Comments on “A novel high input impedance front-end for capacitive biopotential measurement”

  • Original Article
  • Published:
Medical & Biological Engineering & Computing Aims and scope Submit manuscript

Abstract

A front-end for biopotential sensing in wearable medical devices has been recently proposed which is claimed to provide 100 GΩ input impedance by manually matching two resistor pairs in a positive- and a negative-feedback loop around an operational amplifier (op amp); the cost being that the equivalent input noise voltage doubles with respect to a simple non-inverting amplifier. The ECG acquired with capacitive (sic) electrodes through a cotton shirt is presented as a proof of the performance of the proposed circuit. It turns out, however, that the analysis ignores op amp’s input capacitance hence the effort to achieve a very high input resistance seems futile. Further, cotton is highly hygroscopic hence not an appropriate dielectric, so that there is no proof that the electrodes tested were actually capacitive. This comment addresses these two problems and some additional conceptual and methodological inaccuracies found in the paper.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Institutional subscriptions

Similar content being viewed by others

References

  1. Wu R, Tang Y, Li Z, Zhang L, Yan F (2018) A novel high input impedance front-end for capacitive biopotential measurement. Med Biol Eng Comput 56:1343–1355. https://doi.org/10.1007/s11517-017-1781-0

    Article  PubMed  Google Scholar 

  2. Chi YM, Jung TP, Cauwenberghs G (2010) Dry-contact and noncontact biopotential electrodes: methodological review. IEEE Reviews Biomed Eng 3:106–119

    Article  Google Scholar 

  3. Pallàs Areny R (2017) Comments on “capacitive biopotential measurement for electrophysiological signal acquisition: a review”. IEEE Sensors J 17:2607–2609

    Article  Google Scholar 

  4. Fernandez M, Pallàs Areny R (2000) Ag-AgCl electrode noise in high-resolution ECG measurements. Biomed Instrum Tech 34:125–130

    CAS  Google Scholar 

  5. Huigen E, Peper A, Grimbergen CA (2003) Investigation into the noise of surface electrodes. Med Biol Eng Comput 40:332–338

    Article  Google Scholar 

  6. Buxi D, Kim S, van Helleputte N, Altini M, Wijsman J, Yazicioglu RF, Penders J, van Hoof C (2012) Correlation between electrode-tissue impedance and motion artifact in biopotential recordings. IEEE Sensors J 12:3373–3383

    Article  Google Scholar 

  7. Motchenbacher CD, Connelly JA (1993) Low-noise electronic design. John Wiley & Sons, New York

    Google Scholar 

  8. Chi YM, Maier M, Cauwenberghs G (2011) Ultra-high input impedance, low noise integrated amplifier for noncontact biopotential sensing. IEEE Journal on Selected Topics in Circuits and Systems 1:526–535

    Article  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Additional information

Publisher’s note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Pallàs Areny, R. Comments on “A novel high input impedance front-end for capacitive biopotential measurement”. Med Biol Eng Comput 58, 267–269 (2020). https://doi.org/10.1007/s11517-019-02083-x

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s11517-019-02083-x

Keywords

Navigation