Effects of indigo carmine intravenous injection on noninvasive and continuous total hemoglobin measurement with using the Revision L sensor
- 116 Downloads
- 1 Citations
Abstract
The effects of intravenous injection of indigo carmine on noninvasive and continuous total hemoglobin (SpHb) measurement were retrospectively evaluated with the Revision L sensor. The subjects were 18 patients who underwent elective gynecologic surgery under general anesthesia. During surgery, 5 mL of 0.4 % indigo carmine was injected intravenously, and changes in SpHb concentrations between before and after the injection were evaluated. The mean age was 52.4 ± 12.8 years. Before injection, the median SpHb level was 10.1 (range, 6.8–13.4) g/dL. The results demonstrated no change in SpHb concentration between before and after indigo carmine injection as detected by the Revision L sensor. SpHb measurements as determined with the Revision L sensor were not affected, even after the intravenous injection of indigo carmine.
Keyword
SpHb Revision L CO-oximetry Indigo carmineNotes
Compliance with ethical standards
Conflict of interest
The authors declare that they have no conflict of interest.
Ethical Approval
The study protocol has been approved by the Ethics Committee at Fukushima Medical University.
Financial Disclosure
The authors do not have a financial relationship with an organization.
References
- 1.Isosu T, Obara S, Hosono A, Ohashi S, Nakano Y, Imaizumi T, Mogami M, Murakawa M. Validation of continuous and noninvasive hemoglobin monitoring by pulse CO-oximetry in Japanese surgical patients. J Clin Monit Comput. 2013;27(1):55–60. doi: 10.1007/s10877-012-9397-2.CrossRefPubMedGoogle Scholar
- 2.Miller RD, Ward TA, Shiboski SC, Cohen NH. A comparison of three methods of hemoglobin monitoring in patients undergoing spine surgery. Anesth Analg. 2011;112(4):858–63. doi: 10.1213/ANE.0b013e31820eecd1.CrossRefPubMedGoogle Scholar
- 3.Macknet MR, Allard M, Applegate RL II, Rook J. The accuracy of noninvasive and continuous total hemoglobin measurement by pulse CO-oximetry in human subjects undergoing hemodilution. Anesth Analg. 2010;111(6):1424–6. doi: 10.1213/ANE.0b013e3181fc74b9.CrossRefPubMedGoogle Scholar
- 4.Berkow L, Rotolo S, Mirski E. Continuous noninvasive hemoglobin monitoring during complex spine surgery. Anesth Analg. 2011;113(6):1396–402. doi: 10.1213/ANE.0b013e318230b425.CrossRefPubMedGoogle Scholar
- 5.Frasca D, Dahyot-Fizelier C, Catherine K, Levrat Q, Debaene B, Mimoz O. Accuracy of a continuous noninvasive hemoglobin monitor in intensive care unit patients. Crit Care Med. 2011;39(10):2277–82. doi: 10.1097/CCM.0b013e3182227e2d.CrossRefPubMedGoogle Scholar
- 6.Scheller MS, Unger RJ, Kelner MJ. Effects of intravenously administered dynes on pulse oximetry readings. Anesthesiology. 1986;65(5):550–1.CrossRefPubMedGoogle Scholar