Micromotion of implant-abutment interfaces (IAI) after loading: correlation of finite element analysis with in vitro performances

  • Zhi Li
  • Shanshan Gao
  • Hongyu Chen
  • Ruiyang Ma
  • Tingting Wu
  • Haiyang Yu
Original Article
  • 66 Downloads

Abstract

Micromotion between IAI affects long-term survival rate of dental implants. The use of practical implants for mechanical test is costly. Finite element analysis (FEA) could test the micron level deformation changes, but whether it reflects the in vitro mechanical performances remains unknown. This study aims to investigate the correlation between IAI micromotion of FEA and in vitro performances. The two-step–two-component FEA method was used to test the relative deformation between IAI for three implant designs (M1, M2, and M3) during torque loading and cyclic oblique loading. The micromotion was divided into directions that perpendicular to (x-axis) and parallel to (y-axis) IAI. In vitro experiments on the micromotion relevant performance of IAI microleakage (tested by toluidine blue releasing with a spectrometer) and IAI locked condition (tested by abutment removal force tests after detaching the central screws) were also conducted for the identical implant systems (G1, G2, and G3). One-way ANOVA and Pearson’s correlation tests were performed for data analysis. FEA illustrated that the three implant systems performed different micromotion patterns. Significant differences were found in the IAI microleakage and removal force among the groups. Positive correlations were found between FEA and in vitro outcomes. Therefore, the two-step–two-component FEA method is an appropriate method to evaluate the IAI micromotion after loading.

Graphical abstract

The correspondence of IAI micromotion between FEA analysis and in vitro performances.

Keywords

Dental implants Finite element analysis Microleakage Micromotion Locked condition 

Notes

Acknowledgements

The authors gratefully thank Shandong Weigao Group Medical Polymer Co., Ltd. (Weihai, Shandong, China) for providing WEGO implants.

Funding

This study was supported by the National Natural Science Foundation of China (No. 81771113), the National Key Research and Development Project (2016YFC1102704), the Provincial Key Technology Research and Development Program of the Ministry of Science and Technology of Sichuan Province (No. 2014SZ0037), and the Sichuan Province Science and Technology Innovation Team Program (2011JTD0006).

Compliance with ethical standards

Conflict of interest

The authors declare that they have no conflict of interest.

Supplementary material

11517_2018_1937_MOESM1_ESM.xlsx (11 kb)
ESM 1 (XLSX 11 kb)

References

  1. 1.
    14801:2007(E) I. (2007) Dentistry - implants- dynamic fatigue test for endosseous dental implantsGoogle Scholar
  2. 2.
    Abdelhamed MI, Galley JD, Bailey MT, Johnston WM, Holloway J, McGlumphy E, Leblebicioglu B (2015) A comparison of zirconia and titanium abutments for microleakage. Clin Implant Dent Relat Res 17(Suppl 2):e643–e651CrossRefGoogle Scholar
  3. 3.
    Aguirrebeitia J, Muftu S, Abasolo M, Vallejo J (2014) Experimental study of the removal force in tapered implant-abutment interfaces: a pilot study. J Prosthet Dent 111:293–300CrossRefGoogle Scholar
  4. 4.
    Aguirrebeitia J, Abasolo M, Muftu S, Vallejo J (2017) Influence of design and clinical factors on the removal force ratio in tapered implant-abutment interfaces. J Prosthet Dent 117:493–498CrossRefGoogle Scholar
  5. 5.
    Algraffee H, Borumandi F, Cascarini L (2012) Peri-implantitis. Br J Oral Maxillofac Surg 50:689–694CrossRefGoogle Scholar
  6. 6.
    Binon P (2000) Implants and components: entering the new millennium. Int J Oral Maxillofac Implants 15:76–94PubMedGoogle Scholar
  7. 7.
    Bozkaya De (2004) Efficiency considerations for the purely tapered interference fit (TIF) abutments used in dental implants. J Biomech Eng 126:393–401CrossRefGoogle Scholar
  8. 8.
    da Silva-Neto JP, Prudente MS, Dantas TS, Senna PM, Ribeiro RF, das Neves FD (2017) Microleakage at different implant-abutment connections under unloaded and loaded conditions. Implant Dent 26:388–392CrossRefGoogle Scholar
  9. 9.
    Gehrke SA, Poncio da Silva PM, Calvo Guirado JL, Delgado-Ruiz RA, Dedavid BA, Aline Nagasawa M, Shibli JA (2016) Mechanical behavior of zirconia and titanium abutments before and after cyclic load application. J Prosthet Dent 116:529–535CrossRefGoogle Scholar
  10. 10.
    Gomes RS, Bergamo ETP, Bordin D, Del Bel Cury AA (2017) The substitution of the implant and abutment for their analogs in mechanical studies: in vitro and in silico analysis. Mater Sci Eng C Mater Biol Appl 75:50–54CrossRefGoogle Scholar
  11. 11.
    Gratton D, Aquilino S, Stanford C (2001) Micromotion and dynamic fatigue properties of the dental implant-abutment interface. J Prosthet Dent 85:47–52CrossRefGoogle Scholar
  12. 12.
    Grobecker-Karl T, Karl M (2017) Correlation between micromotion and gap formation at the implant-abutment Interface. Int J Prosthodont 30:150–152CrossRefGoogle Scholar
  13. 13.
    Hagiwara Y (2010) Does platform switching really prevent crestal bone loss around implants? Jpn Dent Sci Rev 46:122–131CrossRefGoogle Scholar
  14. 14.
    Heydecke G, Zwahlen M, Nicol A, Nisand D, Payer M, Renouard F, Grohmann P, Muhlemann S, Joda T (2012) What is the optimal number of implants for fixed reconstructions: a systematic review. Clin Oral Implants Res 23(Suppl 6):217–228CrossRefGoogle Scholar
  15. 15.
    Jorn D, Kohorst P, Besdo S, Borchers L, Stiesch M (2016) Three-dimensional nonlinear finite element analysis and microcomputed tomography evaluation of microgap formation in a dental implant under oblique loading. Int J Oral Maxillofac Implants 31:e32–e42CrossRefGoogle Scholar
  16. 16.
    Karl M, Taylor TD (2014) Parameters determining micromotion at the implant-abutment interface. Int J Oral Maxillofac Implants 29:1338–1347CrossRefGoogle Scholar
  17. 17.
    Karl M, Taylor TD (2016) Effect of cyclic loading on micromotion at the implant-abutment Interface. Int J Oral Maxillofac Implants 31:1292–1297CrossRefGoogle Scholar
  18. 18.
    Khorshidi H, Raoofi S, Moattari A, Bagheri A, Kalantari MH (2016) In vitro evaluation of bacterial leakage at implant-abutment connection: an 11-degree Morse taper compared to a butt joint connection. Int J Biomater 2016:8527849CrossRefGoogle Scholar
  19. 19.
    Khraisat A (2013) Two implant/abutment joint designs: a comparative finite element analysis. Int J Oral Maxillofac Implants 28:e83–e87CrossRefGoogle Scholar
  20. 20.
    Lee CK, Karl M, Kelly JR (2009) Evaluation of test protocol variables for dental implant fatigue research. Dental Materials: Official Publication of the Academy of Dental Materials 25:1419–1425CrossRefGoogle Scholar
  21. 21.
    Lin C, Chang C, Cheng C, Wang C, Lee H (1999) Automatic finite element mesh generation for maxillary second premolar. Comput Methods Prog Biomed 59:187–195CrossRefGoogle Scholar
  22. 22.
    Meng J, Everts J, Qian F, Gratton D (2007) Influence of connection geometry on dynamic micromotion at the implant-abutment interface. Int J Prosthodont 20:623–625PubMedGoogle Scholar
  23. 23.
    Messias A, Rocha S, Calha N, Neto MA, Nicolau P, Guerra F (2017) Effect of intentional abutment disconnection on the micro-movements of the implant-abutment assembly: a 3D digital image correlation analysis. Clin Oral Implants Res 28:9–16CrossRefGoogle Scholar
  24. 24.
    Ormianer Z, Palti A, Demiralp B, Heller G, Lewinstein I, Khayat P (2012) Implant-supported first molar restorations: correlation of finite element analysis with clinical outcomes. Int J Oral Maxillofac Implants 27:e1–e12PubMedGoogle Scholar
  25. 25.
    Passos SP, Gressler May L, Faria R, Ozcan M, Bottino MA (2013) Implant-abutment gap versus microbial colonization: clinical significance based on a literature review. J Biomed Mater Res B Appl Biomater 101:1321–1328CrossRefGoogle Scholar
  26. 26.
    Pintinha M, Camarini ET, Sabio S, Pereira JR (2013) Effect of mechanical loading on the removal torque of different types of tapered connection abutments for dental implants. J Prosthet Dent 110:383–388CrossRefGoogle Scholar
  27. 27.
    Pjetursson BE, Thoma D, Jung R, Zwahlen M, Zembic A (2012) A systematic review of the survival and complication rates of implant-supported fixed dental prostheses (FDPs) after a mean observation period of at least 5 years. Clin Oral Implants Res 23(Suppl 6):22–38CrossRefGoogle Scholar
  28. 28.
    Quirynen M, Steenberghe DV (1993) Bacterial colonization of the internal part of two-stage implants. An in vivo study. Clin Oral Implants Res 4:158–161CrossRefGoogle Scholar
  29. 29.
    Quirynen M, Bollen C, Eyssen H, Steenberghe DV (1994) Microbial penetration along the implant components of the Brånemark system. An in vitro study. Clin Oral Implants Res 5:239–244CrossRefGoogle Scholar
  30. 30.
    Saidin S, Abdul Kadir MR, Sulaiman E, Abu Kasim NH (2012) Effects of different implant-abutment connections on micromotion and stress distribution: prediction of microgap formation. J Dent 40:467–474CrossRefGoogle Scholar
  31. 31.
    Santiago Junior JF, Verri FR, Almeida DA, de Souza Batista VE, Lemos CA, Pellizzer EP (2016) Finite element analysis on influence of implant surface treatments, connection and bone types. Mater Sci Eng C Mater Biol Appl 63:292–300CrossRefGoogle Scholar
  32. 32.
    Smith NA, Turkyilmaz I (2014) Evaluation of the sealing capability of implants to titanium and zirconia abutments against Porphyromonas gingivalis, Prevotella intermedia, and Fusobacterium nucleatum under different screw torque values. J Prosthet Dent 112:561–567CrossRefGoogle Scholar
  33. 33.
    Steinebrunner L, Wolfart S, Bössmann K, Kern M (2005) In vitro evaluation of bacterial leakage along the implant-abutment interface of different implant systems. Int J Oral Maxillofac Implants 20:875–881PubMedGoogle Scholar
  34. 34.
    Van Staden RC, Guan H, Loo YC (2006) Application of the finite element method in dental implant research. Comput Methods Biomech Biomed Eng 9:257–270CrossRefGoogle Scholar
  35. 35.
    Verri FR, Batista VE, Santiago JF Jr, Almeida DA, Pellizzer EP (2014) Effect of crown-to-implant ratio on peri-implant stress: a finite element analysis. Mater Sci Eng C Mater Biol Appl 45:234–240CrossRefGoogle Scholar
  36. 36.
    Wachtel A, Zimmermann T, Spintig T, Beuer F, Müller W, Schwitalla A (2016) A novel approach to prove bacterial leakage of implant-abutment connections in vitro. J Oral Implantol 42:452–457CrossRefGoogle Scholar
  37. 37.
    Winter W, Klein D, Karl M (2013) Effect of model parameters on finite element analysis of micromotions in implant dentistry. J Oral Implantol 39:23–29CrossRefGoogle Scholar
  38. 38.
    Wu J, Chen C, Yip S, Hsu M (2012) Stress distribution and micromotion analyses of immediately loaded implants of varying lengths in the mandible and fibular bone grafts: a three-dimensional finite element analysis. Int J Oral Maxillofac Implants 27:e77–e84PubMedGoogle Scholar
  39. 39.
    Wu T, Fan H, Ma R, Chen H, Li Z, Yu H (2017) Effect of lubricant on the reliability of dental implant abutment screw joint: an in vitro laboratory and three-dimension finite element analysis. Mater Sci Eng C Mater Biol Appl 75:297–304CrossRefGoogle Scholar
  40. 40.
    Zhuang LF, Watt RM, Mattheos N, Si MS, Lai HC, Lang NP (2016) Periodontal and peri-implant microbiota in patients with healthy and inflamed periodontal and peri-implant tissues. Clin Oral Implants Res 27:13–21CrossRefGoogle Scholar
  41. 41.
    Zipprich H, Weigl P, Lange B, Lauerr H-C (2007) Micromovements at the implant-abutment interface: measurement, causes, and consequence. Implantologie 15:31–46Google Scholar
  42. 42.
    Zipprich H, Miatke S, Hmaidouch R, Lauer HC (2016) A new experimental design for bacterial microleakage investigation at the implant-abutment interface: an in vitro study. Int J Oral Maxillofac Implants 31:37–44CrossRefGoogle Scholar

Copyright information

© International Federation for Medical and Biological Engineering 2019

Authors and Affiliations

  • Zhi Li
    • 1
    • 2
  • Shanshan Gao
    • 1
  • Hongyu Chen
    • 1
  • Ruiyang Ma
    • 1
  • Tingting Wu
    • 1
  • Haiyang Yu
    • 1
    • 3
  1. 1.State Key Laboratory of Oral Diseases, National Clinical Research Center for Oral Diseases, West China Hospital of StomatologySichuan UniversityChengduChina
  2. 2.Aerospace Center HospitalBeijingChina
  3. 3.ChengduPeople’s Republic of China

Personalised recommendations