 31.
    Head SJ, Mokhles MM, Osnabrugge RLJ, et al. The impact of prosthesis–patient mismatch on long-term survival after aortic valve replacement: a systematic review and meta-analysis of 34 observational studies comprising 27 186 patients with 133 141 patient-years. Eur Heart J. 2012;33:1518–29.CrossRefGoogle Scholar
  32. 32.
    Chen J, Lin Y, Kang B, Wang Z. Indexed effective orifice area is a significant predictor of higher mid- and long-term mortality rates following aortic valve replacement in patients with prosthesis-patient mismatch. Eur J Cardiothorac Surg. 2014;45:234–40.CrossRefGoogle Scholar
  33. 33.
    Dayan V, Vignolo G, Soca G, Paganini JJ, Brusich D, Pibarot P. Predictors and outcomes of prosthesis-patient mismatch after aortic valve replacement. JACC Cardiovasc Imaging. 2016;9:924–33.CrossRefGoogle Scholar
  34. 34.
    Joshi SS, Ashwini T, George A, Jagadeesh AM. Patient prosthesis mismatch after aortic valve replacement: an Indian perspective. Ann Card Anaesth. 2016;19:84–8.CrossRefGoogle Scholar
  35. 35.
    Singh A, Sinha VK, Khandekar J, Agrawal N, Patwardhan A, Khandeparkar J. Left ventricular mass regression following aortic valve replacement with mechanical valves. Indian J Thorac Cardiovasc Surg. 2006;22:121–5.CrossRefGoogle Scholar
  36. 36.
    Dare AJ, Veinot JP, Edwards WD, Tazelaar HD, Schaff HV. New observations on the etiology of aortic valve disease: a surgical pathologic study of 236 cases from 1990. Hum Pathol. 1993;24:1330–8.CrossRefGoogle Scholar

Copyright information

© Indian Association of Cardiovascular-Thoracic Surgeons 2018

Authors and Affiliations

  1. 1.Department of Cardio Vascular and Thoracic SurgerySree Chitra Tirunal Institute for Medical Sciences and TechnologyThiruvananthapuramIndia

Personalised recommendations