Effects of indigo carmine intravenous injection on noninvasive and continuous total hemoglobin measurement with using the Revision L sensor

  • Tsuyoshi Isosu
  • Shinju Obara
  • Takahiro Hakozaki
  • Tsuyoshi Imaizumi
  • Yuzo Iseki
  • Midori Mogami
  • Satoshi Ohashi
  • Yukihiro Ikegami
  • Shin Kurosawa
  • Masahiro Murakawa
Letter to the Editor

Abstract

The effects of intravenous injection of indigo carmine on noninvasive and continuous total hemoglobin (SpHb) measurement were retrospectively evaluated with the Revision L sensor. The subjects were 18 patients who underwent elective gynecologic surgery under general anesthesia. During surgery, 5 mL of 0.4 % indigo carmine was injected intravenously, and changes in SpHb concentrations between before and after the injection were evaluated. The mean age was 52.4 ± 12.8 years. Before injection, the median SpHb level was 10.1 (range, 6.8–13.4) g/dL. The results demonstrated no change in SpHb concentration between before and after indigo carmine injection as detected by the Revision L sensor. SpHb measurements as determined with the Revision L sensor were not affected, even after the intravenous injection of indigo carmine.

Keyword

SpHb Revision L CO-oximetry Indigo carmine 

Notes

Compliance with ethical standards

Conflict of interest

The authors declare that they have no conflict of interest.

Ethical Approval

The study protocol has been approved by the Ethics Committee at Fukushima Medical University.

Financial Disclosure

The authors do not have a financial relationship with an organization.

References

  1. 1.
    Isosu T, Obara S, Hosono A, Ohashi S, Nakano Y, Imaizumi T, Mogami M, Murakawa M. Validation of continuous and noninvasive hemoglobin monitoring by pulse CO-oximetry in Japanese surgical patients. J Clin Monit Comput. 2013;27(1):55–60. doi: 10.1007/s10877-012-9397-2.CrossRefPubMedGoogle Scholar
  2. 2.
    Miller RD, Ward TA, Shiboski SC, Cohen NH. A comparison of three methods of hemoglobin monitoring in patients undergoing spine surgery. Anesth Analg. 2011;112(4):858–63. doi: 10.1213/ANE.0b013e31820eecd1.CrossRefPubMedGoogle Scholar
  3. 3.
    Macknet MR, Allard M, Applegate RL II, Rook J. The accuracy of noninvasive and continuous total hemoglobin measurement by pulse CO-oximetry in human subjects undergoing hemodilution. Anesth Analg. 2010;111(6):1424–6. doi: 10.1213/ANE.0b013e3181fc74b9.CrossRefPubMedGoogle Scholar
  4. 4.
    Berkow L, Rotolo S, Mirski E. Continuous noninvasive hemoglobin monitoring during complex spine surgery. Anesth Analg. 2011;113(6):1396–402. doi: 10.1213/ANE.0b013e318230b425.CrossRefPubMedGoogle Scholar
  5. 5.
    Frasca D, Dahyot-Fizelier C, Catherine K, Levrat Q, Debaene B, Mimoz O. Accuracy of a continuous noninvasive hemoglobin monitor in intensive care unit patients. Crit Care Med. 2011;39(10):2277–82. doi: 10.1097/CCM.0b013e3182227e2d.CrossRefPubMedGoogle Scholar
  6. 6.
    Scheller MS, Unger RJ, Kelner MJ. Effects of intravenously administered dynes on pulse oximetry readings. Anesthesiology. 1986;65(5):550–1.CrossRefPubMedGoogle Scholar

Copyright information

© Springer Science+Business Media Dordrecht 2016

Authors and Affiliations

  • Tsuyoshi Isosu
    • 1
  • Shinju Obara
    • 1
  • Takahiro Hakozaki
    • 1
  • Tsuyoshi Imaizumi
    • 1
  • Yuzo Iseki
    • 1
  • Midori Mogami
    • 1
  • Satoshi Ohashi
    • 1
  • Yukihiro Ikegami
    • 1
  • Shin Kurosawa
    • 1
  • Masahiro Murakawa
    • 1
  1. 1.Department of AnesthesiologyFukushima Medical UniversityFukushima CityJapan

Personalised recommendations