Advertisement

Measurement Techniques

, Volume 61, Issue 12, pp 1166–1173 | Cite as

Measurement of the Efficiency of Detection by Single-Photon Counters Based on Avalanche Photodiodes by the Method of Spontaneous Parametric Down Conversion with Spectrally Asymmetric Channels

  • P. P. Gostev
  • D. P. Agapov
  • A. V. DeminEmail author
  • G. G. Levin
  • E. A. Mamonov
  • S. A. Magnitsky
Article
  • 17 Downloads

We consider a frequency-diaphragm technique of simultaneous standardless measurements of the efficiency of photon detection by two single-photon counters. This technique is based on the well-known scheme of measurements of the efficiency of detection of single photons with the help of photon pairs generated in a nonlinear crystal in the case of spontaneous parametric scattering. As a specific feature of the proposed procedure, we can mention the application of spectrally asymmetric optical channels. An experimental installation aimed at the realization of the proposed procedure in the degenerate mode of spontaneous parametric scattering is described. The results of simultaneous measurements of the detection efficiency of two single-photon counters at a wavelength of 810 nm are presented. The counters are constructed on the basis of SAP-500 silicon avalanche photodiodes with fiber inputs. It is shown that this technique can be also used for measuring the detection efficiency of single-photon counters sensitive in a more long-wave region of the spectrum. As compared with the pure diaphragm scheme, the proposed procedures has advantages in the simplicity and reliability of adjustment.

Keywords

single-photon counters spontaneous parametric scattering quantum efficiency photon detection efficiency 

References

  1. 1.
    S. A. Magnitskiy, D. N. Frolovtsev, D. P. Agapov, et al., “Metrology of single photons for quantum information technologies,” Izmer. Tekhn., No. 3, 24–29 (2017).Google Scholar
  2. 2.
    D. C. Burnham and D. L. Weinberg, “Observation of simultaneity in parametric production of optical photon pairs,” Phys. Rev. Lett., 25, No. 2, 84–87 (1970).ADSCrossRefGoogle Scholar
  3. 3.
    A. N. Penin and A. Sergienko, “Absolute standardless calibration of photodetectors based on quantum two-photon fields,” Appl. Opt., 30, No. 25, 3582–3588 (1991).ADSCrossRefGoogle Scholar
  4. 4.
    P. G. Kwiat, A. M. Steinberg, R. Y. Chiao, et al., “Absolute efficiency and time-response measurement of single-photon detectors,” Appl. Opt., 33, No. 10, 1844–1853 (1994).ADSCrossRefGoogle Scholar
  5. 5.
    V. M. Ginzburg, N. G. Keratishvili, Ye. L. Korzhenevich, et al., “Absolute measurement of quantum efficiency based on parametric down-conversion effect,” Metrologia, 30, No. 4, 367–368 (1993).ADSCrossRefGoogle Scholar
  6. 6.
    A. Migdall, R. Datla, A. V. Sergienko, et al., “Measuring absolute infrared spectral radiance with correlated visible photons: technique verification and measurement uncertainty,” Appl. Opt., 37, No. 16, 3455–3463 (1998).ADSCrossRefGoogle Scholar
  7. 7.
    G. Brida, S. Castelletto, C. Novero, and M. L. Rastello, “Quantum-efficiency measurement of photodetectors by means of correlated photons,” JOSA B, 16, No. 10, 1623–1627 (1999).ADSCrossRefGoogle Scholar
  8. 8.
    M. Stipčević, H. Skenderović, and D. Gracin, “Characterization of a novel avalanche photodiode for single photon detection in VISNIR range,” Opt. Express., 18, No. 16, 17448–17459 (2010).ADSCrossRefGoogle Scholar
  9. 9.
    M. Ware and A. Migdall, “Single-photon detector characterization using correlated photons: the march from feasibility to metrology,” J. Modern Optics, 51, Nos. 9–10, 1549–1557 (2004).Google Scholar
  10. 10.
    A. P. Worsley, H. B. Coldenstrodt-Ronge, J. S. Lundeen, et al., “Absolute efficiency estimation of photon-number resolving detectors using twin beams,” Opt. Express, 17, No. 6, 4397–4412 (2009).ADSCrossRefGoogle Scholar
  11. 11.
    A. Avella, G. Brida, I. P. Degiovanni, et al., “Self-consistent, absolute calibration technique for photon number resolving detectors,” Opt. Express, 19, No. 23, 23249–23257 (2011).ADSCrossRefGoogle Scholar
  12. 12.
    J. Peřina, Jr., O. Haderka, A. Allevi, and M. Bondani, “Absolute calibration of photon-number-resolving detectors with an analog output using twin beams,” Appl. Phys. Lett., 104, No. 4, 041113-1–041113-4 (2014).Google Scholar
  13. 13.
    L. Cohen, Y. Pilnyak, D. Istrati, et al., “Absolute calibration of single-photon and multiplexed photon-number-resolving detectors,” Phys. Rev. A, 98, No. 1, 013811-1–013811-8 (2018).Google Scholar
  14. 14.
    S. Magnitskiy, D. Frolovtsev, V. Firsov, et al., “An SPDC-based source of entangled photons and its characterization,” J. Russ. Laser Res., 36, No. 6, 618–629 (2015).CrossRefGoogle Scholar
  15. 15.
    A. L. Migdall, “Absolute quantum effi ciency measurements using: toward a measurement protocol,” IEEE Trans. Instrum. Measur., 50, No. 2, 478–481 (2001).Google Scholar

Copyright information

© Springer Science+Business Media, LLC, part of Springer Nature 2019

Authors and Affiliations

  • P. P. Gostev
    • 1
    • 2
  • D. P. Agapov
    • 1
    • 2
  • A. V. Demin
    • 3
    Email author
  • G. G. Levin
    • 3
  • E. A. Mamonov
    • 1
    • 2
  • S. A. Magnitsky
    • 1
    • 2
  1. 1.Faculty of PhysicsLomonosov Moscow State UniversityMoscowRussia
  2. 2.International Laser CenterLomonosov Moscow State UniversityMoscowRussia
  3. 3.All-Russia Research Institute of Optophysical Measurements (VNIIOFI)MoscowRussia