Articulated minimally invasive surgical instrument based on compliant mechanism

  • Jumpei Arata
  • Shinya Kogiso
  • Masamichi Sakaguchi
  • Ryu Nakadate
  • Susumu Oguri
  • Munenori Uemura
  • Cho Byunghyun
  • Tomohiko Akahoshi
  • Tetsuo Ikeda
  • Makoto Hashizume
Original Article

Abstract

Purpose

In minimally invasive surgery, instruments are inserted from the exterior of the patient’s body into the surgical field inside the body through the minimum incision, resulting in limited visibility, accessibility, and dexterity. To address this problem, surgical instruments with articulated joints and multiple degrees of freedom have been developed. The articulations in currently available surgical instruments use mainly wire or link mechanisms. These mechanisms are generally robust and reliable, but the miniaturization of the mechanical parts required often results in problems with size, weight, durability, mechanical play, sterilization, and assembly costs.

Methods

We thus introduced a compliant mechanism to a laparoscopic surgical instrument with multiple degrees of freedom at the tip. To show the feasibility of the concept, we developed a prototype with two degrees of freedom articulated surgical instruments that can perform the grasping and bending movements. The developed prototype is roughly the same size of the conventional laparoscopic instrument, within the diameter of 4 mm. The elastic parts were fabricated by Ni-Ti alloy and SK-85M, rigid parts ware fabricated by stainless steel, covered by 3D- printed ABS resin. The prototype was designed using iterative finite element method analysis, and has a minimal number of mechanical parts.

Results

The prototype showed hysteresis in grasping movement presumably due to the friction; however, the prototype showed promising mechanical characteristics and was fully functional in two degrees of freedom. In addition, the prototype was capable to exert over 15 N grasping that is sufficient for the general laparoscopic procedure. The evaluation tests thus positively showed the concept of the proposed mechanism.

Conclusion

The prototype showed promising characteristics in the given mechanical evaluation experiments. Use of a compliant mechanism such as in our prototype may contribute to the advancement of surgical instruments in terms of simplicity, size, weight, dexterity, and affordability.

Keywords

Minimally invasive surgery Articulated surgical instrument Compliant mechanism Robotic surgery 

References

  1. 1.
  2. 2.
  3. 3.
  4. 4.
    Miniati R, Dori F, Cecconi G, Frosini F, Sacca F, Gentili BG, Petrucci F, Franchi S, Gusinu R (2014) Hospital based economic assessment of robotic surgery. IFMBE Proc 41:1144–1146CrossRefGoogle Scholar
  5. 5.
  6. 6.
    Romanelli JR, Earle DB (2009) Single-port laparoscopic surgery: an overview. Surg Endosc 23:1419–1427CrossRefPubMedGoogle Scholar
  7. 7.
    Tacchino R, Greco F, Matera D (2009) Single-incision laparoscopic cholecystectomy: surgery without a visible scar. Surg Endosc 23:896–899CrossRefPubMedGoogle Scholar
  8. 8.
  9. 9.
  10. 10.
    Stammberger H, Posawetz W (1990) Functional endoscopic sinus surgery. Eur Arch Otohinolaryngol 247:63–76Google Scholar
  11. 11.
    Jho HD (1999) Endoscopic pituitary surgery. Pituitary 2:139–154CrossRefPubMedGoogle Scholar
  12. 12.
  13. 13.
  14. 14.
  15. 15.
    Lim JJB, Erdman AG (2003) A review of mechanism used in laparoscopic surgical instruments. Mech Mach Theory 38:1133–1147CrossRefGoogle Scholar
  16. 16.
    Howell LL (2001) Compliant mechanisms. Wiley-Interscience. ISBN 978-0-471-38478-6Google Scholar
  17. 17.
    Smith ST (2000) Flexures elements of elastic mechanisms. CRC Press. ISBN 90-5699-261-9Google Scholar
  18. 18.
    Lobontiu N (2002) Compliant mechanisms. CRC Press. ISBN 978-0-8493-1367-7Google Scholar
  19. 19.
    Lange D, Langelaar MN, Herder JL (2008) Towards the design of a statically balanced compliant laparoscopic grasper using topology optimization. In: Proceedings of ASME 2008 international design engineering technical conferences and computers and information in engineering conference (IDETC/CIE2008), pp 293–305Google Scholar
  20. 20.
    Sato R, Nokata M (2011) Development of one part grasping forceps for vascular catheter. J JSCAS 13(3):300–301 (in Japanese)Google Scholar
  21. 21.
    Arata J, Saito Y, Fujimoto H (2010) Outer shell type 2 DOF bending manipulator using spring-link mechanism for medical applications. In: Proceedings of international conference robotics and automation, pp 1041–1046Google Scholar
  22. 22.
    Arata J, Kogiso S (2013) Monolithically designed minimally invasive surgical tool using compliant mechanism. Int J CARS Suppl 1:S369–371Google Scholar
  23. 23.
    Arata J, Kogiso S, Sakaguchi M, Nakadate R, Oguri S, Uemura M, Cho B, Akahoshi T, Ikeda T, Hashizume M (2014) Articulated minimally invasive surgical tool for laparoscopy based on compliant mechanism. Int J CARS 9(Suppl 1):S183–184Google Scholar
  24. 24.
  25. 25.
    Tolou N, Herder JL (2008) Towards the design of a statically balanced compliant laparoscopic grasper using topology optimization. In: International design engineering technical conference and computers and information in engineering conference, pp 1–13Google Scholar
  26. 26.
    Kozuka H, Arata J, Okuda K, Onaga A, Ohno M, Sano A, Fujimoto H (2012) Compliant-parallel mechanism for high precision machine with a wide range of working area. In: Proceedings of international conference on intelligent robots and systems, pp 2519–2524Google Scholar

Copyright information

© CARS 2015

Authors and Affiliations

  • Jumpei Arata
    • 1
  • Shinya Kogiso
    • 3
  • Masamichi Sakaguchi
    • 3
  • Ryu Nakadate
    • 2
  • Susumu Oguri
    • 4
  • Munenori Uemura
    • 4
  • Cho Byunghyun
    • 4
  • Tomohiko Akahoshi
    • 4
  • Tetsuo Ikeda
    • 4
  • Makoto Hashizume
    • 2
  1. 1.Department of Mechanical Engineering, Faculty of EngineeringKyushu UniversityFukuokaJapan
  2. 2.Center for Advanced Medical InnovationKyushu UniversityFukuokaJapan
  3. 3.Department of Engineering Physics, Electronics and MechanicsNagoya Institute of TechnologyNagoyaJapan
  4. 4.Department of Advanced Medical Initiatives, Faculty of Medical SciencesKyushu UniversityFukuokaJapan

Personalised recommendations