Physical Simulation of the Random Failure of Implanted Braided NiTi Stents

  • Klára Hirmanová
  • Jan Pilch
  • Jan Racek
  • Luděk Heller
  • Petr Šittner
  • Lukáš Recman
  • Martin Petrenec
  • Petr Sedlák
Article

Abstract

A problem of random clinical failures of the braided esophageal NiTi stents has been addressed by performing physical simulation experiments on helical NiTi springs loaded in cyclic tension in air, water, and simulated biological fluid. Strains and stresses involved in spring deformation were analyzed through simulation by FEM implemented SMA model. It was found that the fatigue life of NiTi springs is significantly lower in fluids than in the air pointing toward the corrosion fatigue mechanism. There is, however, a fatigue limit roughly corresponding to the onset of martensitic transformation in the wire, which is not common for corrosion fatigue. It is proposed that surface TiO2 oxide cracking plays major role in that. Once the oxide layer on the NiTi wire surface fractures, typically during the first mechanical cycle, cracks in the oxide layer periodically open and close during subsequent mechanical cycling. This leads to the localization of mechanical and corrosion attacks under the oxide cracked regions. Microcracks within the surface oxide layer crossing over into the NiTi matrix were indeed revealed by scanning electron microscopy of FIB sections of fatigued wires. A corrosion assisted mechanism for fatigue crack nucleation at the interface between the surface oxide and NiTi matrix is proposed based on the available evidence. The approach opens a space for a better assessment of the corrosion fatigue performance of superelastic NiTi and ultimately for estimation of the lifetime of implanted braided NiTi stents.

Keywords

electron microscopy failure analysis mechanical testing 

References

  1. 1.
    K. Volenec, On the use of Nitinol in Corrosive Environment in Clinical Practice, Proceedings of SMST, 2013Google Scholar
  2. 2.
    L. Zhu, C. Trepanier, A.R. Pelton, and J. Fino, Oxidation of Nitinol and its Effect on Corrosion Resistance, ASM Materials & Processes for Medical Device Conference, 2003Google Scholar
  3. 3.
    A. Undisz, F. Schrempel, W. Wesh, and M. Rettenmayer, In situ observation of surface oxide layers on medical grade Ni-Ti alloy during straining, J. Biomed. Mater. Res. A, 2009, 15, p 88Google Scholar
  4. 4.
    S.W. Robertson, A.R. Pelton, and R.O. Ritchie, Mechanical fatigue and fracture of Nitinol, Int. Mater. Rev., 2012, 57, p 1–36CrossRefGoogle Scholar
  5. 5.
    Y. Kim and S. Miyazaki, Fatigue Life of Ti-50 at.% Ni and Ti-40Ni-10Cu (at.%) Shape Memory Alloy Wires, Proceedings of SMST 1997, USA, March 1997, p 473–478Google Scholar
  6. 6.
    A.R. Pelton, V. Schroeder, M.R. Mitchell, X.-Y. Gong, M. Barney, and S.W. Robertson, Fatigue and durability of nitinol stents, J. Mech. Behav. Biomed. Mater., 2008, 1, p 153–164CrossRefGoogle Scholar
  7. 7.
    J. Racek, P. Šittner, L. Heller, J. Pilch, M. Petrenec, and P. Sedlák, Corrosion of NiTi Wires with Cracked Oxide Layer, J. Mater. Eng. Perform., 2014. doi:10.1007/s11665-014-0925-8
  8. 8.
    S. Shabalovskaya, J. Anderegg, and J. Van Humbeeck, Critical Overview of Nitinol Surfaces and their Modifications for Medical Applications, Acta Biomater., 2008, 4, p 447–467CrossRefGoogle Scholar
  9. 9.
    L. Heller, P. Sittner, J. Pilch, Impact of Heat Effects on Superelasticity, Proceedings of ICOMAT 2008, Wiley, 2013Google Scholar
  10. 10.
    P. Sedlak, M. Frost, B. Benesova, P. Sittner, and T. Ben Zineb, Thermomechanical Model for NiTibased Shape Memory Alloys Including R-Phase and Material Anisotropy Under Multi-Axial Loadings, Int. J. Plast., 2012, 39, p 132–151CrossRefGoogle Scholar
  11. 11.
    P. Sedlák, M. Frost, A. Kruisová, K. Hiřmanová, L. Heller, and P. Šittner, Simulations of Mechanical Response of Superelastic NiTi Helical Spring and Its Relation to Fatigue Resistance, J. Mater. Eng. Perform., 2014. doi:10.1007/s11665-014-0906-y
  12. 12.
    B. Reedlunn, Ch. B. Churchill, E.E. Nelson, J.A. Shaw, and S.H. Daly, Tension, Compression, and Bending of Superelastic Shape Memory Alloy Tubes, J. Mech. Phys. Solids, 2014, 63, p 506–537Google Scholar
  13. 13.
    O.W. Bertacchini, D.C. Lagoudas, and E. Patoor, Thermomechanical Transformation Fatigue of TiNiCu SMA Actuators Under a Corrosive Environment—Part I: Experimental Results, Int. J. Fatigue, 2009, 31, p 1571–1578CrossRefGoogle Scholar
  14. 14.
    M.M. Patel, R.F.Gordon, An Investigation of Diverse Surface Finishes on Fatigue Properties of Superelastic Nitinol Wire, Proceedings of SMST (2006)Google Scholar
  15. 15.
    C.C. Shih, S.J. Lin, K.H. Chng, Y.L. Chen, and Y.Y. Su, Increased Corrosion Resistance of Stent Materials by Converting Current Surface Film of Polycrystalline Oxide into Amorphous Oxide, J. Biomed. Mater. Res., 2000, 52, p 323–332CrossRefGoogle Scholar

Copyright information

© ASM International 2014

Authors and Affiliations

  • Klára Hirmanová
    • 1
  • Jan Pilch
    • 1
  • Jan Racek
    • 1
  • Luděk Heller
    • 1
  • Petr Šittner
    • 1
  • Lukáš Recman
    • 2
  • Martin Petrenec
    • 3
  • Petr Sedlák
    • 4
  1. 1.Functional Materials DepartmentInstitute of Physics ASCRPragueCzech Republic
  2. 2.Ella-CSHradec KraloveCzech Republic
  3. 3.TESCANBrnoCzech Republic
  4. 4.Institute of Thermomechanics ASCRPragueCzech Republic

Personalised recommendations