Robotically assisted long bone biopsy under MRI: cadaver study results

  • Sunghwan Lim
  • Karun Sharma
  • Pan Li
  • Doru Petrisor
  • Stanley Fricke
  • Dan Stoianovici
  • Kevin ClearyEmail author
Original Article
  • 108 Downloads

Abstract

Rationale and objectives

We have designed and constructed an MR-safe robot made entirely of nonmetallic components with pneumatic actuators and optical encoders. The robot was developed to enable bone biopsies to be performed under magnetic resonance imaging (MRI) guidance in pediatric patients. The purpose of this study was to show the feasibility of using the robot for biopsy of the femur and tibia in a cadaver leg. Our long-term goal is to eliminate radiation exposure during bone biopsy procedures and provide more timely and accurate diagnosis for children with bone cancers and bone infections.

Methods

The MR-safe robot was mounted on the MRI table. A cadaver leg was procured from an anatomy supply house and placed on the MRI table. All required hospital precautions for infection control were taken. A total of 10 biopsy targets were sampled using MRI guidance: five from the femur and five from the tibia. A handheld, commercially available battery-powered bone drill was used to facilitate drilling through the cortex. After the study, the leg was scanned with CT to better visualize and document the bone biopsy sites. Both the MRI and CT images were used to analyze the results.

Results

All of the targets were successfully reached with an average targeting accuracy of 1.43 mm. A workflow analysis showed the average time for the first biopsy was 41 min including robot setup time and 22 min for each additional biopsy including the time for the repeat MRI scan used to confirm accurate targeting. The robot was shown to be MRI transparent, as no image quality degradation due to the use of the robot was detected.

Conclusion

The results showed the feasibility of using an MR-safe robotic system to assist the interventional radiologist in performing precision bone biopsy under MRI guidance. Future work will include developing an MR-safe drill, improving the mounting of the robot and fixation of the leg, and moving toward first in child clinical trials.

Keywords

MRI Long bone biopsy MR-safe Robot Cadaver 

Notes

Acknowledgements

We would also like to thank the MRI technologists Lu Vargas and Shena Phillips for their help with the experimental studies, as well as Dr. Xiaoyan Song from infection control.

Funding

This research was funded by the NIH Grant R01CA172244, Pneumatic Robot for MRI-Guided Long Bone Biopsy. The MR-safe robot technology developments were supported in part by awards RC1EB010936 from the National Institute of Biomedical Imaging and Bioengineering, and W81XWH0810221 from the Department of Defense. This research was also supported by the Intramural Research Program of the National Institutes of Health, National Heart, Lung, and Blood Institute, HHSN268200900052C.

Compliance with ethical standards

Conflict of interest

The authors declare that they have no conflict of interest.

Human and animal rights

This article does not contain any studies with human participants or animals performed by any of the authors.

References

  1. 1.
    National Cancer Institute. Cancer in Children and Adolescents [Internet]. National Cancer Institute. [cited 2016 Dec 10]. Available from: https://www.cancer.gov/types/childhood-cancers/child-adolescent-cancers-fact-sheet
  2. 2.
    Vasudevan V, Cheung MC, Yang R, Zhuge Y, Fischer AC, Koniaris LG et al (2010) Pediatric solid tumors and second malignancies: characteristics and survival outcomes. J Surg Res 160(2):184–189CrossRefGoogle Scholar
  3. 3.
    Dartnell J, Ramachandran M, Katchburian M (2012) Haematogenous acute and subacute paediatric osteomyelitis: a systematic review of the literature. J Bone Joint Surg Br 94(5):584–595CrossRefGoogle Scholar
  4. 4.
    Tsekos NV, Khanicheh A, Christoforou E, Mavroidis C (2007) Magnetic resonance-compatible robotic and mechatronics systems for image-guided interventions and rehabilitation: a review study. Annu Rev Biomed Eng 9(1):351–387CrossRefGoogle Scholar
  5. 5.
    Gassert R, Burdet E, Chinzei K (2008) Opportunities and challenges in MR-compatible robotics. Eng Med Biol Mag IEEE 27(3):15–22CrossRefGoogle Scholar
  6. 6.
    Arnolli MM, Hanumara NC, Franken M, Brouwer DM, Broeders IA (2015) An overview of systems for CT-and MRI-guided percutaneous needle placement in the thorax and abdomen. Int J Med Robot 11(4):458–475CrossRefGoogle Scholar
  7. 7.
    Monfaredi R, Cleary K, Sharma KMRI (2018) Robots for needle-based interventions: systems and technology. Ann Biomed Eng Jun:19Google Scholar
  8. 8.
    Stoianovici D, Song D, Petrisor D, Ursu D, Mazilu D, Muntener M, Schar M, Patriciu A (2007) “MRI Stealth” robot for prostate interventions. Minim Invasive Ther Allied Technol MITAT Off J Soc Minim Invasive Ther. 16(4):241–248CrossRefGoogle Scholar
  9. 9.
    Melzer A, Gutmann B, Remmele T, Wolf R, Lukoscheck A, Bock M (2008) INNOMOTION for percutaneous image-guided interventions. IEEE Eng Med Biol Mag 27(3):66–73CrossRefGoogle Scholar
  10. 10.
    Fischer GS, Iordachita I, Csoma C, Tokuda J, DiMaio SP, Tempany CM (2008) MRI-compatible pneumatic robot for transperineal prostate needle placement. IEEEASME Trans Mechatron 13(3):295–305CrossRefGoogle Scholar
  11. 11.
    Krieger A, Iordachita II, Guion P, Singh AK, Kaushal A, Ménard C (2011) An MRI-compatible robotic system with hybrid tracking for MRI-guided prostate intervention. IEEE Trans Biomed Eng 58(11):3049–3060CrossRefGoogle Scholar
  12. 12.
    Walsh CJ, Shepard J-A, Gupta R, Hanumara NC, Slocum AH (2008) A patient-mounted, telerobotic tool for CT-guided percutaneous interventions. J Med Devices 2(1):011007CrossRefGoogle Scholar
  13. 13.
    Hungr N, Fouard C, Robert A, Bricault I, Cinquin P Interventional radiology robot for CT and MRI guided percutaneous interventions. In: Fichtinger G, Martel A, Peters T (eds) Medical image computing and computer-assisted intervention—MICCAI 2011 [Internet]. Springer Berlin Heidelberg; 2011 [cited 2014 Feb 28]. p. 137–44. (Lecture Notes in Computer Science). Available from: http://link.springer.com/chapter/10.1007/978-3-642-23623-5_18
  14. 14.
    Song S-E, Tokuda J, Tuncali K, Yamada A, Torabi M, Hata N. Design evaluation of a double ring RCM mechanism for robotic needle guidance in MRI-guided liver interventions. In: 2013 IEEE/RSJ international conference on intelligent robots and systems (IROS). 2013. p 4078–4083Google Scholar
  15. 15.
    Wu F, Torabi M, Yamada A, Golden A, Fischer GS, Tuncali K (2013) An MRI coil-mounted multi-probe robotic positioner for cryoablation. In: Proc ASME 2013 Int Des Engineering Tech Conf Comput Inf Eng ConfGoogle Scholar
  16. 16.
    Monfaredi R, Seifabadi R, Iordachita I, Sze R, Safdar N, Sharma K, Cleary K (2014) Design of a body mounted mri-compatible robot for needle guidance in percutaneous procedures such as arthrography. In: Proceedings of the 5th IEEE RAS/embs international conference on biomedical robotics and biomechatronics (BioRob2014)Google Scholar
  17. 17.
    Stoianovici D, Jun C, Lim S, Li P, Petrisor D, Fricke S, Sharma K, Cleary K Multi-imager compatible, MR safe, remote center of motion needle-guide robot. In: IEEE Trans Biomed Eng [Internet]. 2017 [cited 2017 Aug 27]; Available from: http://ieeexplore.ieee.org/abstract/document/7911199/
  18. 18.
    Cleary K, Lim S, Jun C, Monfaredi R, Sharma K, Fricke ST, Vargas L, Petrisor D, Stoianovici D (2018) Robotically assisted long bone biopsy under MRI imaging: workflow and preclinical study. Acad Radiol 25(1):74–81.  https://doi.org/10.1016/j.acra.2017.08.008 CrossRefGoogle Scholar
  19. 19.
    Stoianovici D, Whitcomb LL, Anderson JH, Taylor RH, Kavoussi LR A modular surgical robotic system for image guided percutaneous procedures. In: International conference on medical image computing and computer-assisted intervention [Internet]. Springer; 1998 [cited 2016 Dec 11]. p. 404–410. Available from: http://link.springer.com/chapter/10.1007/BFb0056225
  20. 20.
    Stoianovici D, Cleary K, Patriciu A, Mazilu D, Stanimir A, Craciunoiu N, Kavoussi LR (2003) AcuBot: a robot for radiological interventions. IEEE Trans Robot Autom 19(5):927–930CrossRefGoogle Scholar
  21. 21.
    Stoianovici D, Patriciu A, Petrisor D, Mazilu D, Kavoussi L (2007) A new type of motor: pneumatic step motor. IEEEASME Trans Mechatron. 12(1):98–106CrossRefGoogle Scholar
  22. 22.
    Stoianovici D, Kim C, Srimathveeravalli G, Sebrecht P, Petrisor D, Coleman J, Solomon SB, Hricak H (2014) MRI-safe robot for endorectal prostate biopsy. IEEEASME Trans Mechatron. 19(4):1289–1299CrossRefGoogle Scholar
  23. 23.
    Li G, Su H, Cole GA, Shang W, Harrington K, Camilo A (2015) Robotic system for MRI-guided stereotactic neurosurgery. IEEE Trans Biomed Eng 62(4):1077CrossRefGoogle Scholar
  24. 24.
    Abdelaziz ME, Groenhuis V, Veltman J, Siepel F, Stramigioli S (2017) Controlling the Stormram 2: An MRI-compatible robotic system for breast biopsy. In: IEEE international conference on robotics and automation (ICRA) 2017. p 1746–1753Google Scholar
  25. 25.
    Stoianovici D, Kim C, Petrisor D, Jun C, Lim S, Ball MW (2017) MR safe robot, FDA clearance, safety and feasibility of prostate biopsy clinical trial. IEEEASME Trans Mechatron 22(1):115–126CrossRefGoogle Scholar
  26. 26.
    Cohen MG, McMahon CJ, Kung JW, Wu JS (2016) Comparison of battery-powered and manual bone biopsy systems for core needle biopsy of sclerotic bone lesions. Am J Roentgenol 206(5):W83–W86CrossRefGoogle Scholar

Copyright information

© CARS 2018

Authors and Affiliations

  1. 1.Robotics Laboratory, Brady Urological InstituteJohns Hopkins UniversityBaltimoreUSA
  2. 2.Sheikh Zayed Institute for Pediatric Surgical InnovationChildren’s National Health SystemWashingtonUSA

Personalised recommendations